两种光学生物测量仪和接触式A超对高度近视眼球生物学测量的一致性评价 被引量:1

Comparison and Evaluation of Ocular Biometry Using Lenstar LS 900, IOLMaster, and Ultrasound A-scan in Highly Myopic Eyes
下载PDF
导出
摘要 【目的】评估两种光学眼球生物测量仪Lenstar和IOLMaster与超声生物测量对于高度近视眼的眼球生物测量的准确性和一致性。【方法】对照观察分别由Lenstar、IOLMaster和接触式A超测量的高度近视患者的眼球生物学参数。三种方法的比较采用Friedman检验或Wilcoxon符号秩和检验,测量结果的重复性使用变异系数进行评价,采用Bland-Altman分析进行一致性评价。【结果】纳入了66个高度近视患者(66只眼),平均年龄(31.2±12.4)岁。Lenstar、IOLMaster和接触式A超测量的眼轴长度(AL)的变异系数分别为0.05%,0.07%和0.2%。AL的95%一致性界限(Lo A)范围分别为0.12 mm(Lenstar vs.IOLMaster),1.06 mm(Lenstar vs.A-scan),和1.14 mm(IOLMaster vs.A-scan),前房深度(ACD)的95%Lo A分别为0.28 mm(Lenstar vs.IOLMaster),0.51 mm(Lenstar vs.A-scan)和0.52 mm(IOLMaster vs.A-scan)。在极高度近视(AL≥27.0 mm)和巩膜后葡萄肿患眼中也得到相近的重复性和一致性结果。【结论】高度近视眼中,光学测量方法的测量结果一致性很好,能提供比超声生物测量重复性和准确性更好的眼球生物测量数据。 【Objective】 To evaluate the ocular biometric measurements in highly myopic eyes using a new optical low-coherence reflectometer-Lenstar and compare the measurements with those obtained from the IOLMaster and the applanation A-scan. 【Methods】In this prospective comparative observational study, a series of biometric measurements were taken for highly myopic subjects. The measurements obtained with the Lenstar were compared with those obtained with the IOLMaster or A-scan. The differences between three methods were assessed using Friedman test and Wilcoxon Signed-Rank test. Repeatability was evaluated by coefficient of variation(CV). Agreement was evaluated using Bland-Altman analyses. 【Results】 The study evaluated 66 highly myopic subjects(66eyes)(mean age 31.2 ± 12.4 years). The CVs for repeated axial length(AL) measurements in Lenstar, IOLMaster and A-scan were0.05%, 0.07%,and 0.2%, respectively. The spans of 95% limits of agreement(Lo A) for AL were 0.12 mm for Lenstar vs IOLMaster,1.06 mm Lenstar vs A-scan, and 1.14 mm IOLMaster vs A-scan. These values for anterior chamber depth(ACD) were 0.28 mm,0.51 mm, and 0.52 mm, respectively. The repeatability and agreement results were comparable in extremely myopic(AL ≥ 27.0mm) or staphylomatous eyes. 【Conclusion】 Optical biometry could provide more repeatable and interchangeable measurements in highly myopic eyes compared to ultrasound biometry.
出处 《中山大学学报(医学科学版)》 CAS CSCD 北大核心 2015年第5期774-779,共6页 Journal of Sun Yat-Sen University:Medical Sciences
基金 海南省卫生和计划生育委员会普通医学科研项目(琼卫2014-055)
关键词 高度近视 生物测量 光学低相干反射测量仪 光学相干生物测量仪 接触式A超 high myopia biometry optical low-coherence reflectometry partial coherence interferometry applanation A-scan
  • 相关文献

参考文献14

  • 1Rajan MS, Keilhom I, Bell JA. Partial coherence laser interferometry vs conventional ultrasound biometry in intraocular lens power calculations [ J ]. Eye (Lond), 2002, 16(5): 552-556.
  • 2Findl O, Drexler W, Menapace R, et al. Improved prediction of intraocular lens power using partial coherence interferometry [J]. J Cataract Refract Surg,2001, 27(6) : 861-867.
  • 3Bjelos Roncevic M, Busic M, Cima IK,et al. Intraobserver and interobserver repeatability of ocular components measurement in cataract eyes using a new optical low coherence reflectometer [J]. Graefes Arch Clin Exp Ophthalmol, 2010, 249( 1): 83-87.
  • 4沈沛阳,丁小虎,何明光,钟兴武,陈海波,邢健强.光学相干生物测量仪Lenstar LS 900与IOLMaster对眼球生物学参数测量的一致性评价[J].眼科新进展,2014,34(6):560-563. 被引量:4
  • 5Zaldivar R, Shuhz MC, Davidorf JM, et al. Intraocular lens power calculations in patients with extreme myopia[J].J Cataract Refract Surg, 2000, 26(5) : 668-674.
  • 6Tsang CS, Chong GS, Yiu EP, et al. Intraocular lens power calculation formulas in Chinese eyes with high axial myopia [J]. J Cataract Refract Surg, 2003, 29 (7) : 1358-1364.
  • 7Rohrer K, Frueh BE, Walti R, et al. Comparison and evaluation of ocular biometry using a new noncontact optical low-coherence retlectometer[J]. Ophthalmology, 2009, 116( 11 ) : 2087-2092.
  • 8Bjelos Roncevic M, Busic M, Cima I, et al. Comparison of optical low-coherence reflectometry and applanation ultrasound hiometry on intraocular lens power calculation [J]. Graefes Arch Clin Exp Ophthalmol, 2011, 249 ( 1 ) : 69-75.
  • 9Olsen T. Calculation of intraocular lens power: a review [J]. Acta Ophthalmol Seand, 2007, 85 (5) : 472-485.
  • 10Byrne S. A-scan axial eye length measurements: a handbook for IOL calculations [ M ]. Mars Hill : Grove Park Publishers, 1995 : 45-46.

二级参考文献16

  • 1Olsen T. Calculation of intraocular lens power: a review[J]. Acta Ophthalmol Scand ,2007,85 (5) :472-485.
  • 2Drexler W, Findl O, Menapace R, Rainer G, Vass C, Hitzenberger CK, et al. Partial coherence interferometry:a novel approach to biometry in cataract surgery [J]. Am J Ophthalmol, 1998,126 (4) :524-534.
  • 3Haigis W, Lege B, Miller N, Schneider B. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis [J]. Graefes Arch Clin Exp Ophthalmo1,2000 ,238 (9) :755-773.
  • 4Santodomingo-Rubido J, Mallen EA, Gilmartin B, Wofffsohn JS. A new non-contact optical device for ocular biometry[J]. Br J Ophthalmo1,2002,85 (4) :458-452.
  • 5Narvaez J, Cherwek DH, Stulting RD, Waldron R, Zimmerman G J, Wessels IF, et al. Comparing immersion ultrasound with par- tial coherence interferometry for intraocular lens power calculation[J]. Ophthalmic Surg Lasers Imaging,2008,39( 1 ) :30-34.
  • 6Lavanya R, Teo L, Friedman DS, Aung HT, Baskaran M, Gao H, et al. Comparison of anterior chamber depth measurements u- sing the IOLMaster, scanning peripheral anterior chamber depth analyser, and anterior segment optical coherence tomography [J]. Br J Ophthalmol,2007,91 ( 8 ) : 1023-1026.
  • 7Salouti R, Nowroozzadeh MH, Zamani M, Ghoreyshi M, Salouti R. Comparison of the ultrasonographic method with 2 partial coherence interferometry methods for intraocular lens power ca culation [J]. Optometry,2010,82 ( 3 ) : 140-147.
  • 8Olsen T. Prediction of the effective postoperative (intraocular lens) anterior chamber depth [J]. J Cataract Refract Surg ,2006 , 32(3) :419-424.
  • 9Ashwin PT, Shah S, Wolffsohn JS. Advances in cataract surgery [J]. Clin Exp Optom,2009,92(4) :333-342.
  • 10Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement [ J ]. Lancet. 1986.1 (8476) :307-310.

共引文献3

同被引文献10

引证文献1

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部