确定旋转放疗设备等中心点的平均矢量算法

Average vector method for determining isocenter of rotated radiotherapy equipment
原文传递
导出
摘要 对于旋转放射治疗设备,其等中心点位置对治疗精度存在重要影响。针对设备负载引起的机械结构变形对等中心误差的影响,提出了一种确定旋转放疗设备等中心点的平均矢量算法。以等中心点位置为优化变量,通过限定等中心点与不同位置射线向量的最大距离建立约束条件,以满足约束条件的等中心点与射线向量的平均距离为目标函数,得到最优等中心点位置和等中心包络线。实例分析验证了平均矢量算法的正确性和有效性。平均矢量法与矢量法得出的等中心包络线半径均为0.311mm,比节点法降低约15.5%;平均矢量法给出的等中心点距各射线向量的平均距离为0.206mm,比矢量法降低约11.2%,比节点法降低约28%。 The isocentric accuracy of rotated radiotherapy equipment significantly influences the treatment accuracy.The average vector method is used to find the optimal isocenter considering the mechanical deformation of the rotated radiotherapy equipment.The maximum distance between the isocenter which is treated as the optimization variable and the ray vectors is considered as a constraint.An objective function is then designed to calculate the average distance between the isocenter that meets the constraint and the ray vectors.Finally,the location of the optimal isocenter and the isocentric envelop are obtained.A test with this method gives the radius of the isocentric envelop as 0.311 mm which is equal to that given by the vector-end-effector method,and about 15.5%less than that given by the point-end-effector method.The results also show that the average distance between the optimal isocenter and the ray vectors is 0.206 mm, about 11.2% less than that by the vector-end-effector method and 28% less than that by the point-end-effector method.
出处 《清华大学学报(自然科学版)》 EI CAS CSCD 北大核心 2015年第7期744-749,共6页 Journal of Tsinghua University(Science and Technology)
基金 北京市科技计划项目(Z141100000514015) 清华大学摩擦学国家重点实验室重点项目(SKLT12A03) 清华大学自主科研计划课题学科交叉专项项目(20111081026)
关键词 放疗设备 等中心点 射线向量 平均矢量法 radiotherapy equipment isocenter ray vector average vector method
  • 相关文献

参考文献11

  • 1胡逸民,张红志,戴建荣.肿瘤放射物理[M].北京:原子能出版社,1999.538~605,383~389.
  • 2黄嘉华.放疗设备等中心精度分析与对策[J].上海生物医学工程,2003,24(1):10-14. 被引量:8
  • 3胡杰,陶建民,孙光荣.医用直线加速器等中心的质量控制和质量保证[J].中国医疗器械杂志,2007,31(3):213-215. 被引量:24
  • 4刘丕福,钟毓斌.等中心放射治疗技术[J].中日友好医院学报,1988,2(2):111-114.
  • 5DU Weiliang, GAO Song, WANG Xiaochun, et al. Quantifying the gantry sag on linear accelerators and introducing an MLC-based compensation strategy EJ]. Med Phys, 2012, 39(4): 2156-2162.
  • 6DU Weiliang, GAO Song. Measuring the wobble of radiation field centers during gantry rotation and collimator movement on a linear accelerator [J]. ivied Phys, 2011, 38(8) : 4575 - 4578.
  • 7Gonzalez A, Castro I, Martinez J A. A procedure to determine the radiation isocenter size in a linear accelerator [J]. Med Phys, 2004, 31(6): 1489-1493.
  • 8Tsai J S, Curran B H, Sternick E S, et al. The measurement of linear accelerator isocenter motion using a three-mieometer device and an adjustable pointer EJ3. Int J Radiat Oncol Biol Phys, 1996, 34(1): 189-195.
  • 9Skworcow P, Mills J A, Haas O C, et al. A new approach to quantify the mechanical and radiation isocentres of radiotherapy treatment machine gantries [J]. Phys Med Biol, 2007, 52(23): 7109- 7124.
  • 10Riis H L, Zimmermann S J, Hjelm-Hansen M. Gantry and isocenter displacements of a linear accelerator caused by an add-on micromutileaf collimator I-J]. Med Phys, 2013, 40(3) : 031707-1 - 031707-9.

二级参考文献2

共引文献31

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部