摘要 本文通过对全院临床科室多品牌的182台多参数监护仪进行质量检测,检测项目主要包括仪器的外观、用电安全、主要生理参数性能。对其中15台不合格的监护仪,查出其原因进行了校正,同时总结出多参数监护仪在使用中的注意事项,保证其安全有效地为临床服务。 Quality testing was performed for various brands of 182 multi-parameter monitors in the clinical departments of the hospital in this paper.The test items mainly included the instrument appearance,electrical safety and the main physiological parameters.As for the 15 sets of unqualified monitors,its reason was found and corrected.At the same time,the precautions during the use of multiparameter monitors were summed up to ensure its safety and effectiveness for clinical services.
作者 郑娆
出处 《中国医疗设备》 2015年第11期125-126,共2页 China Medical Devices
关键词 多参数监护仪 用电安全 医疗设备 质量控制 multi-parameter monitors electrical safety medical equipment quality control