期刊文献+

不同功率双极电凝对山羊肠系膜血管热损伤及止血效果的研究 被引量:1

Thermal damage of mesenteric vascular tissues and hemostatic effect in adult goats with bipolar electrocoagulation under different powers
下载PDF
导出
摘要 目的 研究双极电凝不同功率对山羊肠系膜血管组织的热损伤程度及对血管的止血效果的影响,指导腹腔镜手术中双极电凝功率的选择。方法 截取173段山羊肠系膜血管组织,以不同功率行肠系膜血管凝断术测量热损伤宽度;特定功率下电凝肠系膜血管组织记录单次电凝血管闭合成功率。结果 55~80 W之间,功率越小,热传导损伤宽度越小,差异有统计学意义(P〈0.05)。60 W功率双极单次电凝血管闭合成功率较高,热损伤宽度最小,差异有统计学意义(P〈0.05)。结论 双极电凝在60 W左右为较适合工作功率,热损伤范围小,血管止血效果好。 [Objective] To study the thermal damage of mesenteric vascular tissues and hemostatic effects on goats with bipolar elcctrocoagulation under different powers. [Methods] 173 case of mesenteric vessels of goats were performed bipolar electrocoagulations under different powers, and the thermal damage width and the success rates of electric coagulation of vascular occlusion for a single time were detected. [ Results] From 55W to 80W, the power was lower, the thermal damage width was smaller (P 〈 0.05). The success rate was the highest and thermal damage width was the smallest under the 60W (P 〈 0.05). [ Conclusion ] The 60W may decrease thermal damage and has better vascular closure.
出处 《中国内镜杂志》 北大核心 2015年第11期1130-1133,共4页 China Journal of Endoscopy
基金 河北省卫生厅指定课题(No:20120618)
关键词 热损伤 双极 腹腔镜 功率 thermal damage bipolar laparoscope power
  • 相关文献

参考文献10

  • 1HENIFORD BT, MATTHEWS BD, SING RF, et al. Initial re-sults with an electrothermal bipolar vessel sealer[J]. Surg Endosc,2001, 15(8): 799-801.
  • 2LANDMAN J, KERBL K, REHMAN J, et al. Evaluation of avessel sealing system, bipolar electrosurgery, harmonic scalpel, ti-tanium clips, endoscopic gastrointestinal anastomosis vascular sta-ples and sutures for arte rial and venous ligation in a porcinemodel[J]. J Urol, 2003,169(2): 697-700.
  • 3HAROLD KL, POLLINGER H, MATTHEWS BD, et al. Compar-ison of ultrasonic energy, bipolar thermal energy, and vascularclips for the hemostasis of small-, medium-, and large-sized ar-teries[J]. Surg Endosc, 2003, 17(8): 1228-1230.
  • 4BATEMAN BT, MHYRE JM, CALLAGHAN WM, et al. Peripar-tum hysterectomy in the United States: nationwide 14 year expe-rience[J]. Am J Obstet Gynecol, 2012, 206(1): 63.el-8.
  • 5WORLEY MJ, SLOMOVITZ BM, RAMIREZ PT, et al. Compli-cations of laparoscopy in benign and oncologic gynecologicalsurgeij[J]. Rev Obstet Gynecol, 2009, 2(3): 169-175.
  • 6冷金花,郎景和,李志刚,刘珠凤,孙大为,朱兰,黄荣丽.腹腔镜手术并发症及其相关因素分析[J].现代妇产科进展,2002,11(6):430-433. 被引量:157
  • 7CHRISTIAN W, RAJAB TK, KRAMER B, et al. Quantifyingelectrosurgery-induced thermal effects and damage to human tis-sue: An exploratory study with the fallopian tube as a novelin-vivo in-situ model[J]. J Minim Invasive Gynecol, 2010,17(1):70-77.
  • 8TULKANGAS PK, SMITH Y, FALCONE T, et al. Gross andhistologic characteristics of laparoscopic injuries with four differ-ent energy sources[J]. Fertril Steril, 2001, 75(4): 806-810.
  • 9周影,马玲,代涛,朱掌珠.健康人子宫动脉血流动力学检测[J].解剖与临床,2002,7(4):158-159. 被引量:3
  • 10夏恩兰,李自新.妇科内镜学[M].北京:人民卫生出版社,2004:436-441.

二级参考文献2

共引文献158

同被引文献9

引证文献1

二级引证文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史