期刊文献+

血清淀粉酶测定结果的溯源与方法学分组的关系

Relation between traceability of serum Amylase measuring results and methodological group
下载PDF
导出
摘要 从一项血清酶活性可溯源性研究中发现淀粉酶(Amylase,AMY)测定结果与校正物的定值相差悬殊,并且不同来源的试剂盒间也存在高低不一的系统误差这一现象出发,分析了出现这些系统误差的原因,从理论和实验两个方面指出影响AMY测定结果的关键因素是AMY作用底物的结构不同,以此为例进而说明血清酶活性结果的溯源必须考虑构成检测系统的各个要素,其中首先是方法学原理及试剂盒的性能质量。 It was found from a traceability study of serum enzyme activity that the Amylase (Amylase,AMY) measurement results differ greatly on the value determination of calibrator,and there are also either more or less systemic errors between kits from different sources.The causes of the systemic error are analyzed.And from two aspects of theory and experiment it was proved that the key factors for affecting AMY measurement results are structural difference of AMY substrate.It is a good example to demonstrate that primary factors of a detection system,first of all, the methodological principles and kit performance and quality must be considered for serum amylase measuring results traceability.
作者 王卫兵
出处 《实验技术与管理》 CAS 北大核心 2015年第11期66-68,共3页 Experimental Technology and Management
关键词 血清淀粉酶 溯源 检测系统 方法分组 serum amylase traceability detection system method group
  • 相关文献

参考文献13

  • 1王清涛,童清,郭健.血清酶测定结果可溯源性研究[J].中华检验医学杂志,2003,26(12):761-763. 被引量:23
  • 2北京市医疗器械检验所.GB/T21415--2008体外诊断医疗器械生物样品中量的测量校准品和控制物质赋值的计量学溯源性[s].北京:中国标准出版社,2008.
  • 3韩志军,黄志锋,卢业成,等.临床化学常用项目自动化分析法[M].3版.沈阳:辽宁科学技术出版社,2005:826-835.
  • 4Gerhard Schumann, Ryozi Aoki, Carlo A Ferrero. IFCC primary reference procedures for the measurement ofcatalytic activity con- centrations of enzymes at 37~C [-J~. Clin Chem Lab Med ,2006:44 (9).
  • 5Katsuhiko Kuwa. Scientiifc Division of Japan Society of Clinical Chemistry~ Proposed standard for certified enzyme reference mate- rials(ERM) [-J~. Japan J Chem, 2000,29 ~ 77-92.
  • 6Kruse/Jarres J D. C. Kaiser Evalution of a new a Amylase Assay Using 4, 5-Ethylidene-( GT)-l-4-nitrophenyl-( G1 )-a-D-maltohep taosideEJ~. Clin. Biochem, 1989,27 : 103-113.
  • 7张连祥,刘晨梅,蔡忠,王兆成.七种试剂盒检测α-淀粉酶结果的比较及分析[J].天津医药,2002,30(9):554-555. 被引量:1
  • 8张秀明,庄俊华,徐宁,黄宪章,陈茶,尹一兵,梁伟雄.不同检测系统血清酶测定结果的偏倚评估与可比性研究[J].中华检验医学杂志,2006,29(4):346-349. 被引量:137
  • 9Georges F, John E,Takashi K. Validation of an Enzyme Calibrator- An IFCC Guideline~P II S0009-9120 (98) 00039--33 I-J]. Clin Bio- chem, 1998,31(6) :495-500.
  • 10文庆成.临床酶测定标准化的几个问题[J].中华医学检验杂志,1998,21(1):60-61. 被引量:29

二级参考文献38

  • 1陈文祥.临床检验量值溯源与参考系统[J].中华检验医学杂志,2006,29(1):17-19. 被引量:41
  • 2陈映,Tietz,NW.α—淀粉酶测定的评估[J].临床检验信息导报,1996,3(1):29-31. 被引量:2
  • 3李清华.淀粉酶测定:难以标准化的测定方法[J].临床检验信息导报,1996,3(2):40-43. 被引量:1
  • 4.中华人民共和国国家标准.检测和校准实验室能力的通用要求[M].北京:中国标准出版社,2001.1-18.
  • 5孟泽,临床检验信息导报,1997年,1期,5页
  • 6许绍辉,临床检验信息导报,1997年,1期,10页
  • 7王毓三,临床检验信息导报,1996年,1期,10页
  • 8许绍辉,临床检验信息导报,1996年,1期,17页
  • 9杨振华,临床检验信息导报,1996年,1期,4页
  • 10杨振华,中华医学检验杂志,1994年,2期,121页

共引文献217