FLASH扫描在冠状动脉CTA双低中的应用研究 被引量:1

The Application of FLASH Scanning Mode with Double-low Dose Technique in Coronary CT Angiograph
下载PDF
导出
摘要 目的:探讨在100k V管电压条件下应用碘克沙醇低碘浓度(270 mg I/m L)对比剂,结合大螺距前瞻性心电门控螺旋扫描(FLASH spiral)模式和正弦波迭代重建技术(SAFIRE)行冠状动脉CT成像(CCTA)的可行性。方法:收集接受CCTA检查且心率≤70次/min的患者60例,随机分为A、B两组。A组患者采用FLASH spiral扫描模式,扫描管电压100 k V,碘对比剂浓度为270 mg I/m L,应用迭代重建技术(SAFIRE),重建卷积核I26f;B组患者采用前瞻性心电门控序列扫描模式,扫描管电压为120 k V,碘对比剂浓度为370 mg I/m L,采用滤波反投影重建(FBP),重建卷积核B26f,比较两组患者CCTA图像质量与有效射线剂量的差异性。结果:1两组患者年龄、心率和BMI比较,差异无统计学意义(P>0.05)。2图像质量:两组图像质量主观评价比较,差异无统计学意义(P>0.05)。两组图像质量客观评价比较,A、B两组在左冠状动脉主干的SD和SNR数值差异有统计学意义外(P<0.05),其他指标差异均无统计学意义(P>0.05)。3辐射剂量:A组平均有效射线剂量为(0.94±0.05)m Sv,B组(4.31±1.21)m Sv,两组的差异有统计学意义(P<0.05)。A组有效剂量较B组大幅减低,降幅约达78.2%。结论:100k V管电压条件下应用碘克沙醇低碘浓度(270 mg I/m L)对比剂,结合FLASH扫描模式和SAFIRE行CCTA检查,图像质量可以满足临床诊断需要,与一般前瞻性心电门控序列扫描相比,有效剂量降低约78.2%。 Objective: To evaluate the potential of acquiring the diagnostic images of coronary CT angiography with prospective ECG-triggered high pitch flash spiral acquisition in 100 kV tube voltage combined with iterative reconstruction and contrast of iodixanol (270 mgI/mL). Methods: The 60 patients were divided into A and B two groups randomly, whose heart rate 〈 70/min. In the A group, their coronary CT angiography images were collected with prospective ECG-triggered high pitch flash spiral mode in 100 kV tube voltage combined with iterative reconstruction (convolution kernel B26f) and contrast of iodixanol (270 mgI/mL). In the B group, their coronary CT angiography images were collected with prospective ECG-triggered sequence mode in 120 kV tube voltage combined with filtered back projection (convolution kernel B26f) and contrast of iopromide (370 mgI/mL). The image quality and effective radiation dose of two groups were assessed. Results: (1) There were no significant differences in age, heart rate and BMI between the two groups. (2) Image qualities: There were no significant differences in the subjective image quality evaluation between the two groups.In objective image quality assessment, in addition to the left coronary artery of SD and SNR numerical difference was statistically significant, no statistically significant difference were observed in the other indicators. (3) Radiation dose: The average effective radiation dose was (0.94± 0.05)mSv in group A and (4.31± 1.21) mSv in group B respectively. The differences between the two groups have statistical significance. In effective dose group A decreased sharply than group B by about 78.2%. Conclusions: In coronary CT angiography with prospective ECG-triggered high pitch flash spiral acquisition in 100kV tube voltage combined with iterative reconstruction and contrast of iodixanol (270 mgI/mL) which image quality can meet the demand of clinical diagnosis, it compared with prospective ECG-triggered sequence mode in 120 kV tube voltage, could decrease the effective dose by about 78.2%.
出处 《CT理论与应用研究(中英文)》 2015年第5期637-645,共9页 Computerized Tomography Theory and Applications
基金 内蒙古自治区人民医院院内基金(201306)
关键词 冠状动脉CTA 体层摄影术 X线计算机 双源CT 迭代重建 coronary CT angiography tomography X-ray computed dual source CT iterative reconstruction
  • 相关文献

参考文献17

  • 1Lucaya J, Piqueras J, Garcia-Pena P. et al. Low dose high resolution CT of the chest in children and adults: Dose, cooperation, artifact incidence and image quality[J]. American Journal of Roentgeno]ogy, 2000, 175(4): 985-992.
  • 2Romano G, Briguori C, Quintavalle C, et al. Contrast agents and renal cell apoptosis[J]. European Heart Journal, 2008, 29(20): 2569-2576.
  • 3Zhao Y, Tao Z, Xu Z, et al. Toxic effects of a high dose of non-ionic iodinated contrast media on renal glomerular and aortic endothelial cells in aged rats in vivo[J]. Toxicol Lett, 2011, 202(3): 253-260.
  • 4Frush DP. Applegate K. Computed tomography and radiation: Zunderstanding the issues[J]. Journal of the American College of Radiolog, 2004, 1(2): 113-119.
  • 5Austen WG, Edwards JE, Frye RL, et al. A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association[J]. Circulation, 1975, 51(S4): 5-40.
  • 6Halliburton SS, Abbara S, Chen MY, et al. SCCT guidelines on radiation dose and dose-optimization strategies in cardiovascular CT[J]. J Cardiovasc Comput romogr, 2011, 5(4): 198-224.
  • 7Brenner DJ, Elliston CD. Estimated radiation risks potentially associated with full-body CT screening[J3. Radiology, 2004, 232(3): 735-738.
  • 8Tatsugami F, Matsuki M, Nakai G, et al. The effect of adaptive iterative dose reduction on image quality in320-detector row CT coronary angiography[J]. British Journal of Radiology, 2012, 85(i016): 378-382.
  • 9张飞飞,张晓琴,张凯.双管电压和低深度碘对比剂冠状动脉CTA检查的临床研究[J].内蒙古医学杂志,2014,46(4):388-390. 被引量:4
  • 10Wei-Hua YinMD, Bin Lu MD, Jian-Bo GaoMD, et al. Effect of reduced X-ray tube voltage, low iodine concentration contrast medium, and sinogram-affirmed iterative reconstruction on image quality and radiation dose at coronary CT angiography: Results of the prospective multicenter REALISE trial[J]. Journal of Cardiovascular Computed Tomography, 2015, (9): 215-224.

二级参考文献33

  • 1蒋学祥,邱建星,王继琛,孙晓伟,高莉,高福生,王仪生.64层螺旋CT冠状动脉成像质量的研究[J].中国医学影像技术,2006,22(10):1452-1455. 被引量:31
  • 2Husmann L,Valenta I,Gaemperli O,et al.Feasibility of lowdose coronary CT angiography:first experience with prospective ECG-gating Eur Heart J,2008,29:191-197.
  • 3Coles DR,Smail MA,Negus IS,et al.Comparison of radiation doses from multislice computed tomography coronary angiography and conventional diagnostic angiography.J Am Coll Cardiol,2006,47:1840-1845.
  • 4Brenner DJ,Hall EJ.Computed tomography:an increasing source of radiation exposure.N Engl J Med,2007,357:2277-2284.
  • 5Hausleiter J,Meyer T,Hermann F,et al.Estimated radiation dose associated with cardiac CT angiography.JAMA,2009,301:500-507.
  • 6Kalra MK,Maher MM,Toth TL,et al.Strategies for CT radiation dose optimization.Radiology,2004,230:619-628.
  • 7Mollet NR,Cademartiri F,de Feyter PJ.Non-invasive multislice CT coronary imaging.Heart,2005,91:401-407.
  • 8Szucs-Farkas Z,Kurmann L,Strautz T,et al.Patient exposure and image quality of low-dose pulmonary computed tomography angiography:comparison of 100-and 80-kVp protocols.Invest Radiol,2008,43:871-876.
  • 9Abada HT,Larehez C,Daoud B,et al.MDCT of the coronary arteries:feasibility of low-dose CT with ECG-pulsed tube current modulation to reduce radiation dose.AJR,2006,186:S387-S390.
  • 10Sehoenhagen P.Back to the future:coronary CT angiography using prospective ECG triggering.Eur Heart J,2008,29:153-154.

共引文献112

同被引文献7

引证文献1

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部