期刊文献+

人体皮肤在体手持式扫频光学相干层析系统 被引量:10

Handheld Swept Source Optical Coherence Tomography for Imaging Human Skin in vivo
原文传递
导出
摘要 搭建了一台手持式扫频光学相干层析(SSOCT)系统对人体皮肤进行在体成像。对该系统的性能进行了详细的分析和讨论,并与实验室内现有的传统结构谱域光学相干层析(SDOCT)系统进行了对比,其中手持式SSOCT系统的灵敏度为101 d B,SDOCT系统的灵敏度为94 d B。通过一次样本模拟实验和两次人体不同位置皮肤成像的结果证明了手持式SSOCT系统的信噪比和成像深度都优于SDOCT系统。利用手持式SSOCT系统对人体胳膊内侧成像,体现了手持式SSOCT系统能够对任意位置皮肤成像的优越性。所有实验结果证明搭建的手持式SSOCT系统除了具有使用方便的优点外,还能够提供高质量的人体皮肤微观结构图像。 A handheld swept source optical coherence tomography (SSOCT) system is described for imaging human skin in vivo. The performance of the system is analyzed in detail, and compared to the spectral domain optical coherence tomography (SDOCT) system in our laboratory. The results show that the sensitivity of the handheld SSOCT system and the traditional SDOCT system is 101 dB and 94 dB, respectively. A phantom experiment and two in vivo experiments on different human skin areas are carried out to demonstrate the better performance of the handheld SSOCT system in terms of signal to noise ratio (SNR) and the maximum imaging depth. The inner side of a human arm is imaged successfully with the handheld SSOCT system, which suggests that the handheld SSOCT system can image any position of human skin. All the results indicate that the handheld SSOCT system can not only be used to image skin anywhere conveniently, but also offer the structure images with high quality.
出处 《光学学报》 EI CAS CSCD 北大核心 2015年第11期234-242,共9页 Acta Optica Sinica
基金 国家自然科学基金(61275198 60978069) 南京理工大学毕业设计重点课题
关键词 医用光学 扫频光学相干层析术 手持式 最大成像深度 信噪比 灵敏度 medical optics swept source optical coherence tomography handheld maximum imaging depth signal to noise ratio sensitivity
  • 相关文献

参考文献9

二级参考文献108

  • 1Pilger FRANK.Poly(ethylene glycol) conjugated nano-graphene oxide for photodynamic therapy[J].Science China Chemistry,2010,53(11):2265-2271. 被引量:8
  • 2LI ZhaoBo1,2,WANG JianGuang1,2,CHEN JingRong1,LEI WanHua1,WANG XueSong1 & ZHANG BaoWen1 1Key Laboratory of Photochemical Conversion and Optoelectronic Materials,Technical Institute of Physics and Chemistry,Chinese Academy of Sciences,Beijing 100190,China 2Graduate University of Chinese Academy of Sciences,Beijing 100049,China.Hypocrellin B doped and pH-responsive silica nanoparticles for photodynamic therapy[J].Science China Chemistry,2010,53(9):1994-1999. 被引量:2
  • 3俞晓峰,丁志华,陈宇恒,黄丽娜,周琳,周振明,吴兰,刘旭.光纤型光学相干层析成像系统的研制[J].光学学报,2006,26(2):235-238. 被引量:34
  • 4王颖,顾瑛,刘凡光,程刚,黄乃艳.HMME在不同溶液中光漂白反应动力学规律[J].中国激光医学杂志,2007,16(2):73-78. 被引量:2
  • 5D. Huang,E. A. Swanson, C. P. Linet al.. Optical coherence tomography[J]. Science, 1991, 254(5035) : 1178-- 1181.
  • 6S. Yun, G. Tearney, J. de Boer et al.. High speed optical frequency-domain imaging [J]. Opt. E:cpress, 2003, 11 (22) : 2953-2963.
  • 7F. Fercher, C. K. Hitzenberger, G. Kamp et al.. Measurement of intraocular distances by backscattering spectral interferometry[J]. Opt. Commun. , 1995,117:443-448.
  • 8G. Hausler , M. W. Lindner. Coherence radar and spectral radar new tools for dermatological diagnosis[J]. J. Biomed. Opt. , 1998,3:21-31.
  • 9S. R. Chinn,E. A. Swanson, J.G. Fujimoto. Optical coherence tomography using a frequency tunable optical source[J]. Opt. Lett., 1997,22(5):340-342.
  • 10R. Leitgeh, C. Hitzenberger, A. Fercher. Performance of fourier domain vs. time domain optical coherence tomography [J]. Opt. Express, 2003,11(8): 889-894.

共引文献94

同被引文献64

引证文献10

二级引证文献47

相关作者