期刊文献+

电子血压计质量检测方法与系统的研究 被引量:6

Study on testing method and system for electronic sphygmomanometer
下载PDF
导出
摘要 针对传统的血压计质量检查方法存在功能单一、技术落后、自动化程度低、测量精度和灵活性差等问题,创造性的提出了血压计质量检查的相关指标和方法,并基于该理论,设计了一种基于PC全自动、多功能、通用型血压计质量检测系统。实验结果表明,这些质量检查指标,可以全面彻底的判断血压计的质量状态,为企业和医疗机构提供可靠的血压计产品质量分析方案。该系统通过简单的参数设置,能够自动完成各类血压计产品的漏压及工作特性曲线检测,具有较好的推广应用价值。 The traditional quality inspection methods for sphygmomanometer have disadvantages of single function,obsolete technology,low degree of automation and poor precision etc.This paper creatively puts forward the relevant indicators and methods of the sphygmomanometer quality inspection,and based on the theory,an automatic,multifunctional and universal quality inspection system for sphygmomanometer is proposed in the paper.The experimental results show that,these quality inspection indicators can judge the quality state of the blood-pressure meter thoroughly,and the system has functions of gas leak detection,pressure characteristic curve test,power test etc.The system can check the comprehensive characteristics for all kinds of sphygmomanometer products automatically by programming and parameter adjustment.
出处 《国外电子测量技术》 2015年第11期44-47,共4页 Foreign Electronic Measurement Technology
基金 中央高校基本科研业务费(3142013060 3142013099)资助项目
关键词 多功能 通用型 电子血压计 质量检测 multifunctional universal electronic sphygmomanometer quality inspection
  • 相关文献

参考文献10

二级参考文献58

共引文献83

同被引文献44

  • 1YAN M Y, HOU ZH SH, GAO Y. A bilateral brain symmetry index for analysis of EEG signal in stroke pa- tients[ J]. The 4th International Conference on Bio- medical Engineering and Informatics (BMEI), 2011(1):8-11.
  • 2LOPEZ-GORDO M A, SANCHEZ-MORILLO D, VALLE F P. Dry EEG electrodes [ J]. Sensors, 2014, 14 (7) : 12847.
  • 3GARGIULO G, CALVO R A, BIFULCO P, et al. A new EEG recording system for passive dry electrodes [ J ]. Clinical Neurophysiology, 2010, 121 (5) : 686-693.
  • 4CHIY M, WANGYT, WANGY J, et al. Dry and noncontact EEG sensors for mobile brain-computer in- terfaces [ J]. IEEE Transactions on Neural Systems and Rehabilitation Engineering, 2012, 20 ( 2 ) : 228-235.
  • 5VAN PUTI'EN M J A M, PETERS J M, MULDER S M, et al. A brain symmetry index (BSI) for online EEG mo- nitoring in carotid endarterectomy [ J ]. Clinical Neuro- physiology, 2004, 115(5): 1189-1194.
  • 6MENG Y, LI ZH B, CHEN X, et al. A flexible dry mi- cro-dome electrode for ECG monitoring [ J ]. Microsyst Technol, 2015, 21(6) : 1241-1248.
  • 7SITNIKOVA E, HRAMOV A E, KORONOVSKY A A, et al. Sleep spindles and spike - wave discharges in EEG: Their generic features, similarities and distinctions disclosed with Fourier transform and continuous wavelet analysis [ J ]. Journal of Neuroscience Methods, 2009, 180(2) : 304-316.
  • 8LIN CH T, LIAO L D, LIU Y H, et al. Novel dry poly- mer foam electrodes for long-term EEG measurement [ J ]. IEEE Transactions on Bio-medical Engineering, 2011,58 (5) : 1200-1207.
  • 9DIAS N S, CARMO J P, MENDES P M, et al. Wireless instrumentation system based on dry electrodes for acqui- ring EEG signals [ J ]. Medical Engineering & Physics, 2012, 34(7): 972-981.
  • 10王晓韡,石立臣,吕宝粮.干电极脑电采集技术综述[J].中国生物医学工程学报,2010,29(5):777-784. 被引量:23