摘要
本研究设计一个程序软件,用于计算血液透析中温度对水处理出水量的影响。通过数据采集与记录,使用统计回归方法得出水处理系统反渗透膜温度与出水量的几种函数关系曲线,利用C#高级程序编译语言,编写不同温度下水处理出水量的计算程序软件。该软件为医院采购水处理设备与评估反渗膜的更换时间提供必要的数据支持。
This study focused on development of a program for calculation of the impact of temperature on water treatment discharge volume. Through collection and record of data, the statistical regression method was utilized to obtain the correlation function curve between the reverseosmosis membrane temperature and water discharge volume of the water treatment system. Then, the program was compiled for calculation of the water treatment discharge volume under different temperature by using C# advanced programming language. The program could provide the necessary basis for the hospital to procure water treatment equipment and evaluate the replacement time of reverse-osmosis membrane.
出处
《中国医疗设备》
2015年第12期45-48,共4页
China Medical Devices
关键词
血液透析机
反渗膜
出水量
温度
hemodialysis system
reverse-osmosis membrane
water discharge volume
temperature