激光热源止血刀闭合血管的可行性研究 被引量:2

The feasibility study of vascular occlusion using laser heat source scalpel
下载PDF
导出
摘要 目的研发一种具有克服生物组织对激光选择性吸收的激光热源止血刀,通过血管封合在体实验,探讨其闭合猪动脉血管的可行性。方法设计一个激光热源刀头,采用石墨作为激光吸收介质用于激光止血刀刀头,该吸收介质将1470nm激光能量转化为热能。用该刀头闭合小型猪不同部位不同直径的离体动脉血管,闭合后测量血管的爆破压力作为血管闭合强度的指标。术后,将未进行爆破压实验的血管制成切片,观察闭合处血管的对合情况。结果激光热源止血刀成功闭合了4mm以下的猪离体血管,小血管(1~2mm)、中血管(3~4mm)、大血管(5~6mm)的平均爆破压分别为(238.83±26.68)mmHg、(358.83±29.39)mmHg、(60.40±0.17)mmHg。显微镜下观察闭合血管的切片,闭合处血管对合好。结论激光热源止血刀封合小血管和中血管有较高的闭合强度和可靠性。 Objective In order to overcome the selective absorption of tissues to laser,we developed a laser heat source scalpel, and then used it to occlude porcine arteries to investigate its reliability. Methods We designed a laser heat head which use graphite as the absorbing materials to convert 1470 nm laser energy into thermal energy. Then we used this scalpel to occlude different sizes of arteries which are harvested from different parts of a porcine. After occlusion, we measured the burst pressure of each occluded vessel as the index of occluded strength. Finally, we cut the occluded vessel samples which had not burst into pathological slices to examine whether they were closed. Results All vessels whose diameters were under 4 mm were successfully occluded. The burst pressures for small ( 1-2 mm) , medium ( 3-4 mm) , and large ( 5- 6 mm) vessel diameters were (238.83±26.68) mmHg, (358.83±29.39) mmHg,and (60.40±0. 17) mmHg,respectively. The vessel walls clung to each other and there was no gap between them as observed under optical microscope. ConclusionsUse laser heat source scalpel to occlude small and medium diameters vessels have high occlude strength and reliability.
出处 《北京生物医学工程》 2015年第6期573-577,共5页 Beijing Biomedical Engineering
基金 国家自然科学基金(61272255) 北京市教委科技项目(KM201310025009)资助
关键词 激光 血管 闭合血管 止血刀 laser vessel occluded vessel scalpel
  • 相关文献

参考文献18

  • 1康小芳,仲伟国.腹腔镜手术中不同止血方法优缺点的比较[J].腹腔镜外科杂志,2008,13(1):78-81. 被引量:9
  • 2Lee SJ,Park KH. Ultrasonic energy in endoscopic surgery [ J ].Yonsei Medical Journal, 1999,40(6) :545-549.
  • 3Hruby GW, Marruffo FC, Durak E, et al. Evaluation of surgicalenergy devices for vessel sealing and peripheral energy spread in aporcine model[ J]. The Journal of Urology,20071178( 6) :2689-2693.
  • 4Lamberton GR, Hsi RS, Jin DH, et ai. Prospective comparison offour laparoscopic vessel ligation devices [ J ]. Journal ofEndourology.2008,22( 10) :2307-2312.
  • 5Sutton PA, Awad S, Perkins AC , et al. Comparison of lateralthermal spread using monopolar and bipolar diathermy,theHarmonic Scalpel? and the Ligasure? [ J ]. British Journal ofSurgery,2010,97(3) :428-433.
  • 6吕磊,张正厚,隋丽云,李田勋,郭顺生.激光在医学基础和临床研究中的应用[J].中国临床康复,2006,10(17):152-154. 被引量:6
  • 7Niemz MH. Laser-tissue interactions : fundamentals andapplications [ M ]. New York : Springer, 2007.
  • 8Ritz JP,Roggan A,Isbert C,et al. Optical properties of native andcoagulated porcine liver tissue between 400 and 2400 nm [ J ].Lasers in surgery and medicine,2001,29( 3) ;205-212.
  • 9徐国昌.激光加工中金属表面涂敷的研究[J].激光与红外,1997,27(1):27-29. 被引量:4
  • 10Newcomb WL, Hope WW, Schmelzer TM,et al. Comparison ofblood vessel sealing among new electrosurgical and ultrasonicdevices[ J]. Surgical Endoscopy,2009,23( 1) :90-96.

二级参考文献59

共引文献93

同被引文献16

引证文献2

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部