以LDH为例来探讨具有溯源性和互换性的校准品在临床生化应用中的重要性 被引量:3

With Lactate Dehydrogenase as An Example to Evaluate the Importance of the Traceability and Commutability Calibration in Clinical Biochemical Detection
下载PDF
导出
摘要 目的通过在临床实验室使用具有溯源性和互换性的校准品,寻求血清酶测定标准化的新途径。方法以乳酸脱氢酶(LDH)为例,收集混合血清,由4家酶学参考实验室联合赋值,制备具有溯源性和互换性的参考物质。参加试验的8个实验室先使用自己的校准品,在校准通过后,将高中低3个水平的血清样本和参考物质各测定3次,记录检验结果;再使用具有溯源性和互换性的参考物质作为校准品校准后,重复测定高中低3个水平的血清样本和参考物质各3次,记录检验结果。计算8家实验室在使用具有溯源性和互换性的参考物质作为校准品校准前后,参考物质检测结果与靶值的偏倚和高中低3个水平的血清样本LDH检测结果的变异系数(CV%)。结果使用具有溯源性和互换性的参考物质参后,参考物质检测结果与靶值的偏侍由6.80%降到1.40%;高中低3个水平的血清样本LDH检测结果的变异系数CV由11.41%,9.93%和9.87%降到了4.13%,3.55%和3.80%。结论在临床实验室使用具有溯源性和互换性的参考物质作校准品,能大大降低实验室间结果的正确度和精密度,提高检测结果的准确性,有利于各个实验室各设备间结果的标准化和一致性。 Objective With the use of traceability and commutability calibration in clinical laboratory,try to seek new way in the standardized of serum enzyme determination. Methods For example to the lactate dehydrogenase (LDH), to collect mixed serum, which assigned by four enzymology reference laboratory, produced the traceability and commutability reference materials. Firstly, eight laboratories used their own calibration to calibrate their detection systems. After calibration, every la- boratory detected one enzyme calibrator and three patient samples and then recorded the results. Additionally reused the traceability and commutability reference materials to calibrate their detection systems. After calibration,every laboratory detected them again and recorded the results. Calculated the calibration products for the determination of the bias between the results of the overall mean and labeled values, determination of serum samples from the changes of the results of precision before and after calibration. Results After the use of thetraceability and commutabilityreference materials,the bias between results of reference materials test and the target was value from 6. 80% decline to 1.40%. The precision of three levels of se- rum samples were from 11.41%, 9.93% and 9.93% to 4.13%, 3.55% and 3.80%. Conclusion Using the traceability and commutability reference materials as calibration in clinical laboratory, can greatly reduce the accuracy and precision of the results among these laboratory and improve the accuracy of test results, and then is increase the standardization and consistency in clinical laboratory.
出处 《现代检验医学杂志》 CAS 2015年第6期105-108,共4页 Journal of Modern Laboratory Medicine
关键词 乳酸脱氢酶 溯源性 互换性 校准品 lactate dehydrogenase traceability commutability calibrator
  • 相关文献

参考文献9

  • 1Miller WG, Myers GL . Commutability still matters [J].Clin Chem, 2013,59(9) : 1291-1293.
  • 2Zegers I, Beetham R, Keller T, et al. The importance of commutability of reference materials used as cali- brators: the examples of ceruloplasmin [J]. Clin Chem,2013,59(9) :3122-1329.
  • 3Schumann G, Bonora R, Ceriotti F, et al. IFCC prima- ry reference procedures for the measurement of cata- lytic activity concentrations of enzymes at 37"C. Part3. Reference procedure for the measurement of catalytic concentration of lactate dehydrogenase[J]. Clin Chem Lab Med,2002,40 (16) : 642-648.
  • 4张莹,周铁成,童开,郝晓柯.酶学参考实验室室间质量评价技术方案的建立[J].现代检验医学杂志,2010,25(3):146-148. 被引量:5
  • 5International Organization for Stand ardization(ISO). In vitro diagnostic medical devices-measurement of quantitatias in biological samples-metrologicaI tracea- bility of values assigned to calibrators and control materials[S]. ISO 17511, ISO, 2003.
  • 6Greg Miller W, Myers GL, Lou Gantzer M, et al. Roadmap for harmonization of clinical laboratory measurement procedures[J]. Clin Chem, 2011,57 (8) : 1108-1117.
  • 7Carobene A,Guerra E,Ceriotti F. A mechanism-based way to evaluate commutability of control materials for enzymatic measurements. The example of gamma- glutamyltransferase[J].Clinica Chimica Acta, 2013 (424) : 153-158.
  • 8Perieh C, Rie6s C, Alvarez Z, et al. External qualityassurance programs as a tool for verifying standardi- zation of measurement proeedures:Pilot collaboration in Europe [J]. Clinica Chimica Aeta, 2014 ( 432 ) : 82- 89.
  • 9李民,王延伟,陈阵,胡瑛.几种α-淀粉酶校准品和质控品的互换性研究[J].临床军医杂志,2012,40(4):942-943. 被引量:3

二级参考文献9

  • 1徐国宾,吴南,王清涛.要重视血清酶学测定的标准化工作[J].临床检验杂志,2007,25(3):161-164. 被引量:25
  • 2Schumann G ,Bonora R ,Ceriotti F,et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37℃. Part3. Reference procedure for the measurement of catalytic concentration of lactate dehydrogenase[J]. Clin Chem Lab Med, 2002,40(6) : 643-648.
  • 3Schumann G, Bonora R ,Ceriotti F,et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37℃. Part6. Reference procedure for the measurement of catalytic concentration of γ-glutamyltransferase[J]. Clin chem Lab Med ,2002,40( 7 ) : 734-738.
  • 4Vesper HW, Miller WG, Myers GL. Reference materials and commutability[ J]. Clin Biochem Rev, 2007, 28 (4) : 139 - 147.
  • 5CLSI document EP14-A2-Evaluation of matrix effects, approved guideline[J]. 2nd ed. Wayne, PA USA: CLSI, 2005.
  • 6Schumannl G, Aoki R, Ferrero CA, et al. IFCC primary refer- ence procedures for the measurement of catalytic activity concen- trations of enzymes at 37(;. Part 8. Reference procedure for the measurement of catalytic concentration of a-amylase [ J ]. Clin Chem Lab Med.2006.44(9) :1146 -1155.
  • 7Lorentz K. Routine a-Amylase assay using protected 4-Nitrophe- nyl-1,4-a-D-maltoheptaoside and a novel a-Glucosidase[ J]. Cli- ni Chem, 2000,46 (5) :644 - 649.
  • 8汪静,张传宝,居漪,陈宝荣,张曼,徐国宾,但万春,周跃国,宋耀虹,申子瑜,郭健,陈文祥,杨振华.国际临床化学联合会酶学测定参考方法实验室网络的初步建立[J].中华检验医学杂志,2008,31(3):258-263. 被引量:23
  • 9王涛,齐丽丽,刘海燕,田真,何振坤,李淑葵,但万春,徐国宾.六种临床诊断酶学测定参考方法的应用及国际比对结果分析[J].中华检验医学杂志,2008,31(3):264-269. 被引量:15

共引文献5

同被引文献19

引证文献3

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部