SysmexXN-3000全自动血液分析仪性能评价 被引量:2

Performance evaluation of Sysmex XN- 3000 automated hematology analyzer
原文传递
导出
摘要 目的对Sysmex XN-3000全自动血液分析仪性能进行评价。方法依照行业标准对Sysmex XN-3000的本底计数、携带污染率、准确度、精密度、线性范围、可比性等指标进行评价,并以显微镜法为"金标准",评价其白细胞分类的准确性。结果 Sysmex XN-3000对全血细胞各项参数的本底计数均为0,携带污染率≤0.38%;除极低值水平PLT-I/O外,各种浓度水平患者全血样品或室内质控物的批内精密度、日间精密度均符合卫生行业标准,各参数线性范围相关系数均≥0.975,a≤1±0.05,符合CLSI文件和卫生行业标准要求。与Sysmex XE-1800相比,WBC、Ret、Hb、PLT参数线性相关系数≥0.975,a≤1±0.05;白细胞分类方面,与人工显微分类相比,中性粒细胞、淋巴细胞、单核细胞、嗜酸性粒细胞、嗜碱性粒细胞的相关系数分别为0.972、0.991、0.654、0.381、0.518,P<0.05。结论 Sysmex XN-3000具有携带污染率低、精密度高、线性范围宽的特点,并且同Sysmex XE-1800及人工显微镜分类计数具有很好的一致性,可用于临床大批量标本常规分析。 Objective To evaluate the performance of Sysmex XN- 3000 automated hematology analyzer. Methods According to the industry standard,the background count,carryover rate,accuracy,precision,linear range,comparability of Sysmex XN-3000 were evaluated,and with artificial microscopy as the gold standard,its classification accuracy on leukocyte was evaluated.Results The background counts for complete blood parameters were all 0 with Sysmex XN- 3000,the carry- over contamination rate was ≤0. 38%. In addition to the extreme low levels of PLT- I / O,various levels of patients' whole blood samples or indoor quality control in each batch and the daytime precision all met the requirements of industry hygiene standard; the linear correlation coefficient of the parameters were all ≥0. 975,a≤1 ± 0. 05,which met the requirements of the CLSI document and the industry hygiene standards. Compared with Sysmex XE- 1800,the correlation coefficient of WBC,Ret,Hb,PLT were ≥0. 975,a≤1 ± 0. 05. Compared with artificial microscopy leukocyte classification,the correlation coefficient of neutrophils,lymphocytes,monocytes,eosinophils and basophils were respectively 0. 972,0. 972,0. 654,0. 381,0. 518,P〈0. 05. Conclusion Sysmex XN- 3000 has the characteristics of low carry- over rate,high precision,wide linear range,and is consistent with Sysmex XE- 1800 and artificial microscope in leukocyte classification. It could be used for routine analysis of a large quantities of clinical specimens.
出处 《中国卫生检验杂志》 CAS 2015年第24期4277-4280,共4页 Chinese Journal of Health Laboratory Technology
关键词 血液分析仪 性能评价 全血细胞计数 Hematology analyzer Performance evaluation Complete blood count
  • 相关文献

参考文献12

  • 1Sea JY, Lee ST, Kim SH. Performance evaluation of the new he- matology analyzer Sysmex XN - series [ J ]. Int J Lab IIematol,2015, 37(2): 155 - 164.
  • 2Food and Drug Administration, HHS. Hematology and pathology devices; reclassification of the automated differential cell counter. Final mle[J]. Federal register, 2002, 67(9) : 1606.
  • 3Briggs C, Culp N, Davis B, et al. ICSH guidelines for the evalua- tion of blood cell analysers including those used for differential leu- cocyte and retieulocyte counting[J]. Int J Lab Hematol, 2014, 36 (6) : 613 -627.
  • 4中华人民共和国卫生部.WS/T347-2011:血细胞分析的校准指南[S].北京:中国标准出版社,2011.
  • 5中华人民共和国卫生部.WS/T406-2012临床血液学检验常规项目分析质量要求[S].北京:中国标准出版社,2012.
  • 6Carey RN, George H, Hartmann AE, et al. User demonstration of performance for precision and accuracy: approved guideline [ J ]. NCCLS, 2000, 20(1) : 1 -40.
  • 7陈晓玲,武锦彪,王文娟.Sysmex XE-2100血液分析仪临床应用评价[J].实验与检验医学,2014,32(1):26-29. 被引量:12
  • 8Swaim WR. Reference leukocyte differential count (proportional) and evaluation of instrument methods[ J]. Am J Clin Pathol, 1991, 95(3) : 381 -388.
  • 9Barnes PW, McFadden SL, Machin S J, et al. International consen- sus group for hematology. The international consensus group for he- matology review: suggested criteria for action following automated CBC and WBC differential analysis [ J ]. Lab Hematol, 2005, 11 (2) : 83 -90.
  • 10Sehoorl M, Sehoorl M, Oomes J, et al. New fluorescent method ( PLT - F) on Sysmex XN2000 hematology analyzer achieved higher accuracy in low platelet counting[J]. Am J Clin Pathol, 2013, 140 (4) : 495 - 499.

二级参考文献17

  • 1薛冰蓉,徐鹏,吴延,胡东,叶平.LIC提示信息不同检测方法的可信性探讨[J].临床检验杂志(电子版),2012,1(4):238-241. 被引量:1
  • 2彭黎明,丛玉隆.我国血细胞自动分析中存在的问题及对策[J].中华检验医学杂志,2005,28(9):885-887. 被引量:43
  • 3丛玉隆,岳家新,袁家颖.实用血细胞分析技术与临床[M].北京:人民军医出版社,2011:14.
  • 4WS/T347-2011.血液分析仪的校准指南[S].2011.
  • 5CLSI EP15-A.Clinicaland Laboratory Standards Institute.User Demonstration of Performance for Precision and Accuracy;Ap- proved Guideline[S].CLS1,2001.
  • 6International Council for Standardization in Haematology (ICSH). Protocol for evaluation of automated hematology analyzer[S],iCSH, Clin Lab Heamotal, 1984;6:69.
  • 7CLSI EP6-A.Clinicaland Laboratory Standards Institute.Evaluation of the Linearity of Quantitative Measurement Procedures : A Statis- tical Approach ;Approved Guideline[S].CLSI, 2003.
  • 8CLSI EP9-A2.Clinicaland Laboratory Standards Institute.Method Comparison and Bias Estimation Using Patient Samples;Approved Guideline-Second Edition[S].CLS1,2002.
  • 9CLSI H20-A.Clinicaland Laboratory Standards Institute.Reference Leukocyte Differential Count (Proportional) and Evaluation of In- strumental Methods;Approved Standard[S].CLSl, 1992.
  • 10Tsuda I,Hino M ,Takubo T,et al. First basic performance evalua- tion of the XE-2100 haematology analyzer [J].J Autom Methods Manag Chem, 1999,21 (4): 127-133.

共引文献19

同被引文献13

引证文献2

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部