摘要 归纳了EMC实验室CNAS认可评审依据的文件。指出了EMC实验室在CNAS评审时常出现如下几类不合格案例:检测报告中缺少清晰标识、被测设备的连接示意图和检测布置示意图;校准证书中的校准参数不满足标准要求;仪器设备修正因子在测试系统中未及时更新;实验室无法提供标准变更后对测试方法、场地、设备等的验证记录;实验室未能提供屏蔽室满足屏蔽效能要求的证据。针对问题提出建议。 The basis of CNAS accreditation for EMC laboratory are summarized. Some non-conformities in the EMC laboratory accreditation are described and advices are proposed in detail. For example,there is no clear identification,test arrangement and devices connection in final report; calibration parameters in calibration certificate doesn't meet requirement of standards; modifying factor of devices aren't updated in test system; verification record of test methods,sites and devices aren't provided after standards' update; laboratory failed to provide evidence of shielding effectiveness.
作者 靳冬 崔强
出处 《安全与电磁兼容》 2015年第6期48-49,共2页 Safety & EMC
关键词 电磁兼容实验室 CNAS 不合格 认可 EMC laboratory CNAS non-conformity accreditation