期刊文献+

大气预浓缩仪-GC/FPD测定环境空气中的痕量硫化物 被引量:18

Determination of Trace Sulfur Compounds in Environmental Air by GC / FPD Coupled the Air Preconcentrations
下载PDF
导出
摘要 采用大气预浓缩仪-GC/FPD建立了环境空气中7种含硫化合物的测定方法。结果表明,经惰性化改造后的大气预浓缩仪-GC/FPD测定系统的基线低,对硫化物无残留。7种化合物的线性回归方程的相关系数均在0.995以上,高、中、低3种浓度的空白加标回收率分别为98.6%-105.3%、99.1%-103.7%、85.7%-99.7%,除硫化氢的相对标准偏差为10.9%外,其余6种含硫化合物的相对标准偏差可控制在6.9%以内。当进样体积为400 m L时,方法的检出限在0.10-1.1 ng/m^3。CO2、H2O对实际样品的测定无干扰。因此,该方法能够满足目前监测工作的要求,更加适用于环境空气中痕量含硫化合物的测定。 GC / FPD coupled the Air Preconcentrations was applied to the determination of 7 kinds of trace sulfur compounds in environment air. It was found that the system blank of the Air Preconcentration-GC / FPD was clear,and no residuum after the analysis of sulfur compounds. Results showed that the correlation coefficient of 7 sulfur compounds were all above 0. 995,it has high accuracy and precision. In addition,the recovery of sulfur compounds were in the range of 98. 6%-105. 3%,99. 1%-103. 7%,85. 7%-99. 7%. The relative standard deviations of the methods lower than 6. 9% for 6 compounds,but for hydrogen sulfide,the relative standard deviation was 10. 9%. Otherwise,when inject volume was 400 m L,the limit detection of 7 sulfur compounds was 0. 10-1. 1 ng / m3. So the measurement system was sensitive for sulfur compounds and very suitable for determination of the trace sulfur compounds in environment air.
出处 《中国环境监测》 CAS CSCD 北大核心 2015年第6期103-108,共6页 Environmental Monitoring in China
关键词 含硫化合物 环境空气 大气预浓缩 气相色谱法 火焰光度检测器 trace sulfur compounds air preconcentrations environment air gas chromatography flame photometric detector
  • 相关文献

参考文献18

二级参考文献72

共引文献134

同被引文献166

引证文献18

二级引证文献43

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部