摘要
采用一步水热合成法制备了氮掺杂石墨烯,采用滴涂法制备了氮掺杂石墨烯/壳聚糖修饰电极(N-GN-CS/GCE),研究了在大量抗坏血酸存在下尿酸在此修饰电极上的伏安行为。结果表明,尿酸在此修饰电极上有一个明显的氧化峰,其峰电流约为裸电极的7倍;同时,氮掺杂石墨烯/壳聚糖复合修饰材料显著提高了尿酸对抗坏血酸的选择性,尿酸和抗坏血酸的峰电位差可达362 m V,可实现高浓度抗坏血酸存在下尿酸的选择性测定。优化了修饰材料的滴涂量、溶液pH值、扫描速度等实验参数,利用差分脉冲伏安法对尿酸进行测定,尿酸的脉冲峰电流与尿酸浓度分别在0.1~20μmol/L和20~400μmol/L浓度范围呈良好的线性关系,检出限为0.01μmol/L,相对标准偏差为3.3%(n=8)。将本方法用于实际尿样的测定,回收率为99.7%~103.4%。
Nitrogen-doped graphene was synthesized by one-step hydrothermal method. A nitrogen-doped graphene-chitosan modified electrode( N-GN-CS / GCE) was prepared and the electrochemical behavior of uric acid in the presence of a large amount of ascorbic acid was investigated at the modified electrode. The peak current of uric acid at the modified electrode was 7 times as that at unmodified electrode. The peak separation of uric acid and ascorbic acid was 362 m V and uric acid could be accurately determined in the presence of a large amount of ascorbic acid. Some experimental parameters such as dopping volume,p H and scan rate were optimized and a standard curve was also established using differential pulse voltammetry. The linear range between the peak current and the concentration of uric acid was 0. 1 ~ 20 μmol / L and 20- 400 μmol / L,and the detection limit was estimated to be 0. 01 μmol / L. The proposed method was applied to the analysis of uric acid in human urine sample with recoveries between 99. 7% and 103. 4%.
出处
《分析化学》
SCIE
EI
CAS
CSCD
北大核心
2015年第12期1859-1863,共5页
Chinese Journal of Analytical Chemistry
基金
南京大学生命分析国家重点实验室开放研究基金(No.SKLACLS1308)
江苏省大学生创新训练项目(No.201411460025Y)
南京晓庄学院化学一级重点学科项目资助
supported by the National Natural Science Foundation of China(Nos.11254004
31460035)
关键词
氮掺杂石墨烯
抗坏血酸
尿酸
Nitrogen dopped graphene
Ascorbic acid
Uric acid