植入型心律转复除颤器不适当电击的发生特点及优化管理 被引量:2

The occurrence characteristics and optimal management of inappropriate shocks in patients with implantable cardioverter defibrillator
下载PDF
导出
摘要 目的分析植入型心律转复除颤器(ICD)术后不适当电击的发生特点及处理方法。方法通过腔内心电图判定ICD不适当电击的发生原因,针对发生原因进行个体化治疗。结果在我科2011-2014年植入ICD的189例患者中,20例发生了不适当电击,发生比例约10.6%,心房颤动(AF)伴快速心室率、室上性心动过速(SVT)及过感知是主要原因。其中,70%(14/20)患者是由于AF伴快心室率进入了室颤的诊断区间,发生了不适当电击;15%(3/20)误将SVT识别为室速;10%(2/20)肌电干扰过感知是由于发生了电极导线故障;5%(1/20)T波过感知。17例调整药物治疗方案及ICD优化程控设置,2例植入起搏感知电极进行手术干预故障电极,T波过感知通过调整起搏电极的感知极性,均未再发生严重误电击事件。结论AF快心室率是引起不适当电击的主要原因,长期有效的药物治疗结合ICD鉴别诊断参数优化是主要的治疗方法,对于特殊病因需采用必要的手术治疗进行干预。 Objective To evaluate the occurrence characteristics and management of inappropriate shocks in imp]antable cardioverter defibrillator (ICD) patients. Methods The causes of inappropriate shocks were analyzed by reviewing the endoeardial electrocardio- grams, then the patients received the individualized treatment. Results Twenty patients experienced inappropriate shocks among 189 ICD patients (10.6% ,20/189). Atrial fibrillation (AF) with high ventricuiar rate,supraventficular tachycardia (SVT) and oversensing were the main causes for inappropriate shocks. AF with high ventieular rate were involved in the diagnostic interval of ventricular fibrillation (VF) in 70% of the patients,which led to the inappropriate shocks. Three patients( 15% ) underwent inappropriate shocks for misunderstanding SVTs to VTs, and two patients experienced the lead malfunction because of the eleetromyographical interference oversensing, and one patient had T-wave oversensing. Seventeen patients received adjustment of drug therapy and optimal program setting of ICD. Two patients underwent the interventional operations on malfunction leads by implanting an additional pacing/sensing leads. T-wave oversensing was handled by adjusting the sensing polarity of pacing lead, then no severe false shocks event reoccurred. Conclusion AF with high ventrieular rate is the main cause for inappropriate shocks. The long-term drug therapy combined with the optimal ICD differential diagnostic parameters is the main therapeutic method. For some special eases, the necessary interventional operations are required.
出处 《临床军医杂志》 CAS 2015年第10期991-993,共3页 Clinical Journal of Medical Officers
关键词 心血管病学 埋藏式心脏转复除颤器 不适当电击 Cardiology Implantable eardioverter defibrillator Inappropriate shocks
  • 相关文献

参考文献11

  • 1Levy WC,Lee KL,Hellkamp AS,et al.Maximizing survival benefit with primary prevention implantable cardioverter-defibrillator therapy in a heart failure population[J].Circulation,2009,120(10):835-842.
  • 2Moss AJ,Schuger C,Beck CA,et al.Reduction in inappropriate therapy and mortality through ICD programming[J].N Engl J Med,2012,367:2275-2283.
  • 3Gasparini M,Proclemer A,Klersy C,et al.Effect of long-detection interval vs standard-detection interval for implantable cardioverterdefibrillators on antitachycardia pacing andshock delivery:the ADVANCE III randomized clinical trial[J].JAMA,2013,309:1903-1911.
  • 4Ruwald AC,Sood N,Ruwald MH,et al.Frequency of inappropriate therapy in patients implanted with dual-versus single-chamber ICD devices in the ICD armof MADIT-CRT[J].J Cardiovasc Electrophysiol,2013,24:672-679.
  • 5Theuns DA,Rivero-Ayerza M,Boersma E,et al.Prevention of inappropriate therapy in implantable defibrillators:A meta-analysis of clinical trials comparingsingle-chamber and dual-chamber arrhythmia discrimination algorithms[J].Int J Cardiol,2008,125(3):352-357.
  • 6Wathen MS,De Groot PJ,Sweeney MO,et al.Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneousrapid ventricular tachycardia in patients with implantable cardioverter-defibrillators:Pacing Fast Ventricular Tachycardia Reduces Shock Therapies(Pain FREE RxⅡ)trial results[J].Circulation,2004,110(17):2591-2596.
  • 7Sweeney MO,Sherfesee L,De Groot PJ,et al.Differences in effects of electrical therapy type for ventricular arrhythmias on mortality in implantable cardioverter-defibrillator patients[J].Heart Rhythm,2010,7(3):353-360.
  • 8Daubert JP,Zareba W,Cannom DS,et al.Inappropriate implantable cardioverter-defibrillator shocks in MADIT II:frequency,mechanisms,predictors,and survival impact[J].J Am Coll Cardiol,2008,51(14):1357-1365.
  • 9van Rees JB,Borleffs CJ,de Bie MK,et al.Inappropriate implantable cardioverter-defibrillator shocks:incidence,predictors,and impact on mortality[J].J Am Coll Cardiol,2011,57(5):556-562.
  • 10Eckstein J,Koller M T,Zabel M,et al.Necessity for surgical revision of defibrillator leads implanted long-term:causes and management[J].Circulation,2008,117(21):2727-2733.

同被引文献11

  • 1Iori M,Giacopelli D,Quartieri F,et al.Implantable cardioverter defibrillator system with floating atrial sensing dipole:a singlecenter experience[J].Pacing Clin Electrophysiol,2014,37(10):1265-1273.
  • 2Toivonen L,Lommi J.Dependence of atrial sensing function on posture in a single-lead atrial triggered ventricular(VDD)pacemaker[J].Pacing Clin Electrophysiol,1996,19(3):309-313.
  • 3Sticherling C,Zabel M,Spencker S,et al.Comparison of a novel,single-lead atrial sensing system with a dual-chamber implantable cardioverter-defibrillator system in patients without antibradycardia pacing indications:results of a randomized study[J].Circ Arrhythm Electrophysiol,2011,4(1):56-63.
  • 4Prystowsky EN,Benson DW Jr,Fuster V,et al.Management of patients with atrial fibrillation.A Statement for Healthcare Professionals.From the Subcommittee on Electrocardiography and Electrophysiology,American Heart Association[J].Circulation,1996,93(6):1262-1277.
  • 5Hart RG,Halperin JL.Atrial fibrillation and thromboembolism:a decade of progress in stroke prevention[J].Ann Intern Med,1999,131(9):688-695.
  • 6The Antiarrhythmics versus Implantable Defibrillators(AVID)Investigators.A Comparison of Antiarrhythmic-Drug Therapy with Implantable Defibrillators in Patients Resuscitated from Near-Fatal Ventricular Arrhythmias[J].N Engl J Med,1997,337(22):1576-1583.
  • 7Kuck KH,Cappato R,Siebels J,et al.Randomized comparison of antiarrhythmic drug therapy with implantable defibrillators in patients resuscitated from cardiac arrest:the Cardiac Arrest Study Hamburg(CASH)[J].Circulation,2000,102(7):748-754.
  • 8Connolly SJ,Gent M,Roberts RS,et al.Canadian implantable defibrillator study(CIDS):a randomized trial of the implantable cardioverter defibrillator against amiodarone[J].Circulation,2000,101(11):1297-1302.
  • 9Daubert JP,Zareba W,Cannom DS,et al.Inappropriate implantable cardioverter-defibrillator shocks in MADIT II:frequency,mechanisms,predictors,and survival impact[J].J Am Coll Cardiol,2008,51(14):1357-1365.
  • 10于海波,王祖禄.电风暴诊断及治疗进展[J].中国实用内科杂志,2015,35(7):634-637. 被引量:8

引证文献2

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部