摘要
建立一种氢化物电感耦合等离子体发射光谱法测定纺织品中砷(Ⅲ)的方法,纺织品中的微量砷经酸性试液萃取、过滤后,经过直接氢化物发生器形成砷化氢,进入等离子发射光谱仪进样系统进行定量测定.工作曲线线性相关系数大于0.999 9,线性范围为0.003 76-0.24 mg/L.该方法对砷(Ⅲ)的检出限为0.0005mg/L.对0.1-1.0mg/kg范围内的4个添加水平的平均回收率在94.8%-99.0%之间,相对标准偏差(n=6)在2.8%-0.7%之间.与电感耦合等离子发射光谱法相比,该方法具有基体背景干扰小,准确性高,检出限低等特点.
A method has been developed for the determination of As(Ⅲ) in textile by inductively coupled plasma atomic emission spectrometry(ICP-AES) via As(Ⅲ) hydride generation.The As(Ⅲ) in textile wasfirst extracted with an acidic solution,which was passed a hydride generator to convert the As(Ⅲ) into As(Ⅲ) hydride.The amount of As(Ⅲ) in the hydridewas determined by ICP-AES.In this study,the conditions for hydride generation were optimized.Under the optimal hydride generation conditions,the linear correlation coefficient in the range of 0.003 76 - 0.24 mg/LAs(Ⅲ) was greater than 0.999 9 with an minimum detection limit of As(Ⅲ) of 0.000 5 mg/L.The mean recoveries of As(Ⅲ) at four levels(0.1,0.2,0.5,1.0 mg/kg) were 94.8%- 99.0%with relative standard deviations(n =6) between 2.8%-0.7%.As compared to ICP-AES method,the method presented in this study can determine as low as 0.000 5 mg/LAs(Ⅲ) with much higher accuracy due to the smaller interference of matrix background.
出处
《北京服装学院学报(自然科学版)》
CAS
北大核心
2015年第4期56-62,共7页
Journal of Beijing Institute of Fashion Technology:Natural Science Edition
基金
北京市服装材料研究开发与评价重点实验室项目(项目编号:2012ZK-03)