基于两种计算方法对15MV医用直线加速器机房进行辐射屏蔽分析 被引量:2

Radiation shielding analysis for 15 MV medical linear accelerator room by using two calculations
下载PDF
导出
摘要 为医用直线加速器的屏蔽防护最优化设计提供参考依据,本文分别用9分值层法和蒙特卡罗方法对已知屏蔽墙外的辐射剂量率进行计算,并将计算结果与实际测量值进行对比分析,发现两种计算方法所测得的辐射剂量率值与测量值相对误差小于6%,由此得知两种计算方法均可以指导加速器机房辐射屏蔽设计,其中,蒙特卡罗方法误差相对较小,因此,在屏蔽机房设计时推荐使用蒙特卡罗方法。 To provide a reference for the optimal design of the shielding protection of medical linear accelerator, 9 points layer method and Monte Carlo method were respectively applied to calculate the radiation dose rate of the known shield wall. The calculated results and actual measurements were analyzed and compared. The comparative result showed the relative error between the result calculated by two methods and the measured result was less than 6%. The two calculations could guide the radiation shielding design of accelerator, and the Monte Carlo method is more favorable in the shielding room design for the smaller relative error.
作者 胡晓平 洪君
机构地区 解放军
出处 《中国医学物理学杂志》 CSCD 2016年第2期213-216,共4页 Chinese Journal of Medical Physics
关键词 9分值层法 蒙特卡罗方法 医用直线加速器 辐射防护 9 points layer method Monte Carlo method medical linear accelerator radiation protection
  • 相关文献

参考文献12

  • 1JOHN B, DAN L, CLEBER D, et al. Helical tomotherapy radiation leakage and shielding considerations[J]. Med Phys, 2005, 32(3): 710-719.
  • 2JOHN B, T1M H, RICHARD V, et al. Helical tomotherapy dynamic quality assurance[J]. Med Phys, 2006, 33(10): 3939-3950.
  • 3ROBINSON D, SCRIMGER J W, FIELD G C, et al. Shielding considerations tbr tomotherapy I J 1. Med Phys, 2000, 27(10): 2380- 2384.
  • 4CHESTER R R~ REBECCA S, STEPHEN L M, et al. Out-of-field dosimeter measurements for a helical tomotherapy system I Jl. JApplied Clin Med Phys, 2006, 7(3): 1-11.
  • 5NCRP Report No.151. Structural shielding design and evaluation for megavoltage X and gamma ray radiotherapy facilities IRI . Bethesda, Maryland, American. National Council on Radiation Protection and Measurements, 2005.
  • 6许淑燕.蒙特卡罗方法在实验核物理中的应用[M].北京:原子能出版社,1996..
  • 7NCRP Report No.151. Radiation protection design guide lines for 0.1- 100 MeV Particle Accelerator Facilities IRI. NCRP Report 1977.
  • 8卫生部.放射治疗机房的屏蔽规范.第二部分:电子直线加速器放射治疗机房.BZ/T201.2-2011.G[s].北京:中国标准出版社,2011.
  • 9LANGEN K M, PAPANKKOLAOU N, BALOG J, et al. QA for helical tomotherapy: report of the AAPM Task Group 1481J !. Med Phys, 2010, 37(9): 4817-4853.
  • 10FENWICK J D, TOM6 W A, JARADAT H A, et al. Quality assurance of a helical tomotherapy machine IJi]. Phys Med Biot, 2004, 49(13): 2933-2953.

共引文献2

同被引文献15

引证文献2

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部