期刊文献+

一种基于函数型的热电偶非线性校正方法 被引量:1

A method of nonlinear correction of thermocouple based on function
下载PDF
导出
摘要 由于热电偶的输出信号只有几十毫伏,并且存在较大的非线性,因此测温时需要进行非线性校正。热电偶的非线性校正方法多种多样,以K型热电偶为例,介绍一种基于函数型的热电偶非线性校正方法,可将热电偶1%左右的非线性误差降至0.2%以内,具有校准精度高、实时性好等特点,该方法可广泛应用于各种类型热电偶的非线性校正电路中。 The output signal of the thermocouple was only a few tens of millivolts, due to its large nonlinear, so the result must be corrected. There are various corrected methods of thermocouple, K type as an example, the method based on the function is introduced, the nonlinearity error from 1% down to 0.2% or less, it has high precision, good real-time, and it is also applicable to correction method of other types of thermocouples nonlinear. The method can be widely used in various types of the thermocouple non-linear correction circuit.
出处 《电子设计工程》 2016年第4期59-61,共3页 Electronic Design Engineering
关键词 函数 热电偶 非线性特性 校正电路 function thermocouple nonlinear correcting circuit
  • 相关文献

参考文献5

  • 1本田辰夫.传感器实用电路设计与制作[M].北京:科学出版社.2005.
  • 2范玉久.化工测量与仪表[M].北京:化学工业出版社,1981.
  • 3杨振江,曾余庚.热电偶测温系统中的误差分析与处理[J].化工自动化及仪表,1996,23(3):49-52. 被引量:9
  • 4高桥清,小长井诚.传感器电子学[M].北京:宇航出版社,1987.
  • 5Analog Devices Corp. Data-Acquisition Data Book [EB/OL]. 2013.

二级参考文献1

  • 1[苏]诺维茨基(П·В·Новицкий) 编,《非电量的电测量》翻译组.非电量的电测量[M]机械工业出版社,1983.