期刊文献+

基于xy非线性梯度场磁共振成像研究 被引量:2

MRI Imaging Technology Based on xy Nonlinear Gradient
下载PDF
导出
摘要 为解决非线性梯度成像中心分辨率低的问题,提出了一种基于xy非线性梯度场的成像方法,并讨论了非线性梯度场中心分布曲线对磁共振图像的影响.研究结果表明,在半径r=1.5时,圆形中心分布曲线的图像重建结果最优;与O-Space技术对比实验表明,xy非线性梯度场在不同加速因子R下都得到了更好的重建结果. To eliminate center blurring existing in nonlinear gradient technology based on xy nonlinear gradient field was proposed, and displacement curve in magnetic resonance imaging (MRI) was discussed. circle center displacement curve with diameter r= 1.5 achieves the best comparison experiment with O-Space under different acceleration factor R, field achieves a better image result. imaging, an imagmg the effect of center The results reveal that reconstruction, and in xy nonlinear magnetic
出处 《北京理工大学学报》 EI CAS CSCD 北大核心 2016年第1期64-69,共6页 Transactions of Beijing Institute of Technology
基金 国家自然科学基金资助项目(11105096) 苏州市科技资助项目(SYG201125 SH201207)
关键词 非线性梯度场成像 中心分布曲线 CG迭代算法 nonlinear gradient imaging center placement curve CG iterative algorithm
  • 相关文献

参考文献12

  • 1Hennig J, Welz A M, Schuhz G, et al. Parallel imaging in non-bijective, eurvilinear magnetic field gradients: a concept study[J]. MAGMA, 2008,21(1-2):5- 14.
  • 2何文胜.磁共振影像系统双梯度技术的发展和现状[J].中国医疗设备,2010,25(11):51-53. 被引量:9
  • 3GaHichan G, Cocosco C A, Dewdney A. et al. Simultaneously driven linear and nonlinear spatial encoding fields in mri[J].Magn Reson Med, 2011,65(3):702 -714.
  • 4Stockmann J P, Galiana G, Tam I., et al. In vivo o- space imaging with a dedicated 12 cm z2 insert coil on a human 3t scanner using phase map calibration[J]. MagnReson Med, 2013,69(2):444 - 455.
  • 5Stockmann J P, Ciris P A, Galiana G, et al. ()-space imaging: highly efficient parallel imaging using second order nonlinear fields as encoding gradients with nophaseeneoding[J]. Magn Reson Med, 2010, 64(2): 447 - 456.
  • 6Tam L K, Stockmann J P, Galiana G, et al. Null space imaging., nonlinear magnetic encoding fields designed complementary to receiver coil sensitivities for improved acceleration in parallel imaging[J]. Magn Reson Med, 2012,68 (4) : 1166 - 1175.
  • 7Poole M. Improved equipment and techniques for dynamic shimming in high field MRI[D]. Nottingham.. University of Nottingham, 2007.
  • 8Schultz G, Ullmann P, Lehr H, et al. Reconstruction of mri data encoded with arbitrarily shaped, curvilinear, nonbijective magnetic fields [J]. Magn Reson Med,2010,64(5)~1390 -1403.
  • 9Stockmann J P, Ciris P A, Constable R T. Efficient o- space parallel imaging with higher-order encoding gradients and no phase encoding[C]//Proc Intl Soc Mag Reson Med Honolulu..[s. n. ], 2009:762.
  • 10Pruessmann K P, Weiger M, Rnert P B, et al. Advances in sensitivity encoding with arbitrary k-space trajectories [J]. Magnetic Resonance in Medicine, 2001,46:638 - 651.

二级参考文献13

  • 1Harvey PR,et al. The modular gradient coil: an holistic approach to power efficient and high performance whole-body MRI without peripheral nerve stimulation[J].MAGMA,1999(9): 152-155.
  • 2Wong EC, et al. Echo planar imaging of the human brain using a three axis local gradient coil. In: Proceedings of the SMRM, 11th Annual Meeting, Berlin, 1992. p 105.
  • 3Harvey PR,et al.Modular Gradient CoihA New Concept in High-Performance Whole-Body Gradient Coil Design[J]. Magn Reson Med, 1999,42:561-570.
  • 4Harvey PR, et al. Avoiding peripheral nerve stimulation: gradient waveform criteria for optimum resolution in echoplanar imaging[J]. Magn Reson Med,1994,32:236-241.
  • 5Harvey PR. The modular (twin) gradient coil-high resolution, high contrast, diffusion weighted EPI at 1.0 Tesla[J]. Magnetic Resonance Materials in Physics, Biology and Medicine, 1999,8(1):43-47.
  • 6Steckner MC, et al. Advances in MRI Equipment Design, Software, and Imaging Procedures. AAPM 2006, Orlando, FL.
  • 7Katznelson E,et al.Modular whole-body gradient coil comprising first and second gradient coils having linear gradients in the same direction. US Patent 5736858.
  • 8Harvey PR. Twin Gradient Technology-Potential Advantages For Diffusion Weighted MRI. Elscint Ltd. Internal Publications.
  • 9Ganin A, et al. Modular Gradient System for MRI System. US Patent 6313630.
  • 10Morich M, et al. MRI Gradient with variable field-of-view and apparatus and methods employing the same. US Patent 6538443.

共引文献8

同被引文献1

引证文献2

二级引证文献3