摘要 样品用稀盐酸处理,在酸性条件下石膏样品中的三氧化硫以硫酸根状态存在于待测溶液中,过滤除去酸不溶物。处理好的待测溶液可直接上等离子体发射光谱仪ICAP-6300(美国热电制造)上测定。方法的检出限为0.01%,采用本法测定国家标准物质GBW03109,GBW03109a,GBW03110,GBW03111a,结果表明标准样品测定的相对误差为-0.411%-0.274%,相对误差允许限为1.07%,能够满足分析质量要求。 In order to determine sulphur trioxid,first,the sample should be treated with dilute hydrochloric acid.Under acidic conditions,sulphur trioxid in the gypsum samples occurred in sulfate in testing solution as sulphate,then filtrated acid insoluble substances. After processing,the testing solution can be determined directly on the inductively coupled plasma atomic emission spectrometer. The detection limit of the determination method is 0.01%,and national standard materials are GBW03109,GBW03109 a,GBW03110 and GBW03111 a. It is showed that the relative error of standard sample is 0. 26% - 0. 69%,allowed limit of relative error is 1. 45% which can meet the quality requirements.
出处 《山东国土资源》 2016年第1期71-74,共4页 Shandong Land and Resources
基金 山东省省级整装勘查区地质勘查项目(编号:鲁勘字(2013)1号) 中国地质调查局项目<1:5万区域地质图空间数据库建设(山东省)>(编号:1212010815003)
关键词 三氧化硫 电感耦合等离子体发射光谱仪 精密度与准确度 石膏矿 Sulfur trioxide inductively coupled plasma atomic emission spectrometry precision and accuracy gypsum