摘要
目的:建立测定洛伐他汀分散片含量及含量均匀度的方法,为其仿制生产和贮存期质量控制提供参考。方法:采用高效液相色谱法。色谱柱为Waters MS C18,流动相为乙腈-0.01%磷酸(55∶45,V/V),检测波长为238 nm,柱温为30℃,流速为1.0 ml/min,进样量为20μl。结果:洛伐他汀检测质量浓度线性范围为20.32-609.60μg/ml(r=0.999 9);精密度、稳定性、重复性试验的RSD〈1%;回收率为99.60%-101.68%,RSD=0.8%(n=9)。结论:该方法的重复性好、准确度高,适于洛伐他汀分散片的质量控制。
OBJECTIVE:To establish a method for the determination of content and content uniformity in Lovastatin dispersible tablet,to provide reference for the quality control of its peneric production and storage period. METHODS:HPLC method was adopted on the column of Waters MS C_(18) with mobile phase of acetonitrile-0.01% phosphoric acid(55 ∶ 45,V/V),detection wavelength was 238 nm,column temperature was 30 ℃,flow rate was 1.0 ml/min,and the volume injection was 20 μl. RESULTS:The linear range of lovastatin was 20.32-609.60 μg/ml(r=0.999 9);RSDs of precision,stability and reproducibility tests were lower than 1%;recovery was 99.60%-101.68%(RSD=0.8%,n=9). CONCLUSIONS:The method is reproducible with high accuracy,and suitable for the quality control of Lovastatin dispersible tablet.
出处
《中国药房》
CAS
北大核心
2016年第9期1285-1287,共3页
China Pharmacy