期刊文献+

IC-MC-ICP-MS测量环境水样品中钚同位素比值的方法研究 被引量:4

Determination of Pu Isotope Ratio in Environmental Water Samples by IC-MC-ICP-MS
下载PDF
导出
摘要 为了实现快速分离钚与其他基体,准确测定钚同位素比值,建立了离子色谱(IC)与多接收电感耦合等离子体质谱(MC-ICP-MS)联用测定环境水样品中钚同位素比值的方法。实验选取7.5mol/L硝酸作为色谱柱的上柱液,以0.35mol/L HNO3和0.01mol/L HF的混合溶液为淋洗液,用Pu计数最高点法(peak apex ratio)作为最终同位素比值的测定方法。实验对模拟环境水样进行测定,测量相对标准偏差为0.50%,与Pu标准同位素比值的相对偏差为5.7%。采用该方法对实际环境水样品进行测定,测量相对标准偏差为0.97%,与传统的TBP-7402色谱柱离线分离法测量的钚同位素比值进行对比,相对偏差为1.3%。结果表明,该方法的样品预处理简单、分离速度快、结果可靠,具有广阔的应用前景。 The environmental sampling(ES)analysis is one of important nuclear safeguards' technologies of International Atomic Energy Agency(IAEA),which has been used for detecting the undeclared nuclear activities.Plutonium is regarded as the significant element in nuclear safeguards.In the aspect of ES,bulk analysis is adept at the analysis of plutonium ratio in environmental samples.In this paper,the method of online analysis of plutonium ratio in water samples was established by combination of ion chromatography(IC)and multicollector inductively coupled plasma mass spectrometry(MC-ICP-MS).The separation conditions of liquid chromatographic column,the column acidity,eluent composition were optimized.The AS4A-SC anion chromatographic column was made sure as separation medium.The separation conditions inwhich the acidity of column was 7.5mol/L and the eluent composition was 0.35mol/L HNO3 mixed with 0.01mol/L HF were determined by optimization experiments.The concentrations of uranium in various types of environmental water samples were measured.The interface was determined in order to combine IC and MC-ICP-MS.The effect was confirmed better when using PEEK two way valve to connect IC and MC-ICP-MS.The technique of peek apex ratio was adapted as the method for the measurement of dynamic plutonium ratio by MC-ICP-MS.The simulated water samples were measured.The RSD is 0.50%,and the relative bias compared with the plutonium certified reference material is 5.7%.The actual water samples were also measured.The RSD is0.97%,and the relative bias is 1.3% compared with the typical separation methods by TBP and 7402 chromatography column.This method has high efficiency and fast analytical speed,which can be used for the measurement of plutonium ratio in other environmental samples with more complicated matrices.
出处 《质谱学报》 EI CAS CSCD 北大核心 2016年第2期173-179,共7页 Journal of Chinese Mass Spectrometry Society
关键词 环境水样品 离子色谱(IC) 多接收电感耦合等离子体质谱(MC-ICP-MS) environmental water sample Pu ion chromatography(IC) multicollector inductively coupled plasma mass spectrometry(MC-ICP-MS)
  • 相关文献

参考文献12

  • 1李安利,赵永刚,李静,王林博.核保障的微粒分析与二次离子质谱仪[J].质谱学报,2006,27(3):173-177. 被引量:12
  • 2IAEA safeguards: Staying ahead of the game [R]. IAEA Department of Safeguards, 2007.
  • 3VOGT S, ZAHRADINK P, et al. Bulk analysis of environmental swipe samples[R]. IAEA-SM-367.
  • 4吴剑峰,金玉仁,周国庆,张利兴.逆流色谱分离感应耦合等离子质谱在线测量超痕量钚[J].分析化学,2005,33(10):1397-1400. 被引量:8
  • 5GUNTHER-LEOPOLD I, WALDIS J K, WERNLI B, et al. Measurement of plutonium isotope rati- os in nuclear fuel samples by HPLC-MC-ICP- MS[J]. International Journal of Mass Spectrom- etry, 2005, 242(2): 197-202.
  • 6颜启民 刘寿荪 陈爱民.北京地区土壤和水中钚的浓度研究.环境科学,1984,(5):10-13.
  • 7吴锦海,汪铭侠,王力,王凤仙,任礼华.上海天然水中铀含量[J].上海环境科学,1996,15(10):35-36. 被引量:2
  • 8王琛.FI-MC-ICP-MS分析擦拭样品中钚同位素比值方法研究[D].北京:中国原子能科学研究院,2009.
  • 9林木.用铀矿浓缩物稀土杂质溯源产地的统计学方法研究[D].北京:中国原子能科学研究院,2013.
  • 10刘权卫,吴继宗,肖国平,杨大亭,张永震.铀产品中镎和钚的分离与测定方法[J].核化学与放射化学,2008,30(3):178-183. 被引量:4

二级参考文献30

共引文献26

同被引文献44

引证文献4

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部