摘要
目的:比较光学相干生物测量仪(Lenstar)、三维眼前节分析仪(Pentacam)和A型超声测量仪(A-scan)在白内障患者中测量角膜中央厚度、角膜曲率、前房深度和眼轴长度结果的差异。方法:分别用3种仪器对158例(158眼)白内障患者进行眼部生物测量,比较角膜中央厚度、角膜曲率、前房深度和眼轴长度的结果,采用重复测量的方差分析及Pearson相关分析,一致性比较采用Bland-Altman统计分析法。结果:Lenstar和Pentacam测得的角膜中央厚度分别为536.54±27.90μm和541.46±29.85μm,两者差异有统计学意义(t=-5.439;P<0.001);K值分别为43.87±1.45D和43.86±1.44 D,两者差异无明显统计学意义(t=-0.348,P>0.05)。Pentacam、Lenstar和A超测量的前房深度分别是2.73±0.38mm、2.71±0.38mm和2.85±0.40mm,三者差异均有统计学意义(F=309.94,P<0.001),Pearson相关分析显示三者呈正相关(r=0.989,0.978,and 0.977;P<0.001),但变异系数较小(CV=3.12%)。A超和Lenstar测量的眼轴长度分别是24.28±1.70mm和24.52±1.73mm,两者差异有统计学意义(t=-19.482,P<0.001,r=0.996;P<0.001)。BlandAltman分析显示,对于这几种眼前节参数,三种方法测量的一致性较好。结论:三种仪器测量的结果尽管有一定差异,但Lenstar和Pentacam的生物测量可重复性好,操作简便更易用于白内障患者的检查。
AIM:To compare the central corneal thickness(CCT),keratometry(K) reading,anterior chamber depth(ACD),and axial length(AL) measured with Lenstar and Pentacam with those obtained with the ultrasound(US)pachymetry in the cataract patients.METHODS:A total of 158 eyes of 158 patients were examined in this study.The CCT,average K,ACD and AL obtained by Lenstar and /or Pentacam were compared with those obtained from US pachymetry using repeated-measures analysis of variance,Pearson correlation coefficients and Bland-Altm an analyses.RESULTS:The mean CCT obtained using Lenstar and Pentacam were 536.54 ±27.90μm and 541.46 ±29.85μm(t=-5.439;P〈0.001).The mean Km obtained using Lenstar and Pentacam methods were 43.87±1.45 D and 43.86±1.44 D(t=-0.348,P〈0.05).The mean ACD measured using the Pentacam,Lenstar,and US pachymetry were 2.73±0.38 mm,2.71 ±0.38 mm,and 2.85±0.40 mm,respectively(F=309.94,P 0.001),and they were positively correlated(r=0.989,0.978,and 0.977;P〈0.001) and the coefficient of variation was small(3.12%).The mean AL obtained by US pachymetry and Lenstar were 24.28±1.70 mm and 24.52±1.73 mm,respectively(t=-19.482,P 0.001,r=0.996;P 0.001).The Bland-Altman analysis showed that the three methods were comparable for CCT,Km,ACD and AL.CONCLUSION:Although there were statistically significant differences,the measurements obtained by the Lenstar and the Pentacam were highly repeatable and the instruments easy to use.
出处
《国际眼科杂志》
CAS
2016年第4期594-599,共6页
International Eye Science