组织等效正比计数器对单能中子响应的测量及模拟 被引量:1

Measurement and Simulation of TEPC's Response to Mono-energetic Neutron
下载PDF
导出
摘要 在辐射防护中,组织等效正比计数器(TEPC)可作为主动式剂量计用于中子辐射场监测。采用研制的球形TEPC对加速器单能中子进行测量,获取了0.36、0.65、0.8、1.0、1.3 MeV各单能中子的微剂量谱,根据测得的线能谱计算了吸收剂量和平均品质因子,进一步得到了剂量当量,并与利用注量-周围剂量当量转化因子计算的结果进行了对比。采用FLUKA蒙特卡罗软件模拟了TEPC对各单能中子的响应,并与实验进行了比较。结果发现,理论模拟和实验测量符合很好。实验与理论结果均表明研制的球形TEPC对上述各单能中子具有较好的剂量当量响应。 The tissue equivalent proportional counters are used as active neutron dosemeters in radiation protection.A spherical TEPC was designed and employed for monitoring accelerator-based mono-energetic neutron sources.The microdosimetric spectra of 0.36,0.65,0.8,1.0 and 1.3 MeV mono-energetic neutrons were obtained.The absorbed dose,mean quality factor and dose equivalent value were evaluated from the measured spectra.The comparison between H*(10)which was calculated based on the transforming factor of fluence to ambient dose equivalent and measured dose equivalent was conducted for each mono-energetic neutron.The FLUKA simulation of TEPC was also implemented for comparison.It is found that the simulated data agree well with the experimental results and both reveal a good dose equivalent response of the spherical TEPC.
出处 《原子能科学技术》 EI CAS CSCD 北大核心 2016年第3期531-535,共5页 Atomic Energy Science and Technology
关键词 TEPC 单能中子 微剂量谱 剂量当量响应 FLUKA TEPC mono-energetic neutron microdosimetric spectrum dose equivalent response FLUKA
  • 相关文献

参考文献10

  • 1ANJOMANI J, HANU A R, PRESTWICH W V, et al. Monte Carlo design study for thick gas electron multiplier-based multi-element microdo-simetric detector [ J ]. Nucl Instrum Methods Phys Res A: Accel Spectrum Detect Assoc Equip, 2014, 757(1): 67-74.
  • 2WAKER A J, SZORNEL K, NUNES J. TEPC performance in the CANDU workplace[J]. Radi- ation Protection Dosimetry, 1997, 70(1-4) : 197- 202.
  • 3BALLARINI F, BATTISTONI G, CAMPAN- ELLA M, et al. The FLUKA code: An over- view[J]. Journal of Physics: Conference Series, 2006, 41(1): 151-160.
  • 4ROLLET S, COLAUTTI P, GROSSWENDT B, et al. Monte Carlo simulation of mini TEPC microdosimetric spectra: Influence of low energy electrons[J]. Radiation Measurements, 2010, 45 (10): 1 330-1 333.
  • 5NUNOMIYA T, KIM E, KUROSAWA T, et al. Measurements of lineal-energy distributions for neutrons of 8 keV to 65 MeV by using a tissue equivalent proportional counter[J]. Radiation Pro- tection Dosimetry, 2002, 102(1): 49-59.
  • 6SI-YOUNG C, BONG-HWAN K. Understand- ing of the microdosimetric quantities obtained by a TEPC [J]. Journal of Nuclear Science and Technology, 2008(5): 213-216.
  • 7Recommendation of the International Commission on Radiological Protection, ICRP Publication 60 [R]. Sweden: International Commission on Ra- diological Protection, 1991.
  • 8FAWAZ A. Design and characterization of next- generation tissue equivalent proportional counters for use in low energy neutron fields[D]. Canada: Faculty of Energy Systems and Nuclear Science, University of Ontario Institute of Technology, 2014.
  • 9Conversion coefficients for use in radiologieal protection against external radiation, ICRP Pub- lication 74[R]. Sweden: International Commis- sion on Radiological Protection, 1996.
  • 10张伟华,肖雪夫,王志强,刘毅娜,李春娟,骆海龙.γ辐射场中TEPC微剂量谱的测量与模拟[J].同位素,2013,26(3):144-147. 被引量:2

二级参考文献9

  • 1Menzel HG, Schuhmacher H. Experimental Un- certainties and Applicability of the Proportional counter Method for the Dertermination of Photon Dose Fractions in Neutron Beams [R]//In.. Ion Chambers for Neutron Dosimetry. Broese: EUR6782EN, 1980: 337-343.
  • 2Kota C, Maughan RL, Tattam Da. Use of low- pressure tissue equivalent proportional counters for the dosimetry of neutron beams used in BNCT and BNCEFNT[J]. Med Phys, 2000,27(3).
  • 3Burmeister J, Kota C, Maughan RL, et al. Minia- ture tissue-equivalent proportional counters for BNCT and BNCEFNT dosimetry[J]. Med Phys, 2001,28(9).
  • 4Rossi HH, Zaider M. Microdosimetry and Its Ap- plications, Chapter Ⅱ[M]. Berlin.. Springer, Hei- dlberg, 1996.
  • 5Farahmand M. A novel Tissue-Equivalent Propor- tional Counter based on a Gas Electron Multiplier [D]. TU-Delft, 2004.
  • 6Dietze G, Menzel HG, Buhler G. Calibration of tissue-equivalent proportional counters used as ra- diation protection do dosemeters [J]. Radiation Protection Dosimetry, 1984, 9 (3) :245-249.
  • 7Schrewe UJ, Brede H J, Pihet P, et al. On the cal- ibration of tissue-equivalent proportional counters with buil tin a particle sources[J]. Radiation pro- tection dosimetry, 1988, 23 (1):249-252.
  • 8Rollet S, Beck P, Ferrari A, et al. Dosimetric considerations on tepc fluka-lation and measure- ments[J]. Radiation Protection Dosimetry, 2004, 110, (1) :833-837.
  • 9Srdoc D. Experimental technique of measurement of microscopic energy Distribution in irradiated matter using Rossi Counte[J]. Radiat Res, 1970, 43:302.

共引文献1

同被引文献5

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部