期刊文献+

沉淀分离-电感耦合等离子体质谱法测定方铅矿中铊 被引量:5

Determination of thallium in galena by precipitation separation combined with inductively coupled plasma mass spectrometry
原文传递
导出
摘要 当高浓度Pb共存时,采用四极杆ICP-MS测定Tl,由于^(204)Pb、^(206)Pb的强峰拖尾,分别干扰^(203)Tl、^(205)Tl的测定,使测定结果偏高。本文采用样品分解后加入硫酸,使大量Pb以PbSO_4沉淀形式与Tl分离,分取清液稀释后用ICP-MS测定,解决了测定方铅矿中Tl时Pb的干扰问题。方法检出限(3σ)为0.002μg/g,相对标准偏差(n=12)为3.5%~7.5%。通过与传统泡沫塑料吸附富集-石墨炉原子吸收光谱法比对验证,方法准确可靠。 When high concentrations of Pb were coexisted,the determination of Tl using quadrupole ICP-MS were on the high side due to the strong peak tailing of ^(204)Pb and ^(206)Pb which would interfere the determination of ^(203)Tl and ^(205)Tl. In this paper,sulfuric acid was added after the sample was decomposed. A large number of Pb would be precipitated in the form of PbSO_4 and separated from Tl. The supernatant was diluted for the measurement by ICP-MS. The method solved the determination of Tl in galena from the interference of Pb. The detection limit( 3 σ) was 0. 002 μg / g,and the relative standard deviation( n = 12) was 3. 5% ~ 7. 5%. The method was accurate and reliable comparing with the traditional foam adsorption-graphite furnace atomic absorption spectrometry.
出处 《分析试验室》 CAS CSCD 北大核心 2016年第3期263-265,共3页 Chinese Journal of Analysis Laboratory
基金 国家重大科学仪器设备开发专项(2011YQ06010009)资助
关键词 沉淀分离 电感耦合等离子体质谱法 方铅矿 Precipitation separation Inductively coupled plasma mass spectrometry Tl galena
  • 相关文献

参考文献8

二级参考文献75

共引文献128

同被引文献385

引证文献5

二级引证文献87

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部