期刊文献+

基于IAEA398号报告的绝对剂量校准软件开发 被引量:2

Development of a Software for Absolute Dose Calibration Based on IAEA 398
下载PDF
导出
摘要 本研究运用基于面向对象Python程序开发了基于IAEA398号报告的绝对剂量校准软件(ADC),该软件由光子线修正因子计算模块、电子线修正因子计算模块、校准参数输入可视化和数据输出模块构成。基于IAEA398号报告,该软件能够得出加速器不同能量的光子线和电子线绝对剂量校准所需的用户修正因子,可应用于加速器绝对剂量校准中。实际应用结果表明,运用该软件计算的用户修正因子来校准Truebeam加速器不同能量状态下的绝对剂量效果较好,值得临床推广。 This research developed the software Absolute Dose Calibration(ADC) based on the report of IAEA 398 with the application of The Python program.ADC is composed of calculation of the photon correction factor, calculation of the electron correction factor, as well as input data visualization and data output. Based on the report of IAEA 398, the software can quickly calculate the user correction factors, which were used to precisely measure the absorbed dose and to calibrate beam dose. The results of the ADC was used to calibrate absolute dose of the truebeam accelerator for different energy rays, if choosing different phantoms and ionization chambers, ADC should be able to calculate user correction factors for different energy rays. It can be applied to calibrate absolute dose on the accelerator and is worthy of clinical promotion.
出处 《中国医疗设备》 2016年第1期39-41,共3页 China Medical Devices
基金 四川省医学科学院·四川省人民医院2014年院苗圃科研课题(No.2014007)
关键词 绝对剂量 IAEA398号报告 用户修正因子 加速器 absolute dose IAEA398 user correction factor accelerator
  • 相关文献

参考文献10

  • 1American Association of Physicists in Medicine.TG 21.A protocol for the determination of absorbed dose from high-energy photon and electron beams[J].Med Phys, 1983,10(6):741- 771.
  • 2American Association of Physicists in Medicine.TG 51 .Protocol for clinical reference dosimetry of high-energy photon and electron beam[J].Med Phys, 1999,26(9): 1847-1870.
  • 3文婷,黎静,张晋建,余咏兰,尹智宇,靳怀志,王雅冰,叶柳青.螺旋断层放疗的日常质量控制与质量保证[J].中国医疗设备,2013,28(7):50-52. 被引量:9
  • 4陈旭明,姚升宇,许奕,陈智维,胡喆凯,徐冰,赵国旗.MOSFET探测器在全身放疗质量控制中的应用[J].中国医疗设备,2014,29(1):117-118. 被引量:2
  • 5World Health Organisation.Quality assurance inradiotherapy[S]. 1988.
  • 6International Atomic Energy Agency.TRS 277.Absorbed dose determination in photon and electron beams:aninternational code of practice[S]. 1997.
  • 7胡杰,陶建民,张莹,余根培,张颖.放射治疗中绝对剂量的测量[J].中国医疗器械杂志,2011,35(3):225-228. 被引量:10
  • 8International Atomic Energy Agency.TRS 398.Absorbed dose determination in external beam radiotherapy:An International Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water[S].2000.
  • 9王军良,周振山,盛洪国,齐伟华.IAEA TRS398与TRS277应用于加速器输出量校准的比较[J].中国医学物理学杂志,2011,28(3):2594-2596. 被引量:11
  • 10Andreo P, Saiful M,Huq MS,et al.Protocols for the dosimetry of high-energy photon and electron beams:a comparison of the IAEA TRS-398 and previous international Codes of Practice[J].Phys Med Biol,2002,47(17):3033-3053.

二级参考文献24

  • 1戈伟强,王国民,方庆亮,顾香莲.用MOSFET探测器进行IMRT实时剂量监测[J].中华放射医学与防护杂志,2006,26(2):170-171. 被引量:4
  • 2张绍刚,哈思衡.TBI治疗过程中的实时剂量监控[J].中华放射肿瘤学杂志,2006,15(4):339-341. 被引量:3
  • 3李光俊.X(γ)射线全身放射治疗中的剂量控制[J].医疗装备,2007,20(9):5-6. 被引量:2
  • 4ICRU Report No.24.Detennination of absorbed dose in a patient irradiated by beams of X-or γ-rays in mdiotherapy procedures[R].Bethesdel:ICRU,1976.
  • 5International Atomic Energy Agency.TRS 277 Absorhed Dose Determination in Photon and Electron Beams[R].Vienna:IAEA,1987.
  • 6world Health organization.Quality Assurdnce in Radiotherapy[R].Geneva:WHO,1988.
  • 7Van Dyk J,Barrett RB,Cygler JE,et al.Commissioning and quality assurance of treatment planning computers[J].Int J Radiat Oncol Biol Phys,1993,26:261-273.
  • 8Low DA,Mutic S,Dempsey JF,et al.Quantitative dosimetrjc Verification of an IMRT planning and delivery system[J].Radiother Oncol,1998,49(3):305-316.
  • 9马林,王连元,周桂霞.TomoTherapy肿瘤断层放射治疗[M].成都:四川出版集团,2010.
  • 10Ruchala KJ,Olivera GH,Kapatoes JM.Limited-data image registration for radiotherapy positioning and verification[J].Int J Radiat Oncol Biol Phys,2002,5(4):592-605.

共引文献28