近距离放疗施源器挡块对放疗剂量分布的影响

Impact of Shielding in Applicator on Dose Distribution in Brachytherapy
下载PDF
导出
摘要 本文根据在近距离放疗施源器中添加挡块有利于保护正常组织的原则,使用Ir-192放射源和圆柱形施源器制定治疗计划,并在施源器内分别添加90°和180°挡块,得到计划剂量分布结果。随后借助特制的模体,采用二维电离室矩阵,分别从施源器冠状面和矢状面测量剂量分布,采用γ分析法(3%,3 mm)比较计划剂量分布和实测剂量分布差异。结果表明,挡块遮挡区内组织器官的辐射剂量远低于肿瘤的处方剂量,说明使用挡块可以有效降低辐射剂量,保护正常组织器官。 The paper aims to investigate principle of applying shieding application to spare healthy tissues in brachytherapy with the use of Ir-192 in designing the treatment regimen. The 90° and 180° shieldings were added to the applicator to calculate the result of dose distribution. Dose distribution was then measured using a 2D ion chamber array dosimeter in the transverse and coronal planes for these treatment plans. Gama analysis method(3%, 3 mm) was adopted to compare the dose distribution with the TPS calculations. The results indicate that the dose distribution to the organs in the shield shadow were obviously much lower than the prescribed doses for treatment volume and essentially remained within their tolerance limits for clinical dose prescriptions. It is concluded that the shielding application can effectively reduce the radiation dosage and protect normal tissues and organs.
出处 《中国医疗设备》 2016年第1期48-50,34,共4页 China Medical Devices
基金 乌鲁木齐市科技计划项目(Y12310011)
关键词 近距离治疗剂量分布 带挡块施源器 二维矩阵 辐射剂量 brachytherapy dose distribution shielded applicator two-dimensional matrix radiation dose
  • 相关文献

参考文献12

  • 1Miglierini P, Malhaire JP, Goasduff G,et al.Cervix cancer brachytherapy: high dose rate[J].Cancer Radiother,2014,18( 5-6): 452-457.
  • 2Buyyounouski MK,Davis BJ,Prestidge BR, et al.A survey of current clinical practice in permanent and temporary prostate bmchytherapy: 2010update[J].Brachytherapy,2012,11(4):299-305.
  • 3Stmad V, Sauer R,Niehoff P.Evaluation of(balloon) brachytherapy for partial breast irradiation in breast cancer based on current meta-analyses[J].Strahlenther Onkol,2013,189(10 ):899-901.
  • 4Yao R,Templeton AK,Liao Y, et al.Optimization for high-dose- rate brachytherapy of cervical cancer with adaptive simulated annealing and gradient descent[J].Brachytherapy,2014, 13(4):352-360.
  • 5Uniyal SC,Sharma SD,Naithani UC.Dosimetric verification of a high dose rate brachytherapy treatment planning system in homogeneous and heterogeneous media[J].Phys Med, 2013,29(2):171-177.
  • 6姜瑞瑶,熊霏,黄国锋.不同材质金属植入物对放疗剂量分布影响的研究[J].中国医疗器械杂志,2013,37(4):301-303. 被引量:2
  • 7DeWerd LA,Liang Q,Reed JL,et al.The use of TLDs for brachytherapy dosimetry[J].Radia.Meas,2014,71:276-281.
  • 8Weaver M,Petasecca M,Lerch MLF, et aLDosimetry verification in eye brachytherapy using silicon pixelated detectors[J].Radia. Meas,2011,46(12 ):2010-2013.
  • 9Massilion-JL G,Minniti R,Mitch M.G,et al.High-resolution 3D dose distribution measured for two low-energy x-ray brachytherapy seeds: 1251 and 103Pd[J].Radia Meas,2011, 46(2):238-243.
  • 10Yewondwossen M.Characterization and use of a 2D-array of ion chambers for brachytherapy dosimetric quality assurance[J].Med Dosim,2012,37(3):250-256.

二级参考文献32

共引文献21

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部