NPB840呼吸机吸气压力触发与吸气流量触发比较 被引量:6

Comparison of the Triggers for the Inspiratory Pressure and Inspiratory Flow in NPB840 Ventilator
下载PDF
导出
摘要 目的比较NPB840呼吸机在不同PEEP下吸气压力触发与吸气流量触发呼吸机时触发压力、触发流速、触发时间的差异。方法将FLUKE VT-Plus气流分析仪串联在待测呼吸机与主动模拟肺之间,通过主动模拟肺模拟病人的自主呼吸,气流分析仪采集自主呼吸触发呼吸机时的呼吸参数及波形。结果相同呼气末正压(PEEP)下压力触发的触发压力略高于流量触发的触发压力;相同PEEP下流量触发流速要大于压力触发流速。PEEP为0时,压力触发和流量触发时的触发时间差异最大,随着PEEP值的增加,压力触发和流量触发时的触发时间差异逐渐减少,当PEEP为15 cm H2O时,两者的触发时间接近。结论通过对NPB840呼吸机在不同PEEP下吸气压力触发与吸气流量触发呼吸机时触发压力、触发流速、触发时间的测试,证明NPB840呼吸机使用流量触发方式时触发灵敏度较高,是医生在进行呼吸机治疗时灵敏度设定的首选。 Objective To understand the difference between the pressure and the flow velocity as well as the trigger time of the NPB840 ventilator in different modes of positive-end expiratory pressure(PEEP) under the suction pressure and the inspiratory flow. Methods Connect the FLUKE VT-Plus flow analyzer series to the measured ventilator and then to the simulated lungs with the Norway active system. Through the active simulation of lungs and patients' spontaneous breathing, the analyzer acquisited parameters and waveform from the ventilator triggered by spontaneous breathing. ResultsUnder the same mode of PEEP, the trigger for the inspiratory pressure is slightly higher than that of the inspiratory flow. The flow rate of the same PEEP is greater than that of the inspiratory pressure. When the PEEP is 0, the trigger time between the pressure and the flow rate is different. The time difference decreased with the increase of PEEP value. When the PEEP is 15 cm H2 O, the trigger time between the inspiratory pressure and the flow rate is similar. Conclusion After testing the difference between the pressure and the flow velocity, as well as the difference of the trigger time between the inspiratory pressure and flow rate in the NPB840 ventilator under different modes of PEEP, it is proved that the sensitivity of NPB840 ventilator under the flow mode is superior to that under the mode of of inspiratory pressure.
作者 郑吉锋 白玫
出处 《中国医疗设备》 2016年第1期51-54,共4页 China Medical Devices
关键词 呼吸机 吸气压力触发 吸气流量触发 呼气末正压 ventilator inspiratory pressure trigger inspiratory flow trigger positive-end expiratory pressure
  • 相关文献

参考文献9

二级参考文献52

  • 1张从华,龚岚,宿红,杨勇.精确控制的模拟肺装置用于肺功能测量校准的实验研究[J].四川大学学报(工程科学版),2008,40(5):185-189. 被引量:11
  • 2俞阳.医院医疗设备质量控制管理面临的问题与措施[J].生物医学工程学进展,2013,34(2):132-134. 被引量:5
  • 3吕鹏,陈雅楠,曹德森.呼吸机测试仪的研制[J].北京生物医学工程,2006,25(1):70-72. 被引量:8
  • 4张波 高和.实用机械通气治疗手册[M].北京:人民军医出版社,2004.174-182.
  • 5张璐.泰科PB840呼吸机的系统结构及故障检修,2011(6):113-115.
  • 6张宝刚,孙学玲.泰科Pb840呼吸机故障维修二例[J].医疗装备,2010,25(9):132.
  • 7Aslanian P, Atrous S E L, Isabey D, et al. Effects of flow triggering on breathing effort during partial ventilatory support[J]. Am J Respir Crit Care Med, 1998,157 : 135-143.
  • 8Fauroux B, Leroux K, Desmarais G,et al. Performance of ventilators for non-invasive positive-pressure ventilation in children[J]. Eur Respir J, 2008,31 : 1 300-1 307.
  • 9Bunburaphong T, Imanaka H, Nishimura M,et al. Performance char- acteristies of bilevel pressure ventilators:A lung modes study[J]. Chest, 1997,111 : 1 050-1 060.
  • 10Murata S, Yokoyama K,Sakamoto Y,et al. Effects of inspiratory rise time on triggering work load during pressure-support ventilation:A lung model study[J]. Respir Care, 2010,55 : 878-884.

共引文献34

同被引文献34

引证文献6

二级引证文献24

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部