摘要
目的:建立人乳汁中羟氯喹浓度的HPLC测定方法。方法:取乳汁样本200μL,以氯喹为内标,在0.45mmol·L^(-1) NaOH碱性条件下以甲基叔丁基醚液液提取,氮气吹干后用50μL流动相复溶,取20μL进样。Shimadzu Shim-pack CLC-ODS柱(150×6 mm,5μm)色谱柱分离,流动相为45 mmol·m L^(-1)磷酸二氢钾水溶液:乙腈=87:13(H3P04调节流动相p H至3.0),流速为1.0 mL·min^(-1),柱温40℃,荧光检测激发波长和发射波长分别为337 nm和371 nm。结果:标准曲线在200~2000 ng·m L^(-1)范围内线性良好(r=0.999 9),低、中、高浓度质控样本(400、800、1 500 ng·mL^(-1))的批内、批间变异(RSD)分别为1.07%~2.93%(n=5)和1.82%~3.91%(n=5),方法学和提取回收率分别为90.84%~91.90%(n=10)和67.04%~69.61%(n=3),内标的提取回收率为67.70%±3.94%(n=3)。结论:本方法灵敏度高,准确性好,简便快捷,适用于人乳汁中羟氯喹浓度的测定。
Objective: To develop a high-performance liquid chromatographic(HPLC) method for determination of hydroxychloroquine in human breast milk. Methods: Chloroquine was used as the internal standard. 200 mL breast milk sample were extracted with methyl-tert-butyl ether(MTBE) in an alkaline condition after adding 0.45mmol·L^-1 NaOH solution, followed by drying with nitrogen and re-dissolving with 50 μL mobile phase. The separation was carried out on a Shimadzu Shim-pack CLC-ODS column(150×6 mm, 5μm)at 40°C. The mobile phase consisted of acetonitril -45m mol·mL^-1 KH2PO4 buffer solution(13:87)(pH adjusted to 3.0 by H3PO4). The flow rate was 1.0mL·min^-1 and the injection volume was 20 μL. Fluorescence detection was performed at excitation and emission wavelengths of 337 nm and 371 nm, respectively. Results: The calibration curve of hydroxychloroquine was linear in the range of 200-2 000ng·mL^-1(r =0.999 9). The intra-run and inter-run RSDs were 1.07%-2.93%(n =5)and 1.82%-3.91%(n =5)for the three check samples(400, 800, 1500 ng.mL^(-1)) respectively. The accuracy was 90.84%-91.90%(n =10). The extraction recoveries were 67.04%-69.61% and 67.70%±3.94% for hydroxychloroquine and IS, respectively(n=3). Conclusion: The method is simple, sensitive, and accurate, and it can be used for the measurement of hydroxychloroquine concentration in human breast milk.
出处
《临床药物治疗杂志》
2016年第2期23-26,共4页
Clinical Medication Journal
基金
北京药学会2014年临床药学基金项目