低辐射剂量肝脏CT增强扫描:80kVp结合迭代重建技术的初步研究 被引量:13

Low dose in contrast-enhanced CT scan of liver:80kVp with sinogram affirmed iterative reconstruction technique
下载PDF
导出
摘要 目的:探讨80kVp管电压结合迭代重建技术(SAFIRE)在正常体重人群肝脏CT增强扫描中应用的可行性。方法:前瞻性地对身体质量指数(BMI)18~24kg/m2、临床怀疑肝脏疾病的46例患者采用低辐射剂量(80kVp、420mAs)进行CT增强扫描,对比剂注射采用个性化方案。以滤波反投影法(FBP)和5种强度SAFIRE(1~5)方法分别进行图像重建。图像质量的客观评估指标:图像噪声、对比噪声比(CNR)、品质指数(FOM)。图像质量的主观评估:对图像噪声、血管显示、伪影及图像整体质量进行主观评分,并对动脉晚期的期相进行评估。结果:图像质量客观评估结果:SAFIRE-5图像的噪声最小,CNR及FOM值最高,与其余5组图像间差异均有统计学意义(P〈0.05)。图像质量主观评估结果:SAFIRE-5图像的噪声、血管显示情况及硬化伪影的评估结果均优于其它5组,但"蜡样"伪影的评分明显低于其它各组,使其图像整体评分受到影响;SAFIRE-3图像的整体评分最高,与其它各组间差异有统计学意义;95.6%(44/46)的动脉晚期图像质量合格。结论:正常体型人群行肝脏CT增强扫描时,使用80kVp结合迭代重建技术及个性化对比剂注射方案,能得到较好的图像质量;此种条件下,建议使用SAFIRE-3进行图像重建。 Objective:To evaluate the feasibility of 80 kVp with sinogram affirmed iterative reconstruction(SAFIRE)technique in liver contrast-enhanced CT scan of normal weight adults.Methods:Forty-six patients with body mass index of18~24kg/m2 and suspected of hepatic diseases were prospectively recruited and then underwent liver contrast-enhanced CT using low dose(80kVp and 420mAs)and personalized contrast injection protocol.Filtered back projection(FBP)and five SAFIRE algorithms were used respectively for image reconstruction,and 6sets of images were obtained for each patient.For objective evaluation,image noise,contrast-to-noise ratio(CNR)and figure of merit(FOM)were measured and calculated.For subjective evaluation,subjective score of image noise,vessel conspicuity,artifacts and overall image quality were assessed,as well as the timing of late arterial phase.Results:For objective analysis,images with SAFIRE-5algorithm showed the least image noise and the highest CNR and FOM,with statistical difference from those of the other 5groups(P〈0.05).For subjective analysis,images with SAFIRE-5algorithm showed the highest scores of image noise,vessel conspicuity and beam hardening artifacts,and the least score of reconstruction artifacts.As for overall image quality,the image scores of SAFIRE-3were the highest compared with the other 5groups.95.6%(44/46)of the patients could obtain appropriate images in late arterial phase.Conclusion:In liver contrast-enhanced CT,80 kVp with iterative reconstruction technique and personalized contrast media injection protocol can obtain high quality post-contrast images;SAFIRE-3algorithm is recommended for image reconstruction.
出处 《放射学实践》 北大核心 2016年第4期316-320,共5页 Radiologic Practice
关键词 体层摄影术 X线计算机 管电压 迭代重建技术 图像质量 肝脏疾病 Tomography X-ray computed Tube voltage Iterative reconstruction technique Image quality Liver disease
  • 相关文献

参考文献17

  • 1Brenner DJ, Hall [J]. Computed tomography: an increasing source of radiation exposure[J]. N Engl J Med, 2007,357 (22) 2277 2284.
  • 2Sigal-Cinqualbre AB, Hennequin R, Abada HT, et al. Low-kilo- voltage multi-detector row chest CT in adults: feasibility and effect on image quality and iodine dose [J]. Radiology, 2004, 231 (1): 169-174.
  • 3Ertl-Wagner BB, Hoffmann RT,Bruning R, et al. Multi-detector row CT angiography of the brain at various kilovoltage settings [J]. Radiology, 2004,231 (2) : 528-535.
  • 4张晓东,郭小超,王霄英.体型特异性剂量估计的概念和方法[J].放射学实践,2013,28(3):312-314. 被引量:38
  • 5Borg M,Badr I,Royle GJ. The use of a figure-of-merit (FOM) for optimisation in digital mammography a literature review[J]. Ra- diat Prot Dosimetry,2012,151(1)81-88.
  • 6American College of Radiology. Liver imaging reporting and data system version 2014[J]. Accessed 12 2015, from http: //www. act. org/Quality-Safety/ Resources/LIRADS.
  • 7Nakaura T, Awai K, Oda S, et al. Low-kilovoltage, high-tube-cur rent MDCT of liver in thin adults: pilot study evaluating radiation dose,image quality, and display settings[J]. AJR, 2011,196 (6) : 1332-1338.
  • 8王蕊,张保翠,王霄英,罗健,王鹤.80kVp、低浓度对比剂冠状动脉CTA检查的初步研究[J].放射学实践,2013,28(5):501-504. 被引量:18
  • 9蒋孟茜,王鹤,王霄英,邹明宇,衣闯,胡刚,王蕊,赵承琳,姜健,张保翠,罗健,高福生.80kVp条件下碘克沙醇(270mgI/mL)应用于冠状动脉CTA检查的可行性研究[J].放射学实践,2014,29(3):249-253. 被引量:16
  • 10戴丽娟,王霄英,郭小超,高歌.迭代重建对日常CT检查平均辐射剂量的影响[J].放射学实践,2013,28(3):291-293. 被引量:11

二级参考文献62

  • 1王霄英,薛华丹.泌尿生殖系统影像学2011年度进展报告[J].中国继续医学教育,2011,3(8):75-85. 被引量:23
  • 2陈渝晖,杨亚英,韩丹.CT低剂量在临床应用中的研究进展[J].昆明医科大学学报,2009,32(S2):276-279. 被引量:5
  • 3李明利,金征宇,张云庆,王沄,赵文敏,陆菁菁.多层螺旋CT头颈部低剂量对比剂血管成像的可行性研究[J].临床放射学杂志,2005,24(3):214-217. 被引量:42
  • 4蒋学祥,邱建星,王继琛,孙晓伟,高莉,高福生,王仪生.64层螺旋CT冠状动脉成像质量的研究[J].中国医学影像技术,2006,22(10):1452-1455. 被引量:31
  • 5张龙江,祁吉.对比剂肾病:一个值得关注的问题[J].中华放射学杂志,2007,41(8):882-884. 被引量:69
  • 6Leipsic J, Labounty TM, Heilbron B, et al. Estimated radiation dose reduction using adaptive statistical iterative reconstruction in coronary CT angiography The RASIR study[J]. AJR, 2010, 195 (3) : 655-660.
  • 7Nelson RC, Feuerlein S, Boll DT. New iterative reconstruction techniques for cardiovascular computed tomography: How do they work,and what are the advantages and disadvantages[J]. Cardio- vase Comput Tomogr,2011,5(5) :286-292.
  • 8Hara AK, Paden RG, Silva AC, et al. herative reconstruction tech- nique for reducing body radiation dose at CT:feasibility study[J]. AJR,2009,193(3) :764 -771.
  • 9Sagara Y, Hara AK, Pavlicek W, et al. Abdominal CT : comparison of low dose CT with adaptive statistical iterative reconstruction and routine dose CT with filtered back projection in 53 patients [J]. AJR,2010,195(3),713-719.
  • 10Kilic K,Erbas G,Guryildirim M,et al. Lowering the dose in head CT using adaptive statistical iterative reeonstruction[J] . AJNR, 2011,32(9) :1578-1582.

共引文献195

同被引文献85

引证文献13

二级引证文献40

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部