期刊文献+

旋转摆位误差对调强放疗剂量准确性的影响 被引量:20

Impact of rotational set-up error in dose veracity of intensity-modulated radiotherapy
下载PDF
导出
摘要 目的:分析旋转摆位误差对调强放疗剂量准确性的影响。方法:随机选取10例鼻咽癌患者的适形调强放疗(intensity-modulated radiation therapy,IMRT)计划,以患者放疗计划做模板计划(template plan,TP),然后分别旋转治疗床角度为±0.5、±1、±1.5、±2、±2.5、±3、±3.5、±5°形成不同的模拟计划(simulate plan,SP),利用Omni Pro-IMRT软件对SP与TP进行比较,观察不同旋转角度在3%/3 mm标准下对gamma值的影响。结果:在旋转误差≤1.5°时,gamma值之间无统计学意义(P>0.05);旋转角度≥2°时,gamma值之间有统计学意义(P<0.05)。沿相反方向旋转相同角度后,gamma值之间无统计学意义(P>0.05)。结论:旋转摆位误差对靶区剂量准确性的影响与靶区长度有关;旋转误差的校正阈值为1.5°可能更为合理。 Objective To analyze the effect of the rotational set-up error in dose veracity of intensity-modulated radiotherapy. Methods Totally 10 IMRT treatment plans of nasopharyngeal carcinoma patients were selected for the study. The template plans (TP) were designed with the treatment plans. Then the couch was rotated at ±0.5, ±1, ±1.5, ±2, ±2.5, ±3, ±3.5 and ±5° to create the simulate plans (SP). OmniPro-IMRT was used to compare the influences of gamma indexes with different rotate angles between SP vs TP. Then the gamma values were analyzed to evaluate these influences. Results The gamma values of different rotational set-up errors in 3%/3 mm were as follows: when the rotation set-up error was not more than 1.5°, there had no significant differences (P〉0.05), while had significant differences when the rotation set-up error was not less than 2° (P〈0.05). The gamma values with the same rotate degree at the opposite directions had no significant difference (P〉0.05). Conclusion The influence of the rotational position error on the accuracy of the target dose is related to the target length. Rotation set-up error correction threshold of 1.5° may be reasonable.
出处 《医疗卫生装备》 CAS 2016年第4期109-111,共3页 Chinese Medical Equipment Journal
关键词 旋转摆位误差 调强放疗 剂量准确性 rotational set-up error IMRT close veracity
  • 相关文献

参考文献9

二级参考文献78

共引文献153

同被引文献150

引证文献20

二级引证文献76

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部