期刊文献+

石墨炉原子吸收法测定大米中铅的不确定度评定 被引量:3

Evaluation of uncertainty in determination of lead in rice by graphite furnace atomic absorption spectrometry
原文传递
导出
摘要 目的通过分析实验中的各不确定度分量,评定测量结果的不确定度。方法使用GB/T5009.12-2010采用石墨炉原子吸收法测定大米中的铅,参照JJF 1059.1-2012建立数学模型,对每个不确定分量进行评估。结果通过对每个不确定度分量的计算,得出最终扩展不确定度为0.030 mg/kg(包含因子k=2),结果报告为0.70±0.030 mg/kg。结论石墨炉原子吸收法测定大米中铅,对不确定度影响较大的主要为重复性引入的不确定度和拟合曲线引入的不确定度。 Objective To evaluate uncertainty of measurement results by analyzing the uncertainty components in the experiment. Methods The lead in rice was determined by graphite furnace atomic absorption according to GB / T5009. 12- 2010. The mathematical model was established according to JJF 1059. 1- 2012 to evaluate each uncertainty component.Results By the calculation of each uncertainty component,it was concluded that eventually extended uncertainty was 0.030 mg / kg( including factor k = 2),the report result was( 0. 70 ± 0. 030) mg / kg. Conclusion The uncertainty were mainly influenced by uncertainties in the introduction of repeatability and fitting curve
出处 《安徽预防医学杂志》 2016年第1期14-17,共4页 Anhui Journal of Preventive Medicine
关键词 石墨炉原子吸收 不确定度评定 Graphite furnace atomic absorption Lead Uncertainty evaluation
  • 相关文献

参考文献5

  • 1GB2762-2012.食品安全国家标准食品中污染物限量[S].2012:7.
  • 2JJF1059.1-2012,测量不确定度评定与表示[S].1999.
  • 3GB/T5009.12-2010,食品安全国家标准食品中铅的测定[S].中华人民共和国卫生部,2010.
  • 4JJG196-2006,中华人民共和国计量检定规程常用玻璃仪器[s].国家质量监督检验检疫总局,2006.
  • 5武洪丽,鲜青龙,王琴.火焰原子吸收法测定水中镁元素含量的不确定度评定[J].环境卫生学杂志,2014(2):195-199. 被引量:9

二级参考文献5

共引文献70