摘要 目的探讨通过改变K空间填充方式来降低3D DCE-MRA顺磁性造影剂剂量的可行性.方法临床双下肢动脉狭窄病变患者行动脉3D DCE-MRA磁共振检查,60例患者随机分为两组,双倍剂量组,K空间填充方式选择中心填充,单倍剂量组,K空间填充方式选择椭圆中心填充.结果单倍剂量组及双倍剂量组图像质量无差异,单倍剂量组K空间填充方式为椭圆填充,造影剂的剂量减少.结论改变K空间填充方式可以降低3D DCE-MRA顺磁性造影剂剂量,3D DCE-MRA是一种无创、在活体上可重复实施、又能进行形态学评价的检查方法. Objective To explore the feasibility of reducing paramagnetic contrast agent dose for three dimensional dynamic contrast enhanced mr angiography(3D DCE-MRA) through changing K-space filling approach.Method A total of 60 patients with double lower limbs artery stenosis who underwent artery 3D DCEMRA examination were enrolled in this study.The patients were randomly divided into two groups,the single dose one group accepting center filling approach and the double dose group accepting ellipse center filling method.Results There was no difference in image quality between the single dose group and the double dose group.Kspace filling approach of the single group was ellipse filling which reduced the dosage of contrast agent.Conclusion The change of K-space filling approach can reduce the paramagnetic contrast agent dose for 3D DCE-MRA.Three dimensional DCE-MRA is a noninvasive and repeatable in vivo method,which can also be applied for morphological evaluation.
作者 陈彪
出处 《北华大学学报(自然科学版)》 CAS 2016年第3期363-365,共3页 Journal of Beihua University(Natural Science)
基金 吉林省教育厅科学技术研究项目(2012155)
关键词 K空间技术 磁共振动态增强成像(3D DCE-MRA) 造影剂 下肢血管病变 K-space technology 3D DCE-MRA contrast agents lower extremity vascular lesions
  • 相关文献

参考文献6

二级参考文献83

  • 1周敏,张敏鸣,刘琳.磁共振成像中的敏感性编码技术[J].中国医疗器械杂志,2005,29(3):196-198. 被引量:1
  • 2陈武凡.并行磁共振成像的回顾、现状与发展前景[J].中国生物医学工程学报,2005,24(6):649-654. 被引量:19
  • 3夏爽,祁吉.MR并行采集技术的优势[J].国外医学(临床放射学分册),2006,29(5):348-353. 被引量:18
  • 4尹培尧,史进忠,程炜.加快MR扫描速度的常用方法[J].医疗设备信息,2007,22(3):104-105. 被引量:1
  • 5[1]Zhao HT, Fang H, Wu XH. The Concept of K-Space and Its Application of Magnetic Resonance Imaging [J]. Zhongguo Yixue Yingxiangxue Zazhi (Chin J Med Imaging),1998; 7(3):225-227.
  • 6[2]Hennig J. K-space sampling strategies [J]. Eur Radiol, 1999; 9(6): 1020-1031.
  • 7[3]Mason GF, Harshbarger T, Hetherington HP. A method to measure arbitrary K-space trajectories for rapid MR imaging [J]. Mag Reson Imaging, 1997; 38(2): 492-496.
  • 8[4]M.Louis L, Brian KR. Effects of Polar Sampling in K-Space [J]. Magn Reson Med, 1996; 36(3): 940-949.
  • 9[5]Xu ZX, Zhao HT, Xu CJ, Liu MS, Chang YJ. Analysis the nonlinlinearly sample in K-space of magnetic resonance imaging [J]. Di-si Junyi Daxue Xuebao (J Fourth Mil Med Univ), 2000; 21(Suppl): 150-152.
  • 10[6]Scheffler K, Hennig J. Reduced circular field-of-view imaging [J]. Magn Reson Med, 1998; 40(3): 474-480.