COBIO XS尿沉渣分析仪在尿路感染筛查中的应用 被引量:3

Evaluation of COBIO XS automated urine sediment analyzer in screening urinary tract infection
下载PDF
导出
摘要 目的评价COBIO XS尿沉渣分析仪在尿路感染中的应用。方法应用COBIO XS尿沉渣分析仪检测182例疑似尿路感染中段尿标本白细菌和细菌数,以定量尿细菌培养作为诊断尿路感染金标准,应用受试者工作特性(ROC)曲线评价各指标诊断效能。结果尿培养阳性标本56例占30.8%,分离率依次为大肠埃希菌30株占53.6%、肠球菌属9株占16.1%、变形杆菌属6株占10.7%。白细胞计数曲线下面积为0.808(95%CI 0.710~0.867),当白细胞临界值为50/μL,诊断尿路感染敏感度89.0%,特异度83.8%,阳性预测值86.3%,阴性预测值93.7%。细菌计数曲线下面积为0.899(95%CI 0.797~0.923),当细菌临界值为110/μL,诊断尿路感染敏感度82.5%,特异度85.6%,阳性预测值78.5%,阴性预测值88.0%。结论 COBIO XS尿沉渣分析仪具有快速、准确等优点,可作为筛查尿路感染工具。 Objective To evaluate the clinical application of COBIO XS automated urine sediment analyzer for screening urinary tract infection(UTI).Methods Midstream urine samples were collected from 182 patients with suspected UTI.White blood cell counts(WBC)and bacteria counts(BACT)were determined by COBIO XS automated urine sediment analyzer.By comparing with quantitative urine culture as the golden standard of UTI.The area under the ROC curve(AUC)was constructed to evaluate the discriminative power of variables to predict UTI.Results The positive rate of urine culture was 30.8%.The isolated rate of Escherichia coli,Enterococcus species and proteus species were 53.6%,16.1%,10.7%,respectively.ROC curve analysis showed that AUC was 0.808(95%CI0.710-0.867)for WBC and 0.899(95%CI 0.797-0.923)for BACT.The best cutoff value for WBC was 50/μL,sensitivity was 89.0%,specificity was 83.8%,positive predictive value was 86.3% and negative predictive value was93.7%.The best cutoff value for BACT was 110/μL,sensitivity was 82.5%,specificity was 85.6%,positive predictive value was78.5%and negative predictive value was 88.0%.Conclusion COBIO XS analyzer is adopted to conduct rapid and reliable a screening tool for UTI.
出处 《国际检验医学杂志》 CAS 2016年第9期1221-1222,1225,共3页 International Journal of Laboratory Medicine
关键词 尿沉渣分析仪 尿路感染 白细胞计数 细菌计数 urine sediment analyzer urinary tract infection white blood cell counts bacteria counts
  • 相关文献

参考文献14

  • 1Viekers D,Ahmad T,Coulthard MG. Diagnosis of urinary tract in- fection in children: fresh urine microscopy or culture[J]. Lancet, 1991,338(8770) ..767-770.
  • 2Zaman Z, Roggeman S, Verhaegen J. Unsatisfactory performance of flow cytometer UF-100 and urine strips in predicting outcome of urine eultures[J]. J Clin Microbiol,2001,39(11) ..4169-4171.
  • 3Aspevall O, Hallander H, Gant V, et al. European guidelines for urinalysis:a collaborative document produced by European clinical microbiologists and clinical chemists under ECLM in collaboration with ESCMID[J]. Clin Microbiol Infect,2001,7(4) 173-178.
  • 4Shah AP,Cobb BT, Lower DR, et al. Enhanced versus automated urinalysis for screening of urinary tract infections in children in the emergency department[J]. Pediatr Infect Dis J, 2014,33 (3): 272-275.
  • 5Van Der Zwet WC, Hessels J,Canbolat F, et al. Evaluation of the sysmex UF-1000i urine flow cytometer in the diagnostic work- up of suspected urinary tract infection in a Dutch general hospital [J]. Clin Chem Lab Med, 2010,48(12) 1765-1771.
  • 6Pieretti B,Brunati P,Pini B,et al. Diagnosis of bacteriuria and leu- kocyturia by automated flow cytometry compared with urine cul- ture[J]. J Clin Microbiol, 2010,48 (11) 3990-3996.
  • 7范宏佳,汤瑾,王坚镪,陈卫宾,陈瑜,高锋.尿常规检测在尿路感染诊断中的价值[J].检验医学,2014,29(12):1207-1211. 被引量:43
  • 8魏运梅,易光明,熊军.COBIO全自动尿液分析工作站性能评价[J].检验医学,2012,27(7):531-533. 被引量:3
  • 9吴新忠,庞鑫,姚妍娇.UF-100尿液分析仪细菌计数对泌尿道感染的诊断性能评价[J].实用医学杂志,2009,25(16):2764-2766. 被引量:6
  • 10齐杰,潘健,韩江,程实,董全,张婷菊,马蕊,刘贵建.尿流式有形成分及干化学分析在尿路感染诊断中的应用评价[J].中华检验医学杂志,2009,32(6):630-634. 被引量:36

二级参考文献36

  • 1刘金彪,徐国鹏.尿亚硝酸盐试验诊断尿路感染的价值分析[J].中华医学丛刊,2004,4(8):44-45. 被引量:10
  • 2夏邦世,吴金华.Kappa一致性检验在检验医学研究中的应用[J].中华检验医学杂志,2006,29(1):83-84. 被引量:212
  • 3陈国强,沈培琼,沈燕,刘亚新,刘建栋.自动尿液分析仪筛检医院尿路感染临床意义探讨[J].中华医院感染学杂志,2006,16(12):1431-1433. 被引量:3
  • 4贝政平.内科疾病诊断标准.2版.北京:科学出版社,2007:237.
  • 5陈灏珠.实用内科学[M].北京:人民卫生出版社,1973:585.
  • 6Kumar P, Clark M. Clinical Medicine(临床内科学).王德炳,译.6版.北京:北京大学医学出版社,2008:665-670.
  • 7Zaman Z, Roggeman S, Verhaegen J. Unsatisfactory performance of flow cytometer UF-100 and urine strips in predicting outcome of urine cultures. J Clin Microbiol,2001,39:4169-4171.
  • 8Okada H, Sakai Y , Miyazaki S, et al. Detection of significant bacteriuria by automated urinalysis using flow cytometry. J Clin Microbiol, 2000,38 : 2870-2872.
  • 9Buchsbaum GM, Albushies DT, Guziek DS, Utility of urine reagent strip in screening women with incontinence for urinary tract infection. Int Urogynecol J Pelvic Floor Dysfunct, 2004,15:391- 393.
  • 10Armengol CE, Hendley JO, Sehlager TA. Should we abandon standard microscopy when screening for urinary tract infections in young children. Pediatr Infect Dis J, 2001,20 : 1176-1177.

共引文献96

同被引文献19

引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部