期刊文献+

HT系统兆伏级CT图像质量和剂量计算研究

Image quality and dose calculation of megavoltage computed tomography in helical tomotherapy
原文传递
导出
摘要 目的定量对HT系统兆伏级CT图像质量及剂量计算准确性进行评估。方法应用兆伏级CT扫描Cheese phantom模体,分别对图像成像几何精度、噪声、图像均匀性、空间分辨率、密度-CT值的转化及剂量计算准确性进行测试,并与常规千伏级CT进行对比分析。结果兆伏级CT图像的成像几何精度在三维方向均〈mm,MVCT图像噪声、均匀性、空间分辨率都差于千伏级CT。基于CT图像进行剂量重建的DVH均与千伏级CT计划的DVH有很好一致性。结论兆伏级CT图像几何尺寸精确,成像剂量低,剂量计算精确,满足于临床要求。 Objective To quantitatively investigate the image quality and dose calculation accuracy of megavohage computed tomography (CT) in helical tomotherapy. Methods The megavoltage CT was used to scan the Cheese phantom, and the geometric accuracy of images, noise, image uniformity, spatial resolution, density-CT value conversion, and dose calculation accuracy were analyzed and compared with conventional kilovohage CT. Results The geometric accuracy of megavohage CT images was within 2 mm in three directions. The megavoltage CT was inferior to kilovoltage CT in terms of image noise, uniformity, and spatial resolution. The dose-volume histogram (DVH) in dose reconstruction based on CT images was well consistent with DVH in the kilovoltage CT plan. Conclusions The megavohage CT can obtain images with accurate geometric dimensions and has a low imaging dose and accurate dose calculation, which meets clinical requirements.
出处 《中华放射肿瘤学杂志》 CSCD 北大核心 2016年第5期500-503,共4页 Chinese Journal of Radiation Oncology
关键词 提层摄影术 X线计算机 兆伏级 图像质量 剂量重建 Tomography,X-ray computed,megavoltage Image quality Dose calculation
  • 相关文献

参考文献8

  • 1Mackie TR, Holmes T, Swerdloff S, et al. Tomotherapy: a new concept for the delivery of dynamic confomml radiotherapy [ J ]. Med Phys, 1993,20(6) : 1709-1719. DOI : 10.1118/1.596958.
  • 2Langen KM, Papanikolaou N, Balog J, et al. QA for helical tomotherapy:report of the AAPM Task Group 148 [ J] .Med Phys, 2010,37(9) :4817-4853.DOI: 10.1118/1.3462971.
  • 3Ruchala KJ,Olivera GH,Schloesser EA, et al. Megavoltage CT on a tomotherapy system [ J]. Phys Med Biol, 1999 : 44 ( 10 ) : 2597- 2621.DOI : 10.1088/0031-9155/44/10/316.
  • 4Meeks SL, Harmon JF Jr, Langen KM, ct al. Performance characterization of megavoltage computed tomography imaging on a helical tnmotherapy unit [ J ].Med Plays,2005,32 ( 8 ) : 2673-2681. DOI : 10.1118/1.1990289.
  • 5Langen KM, Meeks SL, Poole DO, et al. The use of megavoltage CT (MVCT) images for dose recomputations [J]. Pbys Med Biul, 2005,50(18) :4259-4276.0OI: 10.1088/0031-9155/50/18/002.
  • 6Shah AP, Langen KM, Ruehala K J, et al. Patient dose from megavultage computed tomography imaging [ J] .Int J Radiat Onenl Biol Phys, 2008,70 ( 5 ) : 1579-1587. DOI : 10.1016/j. ijrobp. 2007. 11.048.
  • 7Ramsey CR, Langen KM, Kupelian PA, et al. A technique for adaptive image-guided helical tomotherapy for lung cancer [ J] Ant J Radiat Oncol Biol Phys,2006,64(4) : 1237-1244.DOI: 10.1016/ j.ijrohp.2005.11.012.
  • 8刘景鑫,杨海山.螺旋CT图像噪声影响因素的实验分析[J].中华放射学杂志,2000,34(1):9-12. 被引量:78

二级参考文献1

共引文献77

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部