期刊文献+

基于ARM的32电极电阻抗成像系统设计 被引量:1

Design of 32 Electrodes Electrical Impedance Tomography System Based on ARM
下载PDF
导出
摘要 电阻抗成像是一种新型的医学成像技术,在生物信息检测与成像方面具有广阔的应用前景。针对现有电阻抗成像系统采集速度不高、实时性较低、分辨率不理想的情况,进行了改进设计。采用ARM处理器构建嵌入式电阻抗成像系统,搭建32电极物理模型作为检测对象,利用32通道多路开关实现激励电流和电压采集通道的切换,向模型中注入激励电流信号并通过前置滤波、两级放大等信号处理技术获得成像数据。采用敏感矩阵算法对有无目标时的数据差值进行图像重建。实验结果表明,该系统具有较好的成像效果,分辨率高,能有效地实现对目标区域的检测。 Electrical impedance tomography is a new technique of medical imaging, which has a broad application prospect in biological information detection and imaging. In this paper,improved design is carried out in the case of low acquisition speed,poor real-time per- formance and unsatisfying resolution of the EIT system. In the experiment, an ARM processor is used to construct the embedded EIT sys- tem ,the 32-electrode physical model is acted as a detection object,and the 32-channel multiplexer realizes the switching among electrode excitation and measurement channels. The excitation current signal injected into the model is passed through the pre-filter,two-stage am- plifier and other signal processing technique to get imaging data, from which difference between existence and non-existence of the tes- ting object is obtained. Further,image reconstruction of electrical impedance tomography is achieved through the sensitivity matrix algo- rithm. The experiment shows that the system has a good effect, and high resolution, and could effectively realize the detection of the target area.
作者 饶水发 戎舟
出处 《计算机技术与发展》 2016年第6期138-141,共4页 Computer Technology and Development
基金 国家自然科学基金资助项目(61203213)
关键词 电阻抗成像 ARM处理器 32电极物理模型 32通道多路开关 两级放大 敏感矩阵算法 electrical impedance tomography ARM processor 32 electrodes physical model 32 channels multiplexer two-stage amplifi- cation sensitivity matrix algorithm
  • 相关文献

参考文献15

  • 1Brown B H. Medical impedance tomography and process im- pedance tomography : a brief review [ J ]. Measurement Science and Technology ,2001,12 ( 8 ) :991-996.
  • 2任超世.生物电阻抗测量技术[J].中国医疗器械信息,2004,10(1):21-25. 被引量:95
  • 3Gargiulo G D, Cohen G, McEwan A L, et al. Active electrode design suitable for simultaneous EIT and EEG [ J ]. Electronics Letters ,2012,48 ( 25 ) : 1583-1584.
  • 4徐管鑫,王平,何为.实时电阻抗成像系统及实验研究[J].仪器仪表学报,2005,26(9):886-890. 被引量:11
  • 5Smith R W,Freeston I L,Brown B H. A real-time electrical impedance tomography system for clinical use- design and preliminary results[ J ]. IEEE Transactions on Biomedical En- gineering, 1995,42 ( 2 ) : 133 - 140.
  • 6Yue X, MeLeod C. FPGA design and implementation for EIT data acquisition [ J ]. Physiological Measurement, 2008, 29 (10) : 1233-1246.
  • 7Rafiei-Naeini M, McCann H. Low-noise current excitation sub-system for medical EIT[ J]. Physiological Measurement, 2008,29(6) : 173-184.
  • 8童倜,张伟兴.电阻抗成像系统中电压控制电流源的设计[J].电子设计工程,2012,20(1):116-119. 被引量:9
  • 9Yan W, Hong S, Chaoshi R. Optimum design of electrode structure and parameters in electrical impedance tomography [ J ]. Physiological Measurement,2006,27 (3) :291-306.
  • 10张伟兴,马艺馨.EIT激励电流切换用模拟多路开关的选择[J].仪表技术与传感器,2013(10):105-107. 被引量:7

二级参考文献35

  • 1付峰,秦明新.电阻抗断层图像重构算法[J].国外医学(生物医学工程分册),1995,18(5):255-262. 被引量:12
  • 2李蛟,杨仁锟,肖峻.2.4GHz无线技术标准及ZigBee抗干扰性能[J].电信工程技术与标准化,2006,19(3):31-35. 被引量:34
  • 3汤孟兴.动态电阻抗断层成像技术[J].国外医学(生物医学工程分册),1996,19(4):206-207. 被引量:1
  • 4Franco S. Design with operational amplifiers and analog integrated circuits 2nd edition [M]. New York:McGraw-Hill, 1998.
  • 5Ross A S,Saulnier G J,Newell J C,et al. Current source design for electrical impedance tomography[J]. Physiol. Meas, 2003 (24) : 509-516.
  • 6Bertemes-Filho P,Brown B H,Wilson A J. A compar-sion of modified Howland circuits as current generators with current mirror type circuits[J]. Physiol. Meas,2000(21 ) : 1-6.
  • 7Munoz D R,Moreno J S,Berga S C,et al. Constant current drive for resistive sensors based on generalized impedance converter [J]. IEEE Transactions on Instrumentation and Mea- surement, 2008,57 (10) : 2290-2296.
  • 8Boone K. , Barber D. C. , Brown B. H. , et al.. Imaging with electricity:review of the European concerted action on impedance tomography [J]. Journal of Medical Engineering & Technology, 1997,21 (6) : 201-232.
  • 9Isaacson D. Distinquishability of conductivity by electric current computed tomography [J]. IEEE Trans. , 1986,MI-5(2):91-95.
  • 10Peter Metherall BEng (Hons) MMedSci. Three dimensional electrical impedance tomography of the humanthorax [D]. Department of Medical Physics and Clinical Engineering, University of Sheffield ,January, 1998.

共引文献121

同被引文献13

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部