能谱CT脑动脉造影单能量图像质量优化研究

Imaging quality optimization of CT cerebral angiography by using single source spectral CT
下载PDF
导出
摘要 目的:探讨能谱CT不同单能量水平对脑动脉造影图像质量的影响,以确定最优化的单能量图像。方法:采用能谱CT对53例患者行脑CT动脉造影(CTA)检查。应用能谱分析软件,以10 ke V间隔重建40~140 ke V单能量脑CTA图像。测量并计算基底动脉、颈内动脉颅内段、大脑前、中、后动脉图像对比噪声比(CNR)、信号噪声比(SNR);采用5分法对脑CTA图像的主观质量进行盲法独立评分。分别采用单因素方差分析和非参数秩和检验对各单能量脑CTA图像的客观和主观质量进行比较评价。结果:图像客观质量比较,60、70 ke V脑CTA图像CNR、SNR值较其他各单能量图像显著增加(P〈0.05),但60、70 ke V间差异无统计学意义(P〉0.05)。单能量脑CTA图像噪声以70 ke V为界,70~140 ke V图像噪声显著低于40、50、60 ke V单能量水平。而且70 ke V图像主观质量评分结果较60 ke V及其他各单能量图像评分结果显著增加。结论:能谱70 ke V水平单能量脑CTA图像兼顾图像CNR、SNR及图像噪声,可作为最优化单能量水平。 Objective: To determinate the optimal keV levels of single source spectral CT by evaluating imaging qualities of CT cerebral angiography.Methods: A total of 53 cases underwent spectral CT were enrolled in this study. All images of cerebral angiography were reconstructed by software of spectral CT from 40 to 140 keV in 10 keV increments. The mean vessel attenuation and the contrast-to-noise ratio (CNR), signal-to-noise ra-tio (SNR) of the following arteries: basilar artery, internal carotid artery, anterior cerebral artery, middle cerebral artery and posterior cerebral artery were measured and calculated respectively. Subjective assessment on imaging quality was performed by using a 5-point scoring system. The results of objective image quality among different keV levels were compared using one-way analysis of variance. For subjective assessment, the Nonparametric Wilcoxon test was used to analyze the differences in subjective scores.Results: The image noise at 70-140 keV levels was statistically lower than that at 40, 50 and 60 keV levels. The image CNR and SNR between 60 keV and 70 keV levels showed on statistically differences, but markedly higher than other keV levels (P〈0.05). The scores of subjective image quality at 70 keV level were all higher than other keV levels including 60 keV level (P〈0.05).Conclusion: The 70 keV level can be commended as an optimal keV setting for spectral CT cerebral angiography by improving image CNR, SNR and reducing image noise.
出处 《温州医科大学学报》 CAS 2016年第8期600-603,共4页 Journal of Wenzhou Medical University
基金 温州市科技局科技计划项目(Y20130146 Y20140029)
关键词 体层摄影术 X线计算机 脑动脉造影术 单能量 图像质量 tomography,X-ray computed cerebral angiography single-energy image quality
  • 相关文献

参考文献10

  • 1刘义军,刘爱连,浦仁旺,刘晓锋,田士峰,李烨,张婷,刘磊.腹部能谱CT血管成像低浓度对比剂扫描方案的可行性[J].中华放射学杂志,2014,48(10):822-826. 被引量:21
  • 2BOGOT N R, FINGERLE A, SHAHAM D, et al. Imagequality of low-energy pulmonary CT angiography: comparisonwith standard CT[J]. AJR Am J Roentgenol, 2011,197(2): W273-278.
  • 3李娇娇,张璋,李东,王元全,于铁链.能谱CT冠状动脉造影单能量图像质量的优化[J].中国医学影像技术,2015,31(2):298-301. 被引量:12
  • 4WAAIJER A, PROKOP M, VELTHUIS B K, et al. Circleof Willis at CT angiography: dose reduction and image quality-reducing tube voltage and increasing tube currentsettings[J]. Radiology, 2007, 242(3):832-839.
  • 5唐坤,李瑞,林洁,郑祥武,曹国全.80kV低管电压扫描对脑CT血管成像图像质量及辐射剂量的影响[J].中国医学影像学杂志,2013,21(3):177-180. 被引量:23
  • 6CLOFT H J, JOSEPH G J, DION J E. Risk of cerebral angiographyin patients with subarachnoid hemorrhage, cerebralaneurysm, and arteriovenous malformation a meta-analysis[J]. Stroke, 1999, 30(2): 317-320.
  • 7KAZA R K, PLATT J F, COHAN R H, et al. Dual-energyCT with single-and dual-source scanners: current applicationsin evaluating the genitourinary tract[J]. Radiographics,2012, 32(2): 353-369.
  • 8LV P, LIN X Z, CHEN K, et al. Spectral CT in patientswith small HCC: investigation of image quality and diagnosticaccuracy[J]. Eur Radiol, 2012, 22(10): 2117-2124.
  • 9曹国全,唐坤,潘克华,邰云鹏,陈晓,吴恩福,王镇章.低管电压联合迭代重建技术在脑血管CT成像中的应用[J].温州医学院学报,2014,44(10):718-722. 被引量:7
  • 10ERTL-WAGNER B B, HOFFMANN R T, BRUNING R,et al. Multi-detector row CT angiography of the brain at variouskilovoltage settings[J]. Radiology, 2004, 231(2): 528-535.

二级参考文献45

  • 1Mettler FA Jr, Bhargavan M, Faulkner K, et al. Radiologic and nuclear medicine studies in the United States and worldwide: frequency, radiation dose, and comparison with other radiation sources-1950-2007. Radiology, 2009, 253(2): 520-531.
  • 2CEC. Quality criteria for computed tomography, European guidelines, EUR 16262. Luxembourg: Commiss Eur Comnlunit, 1999: 32.
  • 3Brenner D J, Hall EJ. Computed tomography-an increasing source of radiation exposure. N Engl J Med, 2007, 357(22): 2277-2284.
  • 4Waaijer A, Prokop M, Velthuis BK, et al. Circle of willis at CT angiography: dose reduction and image quality-reducing tube voltage and increasing tube current settings. Radiology, 2007, 242(3): 832-839.
  • 5Bahner ML, Bengel A, Brix G, et al. Improved vascular opacification in cerebral computed tomography angiography with 80 kVp. Invest Radiol, 2005, 40(4): 229-234.
  • 6Frush DE Frush KS. The ALARA concept in pediatric imaging: building bridges between radiology and emergency medicine: consensus conference on imaging safety and quality for children in the emergency setting, Feb. 23-24, 2008, Orlando, FL-executive summary. Pediatr Radiol, 2008, 38(Suppl 4):S629-S632.
  • 7Willis CE, Slovis TL.The ALARA concept in radiographic dose reduction. Radiol Technol, 2004, 76(2): 150-152.
  • 8Ertl-Wagner BB, Hoffmann RT, Bruning R, et al. Multi- detector row CT angiography of the brain at various kilovoltage settings. Radiology, 2004, 231(2): 528-535.
  • 9Huda W,Scalzetti EM,Levin G.Technique factors and image quality as functions of patient weight at abdominal CT[J].Radiology,2000,217:430-435.
  • 10Marin D,Nelson RC,Schindera ST,et al.Low-tube-voltage,high-tube-current multidetector abdominal CT:improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm--initial clinical experience[J].Radiology,2010,254:145-153.

共引文献59

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部