国产封堵器植入后行3.0T磁共振检查的安全性及对影像的影响 被引量:2

Safety of China-made occluders and the influence on MRI images under post-implantation 3.0T MR scan
下载PDF
导出
摘要 目的探讨几种常用国产封堵器植入后行3.0T MR检查时的安全性,及封堵器对MRI图像质量的影响。方法将封堵器悬挂于3.0T磁场中测量其偏转角度,再植入离体猪心内,行3.0T MR扫描,观察封堵器在扫描前后有无变形、移位,并分析图像、判断封堵器对图像质量有无影响。结果所有封堵器在磁场中均产生一定的偏转角度,均远<45°(P<0.01)。扫描前后,植入猪心的封堵器无变形或移位。房间隔缺损及动脉导管未闭封堵器盘片边缘轮廓显示较清晰,其轮廓内及周围可见伪影,且越靠近中心伪影越明显;室间隔缺损封堵器轮廓模糊,伪影范围明显大于其直径;总体上,较大的封堵器产生的伪影范围更大;同时植入3个封堵器,相互间图像无干扰。结论 3.0T MR常规扫描不会导致国产封堵器变形或移位,金属封堵器会对其自身及周围心脏结构的观察产生影响,但对心脏其他部位的图像质量无影响。 Objective To explore the safety of several common China-made occluders and their influence on the quality of MR image when the patient underwent a 3.0T MR scan. Methods The occluders were suspended in a magnetic field at a strength of 3.0T to measure their angle of deflection under such environment. 3.0T MR scans were performed on porcine hearts implanted with the occluders to examine if any deformation or displacement had occurred. Images were also analyzed to detect whether the occluders have affected their quality. Results All occluders in the magnetic field had an angle of de- flection less than 45° (P〈0.01). The occluders implanted into the porcine hearts were neither deformed nor displaced after scanning. The MRI images showed relatively clearer discs for both atrial septal defect (ASD) and patent ductus arteriosus (PDA) occ[uders, while the ventricular septal defect (VSD) occluder had an overa|[ vague contour. The ASD and PDA oc- cluders had artifacts that spreaded from the center, while the VSD occluder was covered by an artifact larger than the device itself. Occluders bigger in size generally produce more noticeable artifacts. The simultaneous implantation of three occlud- ers did not interfere with the image. Conclusion A 3.0T MR regular scan will not lead to the deformation or displacement of the occluders. While a metal occluder may affect the screening of itself and its adjacent cardiac structure, such influence is not present on the images of other parts of the heart.
出处 《中国医学影像技术》 CSCD 北大核心 2016年第7期1115-1118,共4页 Chinese Journal of Medical Imaging Technology
基金 福建省中青年教师教育科研项目(JA15206)
关键词 磁共振成像 先天性心脏病 国产封堵器 安全性 图像质量 Magnetic resonance imaging Congenital heart disease China-made occluders Security Image quality
  • 相关文献

参考文献13

  • 1Shellock FG, Morisoli SM. Ex vivo evaluation of ferromagnetism and artifacts of cardiac occluders exposed to a 1.5-T MR system. J Magn Reson Imaging, 1994,4(2):213-215.
  • 2胡海波,蒋世良,徐仲英,吴文辉,孟亮,陈雪生,曾庆玉.国产室间隔缺损封堵器磁共振检查的实验研究[J].中华放射学杂志,2005,39(7):751-754. 被引量:3
  • 3陈金明,王俭,邵敏伟,廖德宁.磁共振成像检查对先天性心脏病封堵器位置影响的实验研究[J].第二军医大学学报,2007,28(8):898-900. 被引量:3
  • 4ASTM. Standard test method for measurement of magnetically induced torque on medical devices in the magnetic resonance envi- ronment. 1-2015-07-11]. http://www, astm. org/Standards/ F2213. htm. 2011.
  • 5胡丽丽,陈昌胜,张冰,曾仁端.金属植入物在体模状况下的MRI研究[J].中国医学物理学杂志,2003,20(1):13-16. 被引量:13
  • 6李玉欣,王新良,周晓琳.人体内金属植入物MR检查的安全性[J].中国介入影像与治疗学,2009,6(5):485-488. 被引量:4
  • 7ASTM. Standard test method for measurement of magnetically induced displacement force on medical devices in the magnetic res- onance environment. [-2015-07-11]. http://www, astm. org/DA- TABASE. CART/HISTORICAL/F2052-06. htm, 2006.
  • 8Ganapathi M, Joseph J, Savage R, et al. MRI susceptibility artefacts related to scaphoid screws: Effect of screw type, screw orientation and imaging parameters. J Hand Surg Br, 2002,27 : 165-170.
  • 9Teitelbaum GP, Lin MC, Watanabe AT, et al. Ferromagnetism and MR imaging: Safety of carotid vascular clamps. AJNR Am J Neuroradiol, 1990,11(2) :267-272.
  • 10Olsrud J, Latt J, Brockstedt S, et al. Magnetic resonance ima- ging artifacts caused by aneurysm clips and shunt valves: De- pendence on field strength (1.5 and 3T) and imaging parame- ters. J Magn Reson Imaging, 2005,22(3):433-437.

二级参考文献77

共引文献46

同被引文献17

引证文献2

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部