期刊文献+

ADVIA2120i与XN-1000全自动血细胞分析仪结果对比研究 被引量:1

The contrastive study between ADVIA2120i and XN-1000 fully automatic blood cell analyzer
下载PDF
导出
摘要 目的:为探讨两种不同型号的ADVIA2120i和XN-1000全自动血细胞分析仪对同一批临床标本的检测结果的差异,比较两台仪器的检测性能,从而为临床检验工作提供指导性意义。方法:用ADVIA2120i和XN-1000两种仪器对从,临床上收集的标本做同种项目(全血细胞计数及白细胞分类计数)的检测,收集两种仪器实验所得的各项实验数据,并整理、比较、分析。结果:两台全自动血细胞分析仪在红细胞、白细胞、血红蛋白、红细胞压积、红细胞平均体积、平均血红蛋白浓度及血小板等检验项目上都有很好的一致性,两台仪器的批间、批内精密度及总精密度的变异系数均在各自的设定范围内;两台仪器在白细胞分类方面的检测结果除嗜碱性粒细胞外,其它各项分类差异无统计学意义。结论:ADVIA2120i和XN-1000全自动血细胞分析仪的检测性能良好,在做红细胞、血红蛋白、血小板等项目上,两台仪器可交叉使用,但是在做单核细胞、嗜碱性粒细胞项目上两台仪器的检测结果有差别,须与其他仪器或用人工镜检法加以佐证。 Objective : ADVIA2120i and XN 1000 fully automatic blood cell analyzer, Exploring the different types of test instruments for the same test result difference of clinical specimens, comparing the detection performance of two different type instrument and providing guiding for clinical inspection work. Methods : Using ADVIA2120i and XN-1000 fully automatic blood cell analyzer to do the same project for the clinical specimen .We collected two kinds of instruments and experiments of the experimental data to organiz, compar and analyze the experimental data collected.Results Two fully automatic blood cell analyzer in red blood cells, white blood cells, hemoglobin and platelet test projects had good consistency. Between the two instrument of batch, batch precision and variation coefficient of total precision are within the scope of their respective set. Two instrument of testing results and other types of blood cells analyzer compared to test results, in addition to the detection ofbasophils project has certain difference, the various test data correlation is well. Complete blood count and classification of leukocyte count are all accord with the requirement of laboratory. Corlclusi0rls : ADVIA2120i and XN-1000 fully automatic blood cell analyzer have better detection performance for red blood cells, hemoglobin and platelets detecting. But in basophils detection, the two instrument of testing results have certain difference and need to use artificial microscopy method and other instruments to evidence its degree of exactitude.
出处 《医学检验与临床》 2016年第3期40-43,共4页 Medical Laboratory Science and Clinics
关键词 ADVIA2120i全自动血细胞分析仪 XN-1000全自动血细胞分析仪 人工镜检 检测性能 ADVIA2120ifully automatic blood cell analyzer XN-1000 fully automatic blood cell analyzer manual microscopy Detection performance
  • 相关文献

参考文献4

二级参考文献4

  • 1肖秀林,艾红梅,李莉.Excel数据点折线图在血液学检验仪器间比对质量管理中的应用[J].现代检验医学杂志,2005,20(1):78-81. 被引量:7
  • 2张传宝,张克坚.方法比对及偏差评估的方法[M].临床实验室质量文件汇编,卫生部临床检验中心印发,2002,169-185.
  • 3许文融 谷俊侠.临床血液学检验[M].南京:东南大学出版社出版,..
  • 4叶应妩 王毓三 主编.全国临床检验操作规程[M].东南大学出版社,1991.136-136.

共引文献44

同被引文献4

引证文献1

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部