具有心房感知的右心室除颤电极导线临床应用体会 被引量:1

Clinical application of right ventricular defibrillator leads with floating atrial sensing bipolar electrodes
下载PDF
导出
摘要 目的探讨存在右心房漂浮感知双极线圈的右心室除颤电极导线(A+ICD电极)在单腔埋藏式心律转复除颤器(ICD)患者临床应用中的疗效及安全性。方法回顾性分析沈阳军区总医院心内科28例植入配备A+ICD电极(Linox Smart SDX65)的A+ICD心律转复除颤器(Lumax 540 VR-T DX)患者的临床资料,通过家庭监护(HM)长期观察右心房感知、右心室导线阈值、阻抗及感知参数;术后3个月诊室随访,观察卧位、坐位、立位时心房感知的变化,以及心房腔内图在单腔心律转复除颤器中的作用。结果本组28例患者手术均成功,平均随访(704.57±308.73)d。出院前及术后1、3、6、12个月的右心房感知振幅、右心室导线感知、阈值、阻抗间比较,差异均无统计学意义(P>0.05)。所有患者术后3个月卧位、坐位及立位的右心房感知振幅比较,差异均无统计学意义(P>0.05)。右心房腔内心电图早期发现房性心律失常并跟踪房性心律失常的管理,协助室性心动过速与室上性心动过速的鉴别诊断,优化房室顺序起搏。结论 A+ICD电极导线长期心房及心室参数稳定,通过A+ICD电极导线可获取稳定的心房感知、参与房室顺序起搏及房性心律失常的早期识别和管理。 Objective The study was performed to evaluate the clinical effect and safety of ventricular defibrillator lead(A + ICD lead) with floating atrial sensing bipolar electrodes in patients with implantable cardioverter defibrillator(ICD). Methods Totally 28 patients were implanted ICD( Biotronik Germany Lumax 540 VR-T DX) with A + ICD lead( Linox Smart SDX65). Long-term right atrial sensing,right ventricular threshold,lead impedance,sensing parameters were collected by home monitoring( HM) system and analyzed. Clinical follow-up was performed at 3 months after intervention,atrial sensing changes were observed in supine,sitting and standing position; and how atrial intracardiac electrocardiogram works in a single chamber implantable cardioverter defibrillator was observed. Results All 28 patients were implanted successfully,and the mean follow-up periods were( 704. 57 ± 308. 73) days. No significant changes were found in atrial sensing,ventricular sensing,threshold,as well as impedance before discharge,one month,three months,six months,and 12 months after intervention. Three months after 28 patients' intervention,no significant changes in atrial sensing were found while patients were in supine or sitting or standing position. Atrial sensing helped early identify and manage atrial arrhythmias,facilitate differential diagnosis of superventricular tachycardia and ventricular tachycardia,and optimize atrioventricular sequential pacing. Conclusion A + ICD lead could provide stable long-term atrial and ventricular lead parameters. A + ICD lead could help acquire stable atrial sensing,which leads to atrioventricular sequential pacing,and early identification and management of atrial arrhythmias.
出处 《临床军医杂志》 CAS 2016年第7期661-665,共5页 Clinical Journal of Medical Officers
基金 辽宁省科学技术计划面上项目(2015020418)
关键词 心律转复除颤器 心房感知 室性心动过速 Implantable cardioverter defibrillator Atrial sensing Ventricular tachycardia
  • 相关文献

参考文献10

  • 1Iori M,Giacopelli D,Quartieri F,et al.Implantable cardioverter defibrillator system with floating atrial sensing dipole:a singlecenter experience[J].Pacing Clin Electrophysiol,2014,37(10):1265-1273.
  • 2Toivonen L,Lommi J.Dependence of atrial sensing function on posture in a single-lead atrial triggered ventricular(VDD)pacemaker[J].Pacing Clin Electrophysiol,1996,19(3):309-313.
  • 3Sticherling C,Zabel M,Spencker S,et al.Comparison of a novel,single-lead atrial sensing system with a dual-chamber implantable cardioverter-defibrillator system in patients without antibradycardia pacing indications:results of a randomized study[J].Circ Arrhythm Electrophysiol,2011,4(1):56-63.
  • 4Prystowsky EN,Benson DW Jr,Fuster V,et al.Management of patients with atrial fibrillation.A Statement for Healthcare Professionals.From the Subcommittee on Electrocardiography and Electrophysiology,American Heart Association[J].Circulation,1996,93(6):1262-1277.
  • 5Hart RG,Halperin JL.Atrial fibrillation and thromboembolism:a decade of progress in stroke prevention[J].Ann Intern Med,1999,131(9):688-695.
  • 6The Antiarrhythmics versus Implantable Defibrillators(AVID)Investigators.A Comparison of Antiarrhythmic-Drug Therapy with Implantable Defibrillators in Patients Resuscitated from Near-Fatal Ventricular Arrhythmias[J].N Engl J Med,1997,337(22):1576-1583.
  • 7Kuck KH,Cappato R,Siebels J,et al.Randomized comparison of antiarrhythmic drug therapy with implantable defibrillators in patients resuscitated from cardiac arrest:the Cardiac Arrest Study Hamburg(CASH)[J].Circulation,2000,102(7):748-754.
  • 8Connolly SJ,Gent M,Roberts RS,et al.Canadian implantable defibrillator study(CIDS):a randomized trial of the implantable cardioverter defibrillator against amiodarone[J].Circulation,2000,101(11):1297-1302.
  • 9Daubert JP,Zareba W,Cannom DS,et al.Inappropriate implantable cardioverter-defibrillator shocks in MADIT II:frequency,mechanisms,predictors,and survival impact[J].J Am Coll Cardiol,2008,51(14):1357-1365.
  • 10于海波,梁延春,许国卿,刘荣,王祖禄,韩雅玲.植入型心律转复除颤器不适当电击的发生特点及优化管理[J].临床军医杂志,2015,43(10):991-993. 被引量:2

二级参考文献11

  • 1Levy WC,Lee KL,Hellkamp AS,et al.Maximizing survival benefit with primary prevention implantable cardioverter-defibrillator therapy in a heart failure population[J].Circulation,2009,120(10):835-842.
  • 2Moss AJ,Schuger C,Beck CA,et al.Reduction in inappropriate therapy and mortality through ICD programming[J].N Engl J Med,2012,367:2275-2283.
  • 3Gasparini M,Proclemer A,Klersy C,et al.Effect of long-detection interval vs standard-detection interval for implantable cardioverterdefibrillators on antitachycardia pacing andshock delivery:the ADVANCE III randomized clinical trial[J].JAMA,2013,309:1903-1911.
  • 4Ruwald AC,Sood N,Ruwald MH,et al.Frequency of inappropriate therapy in patients implanted with dual-versus single-chamber ICD devices in the ICD armof MADIT-CRT[J].J Cardiovasc Electrophysiol,2013,24:672-679.
  • 5Theuns DA,Rivero-Ayerza M,Boersma E,et al.Prevention of inappropriate therapy in implantable defibrillators:A meta-analysis of clinical trials comparingsingle-chamber and dual-chamber arrhythmia discrimination algorithms[J].Int J Cardiol,2008,125(3):352-357.
  • 6Wathen MS,De Groot PJ,Sweeney MO,et al.Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneousrapid ventricular tachycardia in patients with implantable cardioverter-defibrillators:Pacing Fast Ventricular Tachycardia Reduces Shock Therapies(Pain FREE RxⅡ)trial results[J].Circulation,2004,110(17):2591-2596.
  • 7Sweeney MO,Sherfesee L,De Groot PJ,et al.Differences in effects of electrical therapy type for ventricular arrhythmias on mortality in implantable cardioverter-defibrillator patients[J].Heart Rhythm,2010,7(3):353-360.
  • 8Daubert JP,Zareba W,Cannom DS,et al.Inappropriate implantable cardioverter-defibrillator shocks in MADIT II:frequency,mechanisms,predictors,and survival impact[J].J Am Coll Cardiol,2008,51(14):1357-1365.
  • 9van Rees JB,Borleffs CJ,de Bie MK,et al.Inappropriate implantable cardioverter-defibrillator shocks:incidence,predictors,and impact on mortality[J].J Am Coll Cardiol,2011,57(5):556-562.
  • 10Eckstein J,Koller M T,Zabel M,et al.Necessity for surgical revision of defibrillator leads implanted long-term:causes and management[J].Circulation,2008,117(21):2727-2733.

共引文献1

同被引文献2

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部