摘要 目的 研究利用调频超声激励微气泡群产生的次谐波进行血压测量的有效性。方法 选用参数相同的调频波和正弦波,通过模拟计算和实验研究比较两种激励下次谐波的幅度与激励声压之间的关系,分析次谐波幅度对过压变化的敏感度;同时探讨调频波的带宽和循环数对次谐波幅度所产生的影响。结果 调频波可以提高次谐波的幅度5~10dB,过压测量的敏感度提高3.4dB。结论适当地增大调频信号的带宽或循环数,可以提高次谐波对过压的敏感程度。 Objective To investigate the effectiveness of noninvasive blood pressure measurement by detecting the sub- harmonic response excited by chirp ultrasound scattered from microbubbles. Methods The dependence of sub-harmonic re- sponse respectively on exciting acoustic pressure and ambient pressure while utilizing chirp excitation was investigated, and comparisons were made between the results of chirp and traditional sinuous pulse. Then, in order to optimize the parame- ters of chirp signal, the effect of chirp bandwidth and pulse length on the variation sub-harmonic component were dis- cussed. Results As a superior excitation, chirp ultrasound greatly enhanced the amplitude of sub-harmonic component of 5--10 dB and its sensitivity of 3.4 dB to the ambient pressure. Conclusion In order to optimize the sensitivity of pressure measurement by chirp excitation, a wider frequency range and a longer pulse length can be utilized to increase the reduction when ambient pressure increases
出处 《中国介入影像与治疗学》 CSCD 北大核心 2016年第8期505-509,共5页 Chinese Journal of Interventional Imaging and Therapy
基金 蚌埠医学院自然科学基金重点项目(BYKY201410ZD)
关键词 超声检查 造影剂 血压 Ultrasonography Contrast media Blood pressure
  • 相关文献

参考文献20

  • 1Reddy AK, Taffet GE, Madala S, et al. Noninvasive blood pres- sure measurement in mice using pulsed Doppler ultrasound. Ul- trasound Med Biol, 2003,29(3) :379-385.
  • 2Strauss AL, Roth FJ, Rieger H. Noninvasive assessment of pres- sure gradients across iliac artery stenoses-duplex and catheter cor- relative study. J Ultrasound Med, 1993,12(1):17-22.
  • 3Goldberg BB, Liu JB, Forsberg F. Ultrasound contrast agents: A review. Ultrasound Med Biol, 1994,20(4):319-333.
  • 4Fairbank WM Jr, Seully MO. New noninvasive technique for car diac pressure measurement resonant scattering of ultrasound from bubbles. IEEE Trans Biomed Eng, 1977,24(2):107-110.
  • 5Bouakaz A, Frinking PJ, de Jong N, et al. Noninvasive measure- ment of the hydrostatic pressure in a fluid-filled cavity based on the disappearance time of micrometer-sized free gas bubbles. Ul- trasound Med Biol, 1999,25(9):1407 -1415.
  • 6Brayman AA, Azadniv M, Miller MW, et al. Effect of static pressure on' acoustic transmittance of AlbunexVR microbubhle suspensions. J Aeoust Soc Am, 1996,99(4 Pt 1):2403-2408.
  • 7H6k B. A new approach to noninvasive manometry: Interactionbetween ultrasound and bubbles. Med Bioi Eng Comput, 1981,19 (1) :35-39.
  • 8Shi WT, Forsberg F, Raichlen JS, et al. Pressure dependence of subharmonic signals from contrast microbubbles. Ultrasound Med Biol, 1992,25(2) :275-283.
  • 9Andersen KS, Jensen JA. Impact o acoustic pressure on ambient pressure estimation using ultrasound contrast agent. Ultrasonics, 2010,50(2) : 294-299.
  • 10Frinking PJ, Brochot J, Arditi M. Subha'rmonie scattering of phospholipid-shell mierobubbles at low acoustic pressure ampli- tudes. IEEE Trans Ultrason Ferroelectr Freq Control, 2010,57 (8) : 1762-1771.

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部