一种宽螺纹锥形种植体高扭矩植入后即刻负载的临床研究

The Immediate loading effect of a taper shape and wide screw thread implant with high torque insertion
原文传递
导出
摘要 目的评价一种宽螺纹锥形种植体高扭矩植入后进行即刻负载的临床效果。方法种植体植入后即刻负载,植入1月、2月、3月时对其周围X线、种植体稳定系数、种植体颈部骨吸收等情况进行评价。结果 1所有即刻负载的种植体在1月、2月、3月时均无松动,种植体周围无X线透影区;2种植体颈部边缘有骨吸收,骨吸收在1月和2月时发生率较高,其中,单冠修复发生率多于固定桥修复(P<0.05);3种植体稳定系数在刚植入时较高、1月时最低、随后又升高,单冠及固定桥修复之间无明显差异(P>0.05)。结论此种采用高扭矩植入的宽螺纹锥形种植体,进行即刻负载有较好的临床效果,但是其长期成功率还需进一步观察。 Objective To evaluate the immediate loading effect of a taper and wide screw thread implant after insertion with high torque. Methods The apical X-rays,implants stability quotient(ISQ) and bone absorption in cervical sections were examined after 1 month,2 months and 3 months of immediate loading. Results All implants were stable without peri-implant X-ray defect. Bone absorption rate in cervical area happened more in the first 2 months than in the three months', and which was more in single crown restoration than that in fix restorations(P0.05). ISQ decreased at 1 month of loading,and increased at the following second month time. ISQ were no significant difference in single crown and fix restorations(P 0.05). Conclusion Immediate loading of this taper shape and wide screw thread implant with high insertion torque could result in advanced clinical effects. But farther studies are still needed to evaluate the long term effects.
出处 《现代口腔医学杂志》 CAS CSCD 2016年第4期225-227,共3页 Journal of Modern Stomatology
基金 山西省自然科学基金资助(310068/C10112)
关键词 牙种植 即刻负载 宽螺纹种植体 高扭矩植入 Dental implant Immediate loading Wide screw thread High insertion torque
  • 相关文献

参考文献12

  • 1Grunder U, Polizzi G, Goene R, et al. A 3-year prospective muhicenler follow-up report on the immediale and delayed- immediate placement of implants. Int J Oral Maxillofae Implants. 1999, 14(2): 210-216.
  • 2Jaffin KA, Kumar A, Bermar CL. Immediate loading of implants in partially and fully edentulous jaws: a series of 27 ease reports. J Periodontol, 2000, 71(5): 833-838.
  • 3Rabel A, Kohler SG, Schmidt-Westhausen AM. Clinical study on the primary stability of two dental implant systems with resonance frequency analysis. Clin oral Investig, 2007, 11 (3): 257-265.
  • 4Irinakis T, Wiebe C. initial torque stability of a new bone condensing dental implant. A cohort study of 140 consecutively placed implants. J of Oral hnplantol, 2009, 35(6): 277-282.
  • 5Ueda M, Matsuki M, Jaeobsson M, et al. Relationship between insertion torque and removal torque analyzed in fresh temporal bone. Int J Oral Maxillofac Implants, 1991, 6(4): 442-447.
  • 6Gehrke SA, da Silva UT, Del Fabbro M. Does implant design affect implant prinmry stability? A resonance frequency analysis-based randomized split-mouth clinical trial. J Oral Implantol, 2015, 41(6): e 281-286.
  • 7Ottoni JM, Oliveira ZF, Mansini R, et al. Co~Telation between placement torque and survival of single-tooth implants. Int J Oral Maxillofac Implants, 2005, 20(5): 769-776.
  • 8O'Sullivan D, Sennerby L, Meredith N. Influence of implant taper on the primary and secondary stability of osseointegrated titanium implants. Clin Oral Implants Res, 2004, 15(4): 474- 480.
  • 9刘宝林主编.口腔种植学.第1版.北京:人民卫生出版社,2011:21,90.
  • 10杨德圣,刘洪臣,顾晓明,魏务建,李华.牙种植体挤压植入后早期骨界面的组织学变化[J].中华老年口腔医学杂志,2005,3(2):113-114. 被引量:7

二级参考文献22

  • 1汪昆,李德华,于敬宇,李玉龙,刘宝林.双皮层骨种植影响牙种植体稳定性的有限元固有频率分析[J].实用口腔医学杂志,2005,21(3):356-358. 被引量:3
  • 2[1]Cochran DL,Buser D,ten Bruggenkate CM,et al.The use of reduced healing times on ITI implants with a sand blasted and acid-etched (SLA) surface:early results from clinical trialson ITI SLA implants[J].Clin Oral Impl Res,2002,13:144~153
  • 3[2]Meredith N.Assessment of implant stability as a prognostic determinant[J].International journal of Prosthodontics,1998,11:491~501
  • 4[3]Bahr W.The effects of pre-tapping on the miniscrew-bone interface in the midface [J].J Craniomaxillofac Surg,1989,17:337
  • 5[4]Berglundh T,Abrahamsson I,Lang N.P,et al.De novo alveolar bone formation adjacent to endosseous implants:A model study in the dog[J].Clin Oral Impl Res,2003,14(20):251~262
  • 6[5]Moller W,Lowicke G,Naumann H.Reconstruction of the pars alveolaris with mouded and compressed spongy substance [J].Zahn-,Mund-Kieferheilkunde, 1985,73:464~470
  • 7[6]Nkenke E,Kloss F,Wiltfang J,et al. Histomorphometric and fluorescencemicroscopic analysis of bone remodelling af ter installation of implants using an osteotome technique[J].Clin Oral Impl Res,2002,13:595~602
  • 8Bischof M,Nedir R,Szmukler-Moncler S,et al.Implant stability measurement of delayed and immediately loaded implants during healing.Clin Oral Implants Res,2004,15(5):529
  • 9Lachmann S,Jager B,Axmann D,et al.Resonance frequency analysis and damping capacity assessment.Part I:An in vitro study on measurement reliability and a method of comparison in the determination of primary dental implant stability.Clin Oral Implants Res,2006,17(1):75
  • 10Lachmann S,Laval JY,Jager B,et al.Resonance frequency analysis and damping capacity assessment.Part 2:peri-implant bone loss follow-up.An in vitro study with the Periotest and Osstell instruments.Clin Oral Implants Res,2006,17(1):80

共引文献36

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部