期刊文献+

分次团注结合双能量扫描在CTU检查中的应用价值 被引量:6

Application Value of Split-bolus Injection Combined with Dual Energy CT Scan in CT Urography
下载PDF
导出
摘要 目的探讨分次团注双能量扫描在降低CTU检查辐射剂量的应用价值。方法 50例行CTU检查的患者平均分两组,其中试验组(A组)25例,行两次团注对比剂后实质期常规扫描和排泄期双能量扫描;对照组(B组)25例,行全泌尿系平扫及单次团注对比剂后实质期和肾孟期常规扫描。A组的数据在工作站上使用双能量后处理软件获得排泄期的最佳ke V图像,将图像的集合系统分段,测量各段管腔内的CT值.并计算对比噪声比(contrast to noise ratio,CNR)。B组排泄期图像发送至同一工作站进行相同的测量和CNR计算,同时记录所有病例完成全部检查的有效辐射量((effetive radiation dose,ERD)),比较两组CT值、CNR、ERD的差别。结果分次团注排泄期最佳ke V图像集合系统的均匀度提高,各段CT值均高于常规图像,其差异有统计学意义(P<0.05);与常规排泄期集合系统各段CNR没有统计学意义(P>0.05);分次团注双能量方法完成全部检查的ERD(平均12.56 m Sv)明显低于常规CTU(平均18.27m Sv),差异有统计学意义(P<0.05)。并能利用双能量软件中的其他功能为诊断提供更多的信息。结论分次团注结合双能量扫描能够有效降低CTU的辐射剂量,提高图像质量。 Objective To discuss the application value of split-bolus injection combined with dual energy CT scan in CT urography. Met.hods Fifty patients underwent CT urography were randomly and averagely divided into two groups. Twenty-five cases of the experimental group(A group)were applied with conventional scan in parenchymal phase and dual energy scan in excretory phase after split-bolus injection. Other 25 cases of the control group(B group) underwent plain scan in total urinary and conventional scan in parenchymal and excretory phase after single-bolus injection. The keV images of Group Ns excretory phase data were acquired by using dual energy software. The images were segmented by urinary collection system and CT values of each lumen were measured, and contract to noise ratio (CNR) was calculated. Images of Group B's excretory phase were sent to the same workstation to do the same measurement and CNR calculation. At the same time, the amount of effective radiation in a complete inspection was recorded for all 50 cases. The CT value, CNR and the amount of effective radiation of two groups were compared. Results The uniformity of excretory phase's keV images of split-bolus injection improves compares to that of conventional method. CT value is significantly higher(P〈0.05) while CNR is not significantly different(P〉0.05).The effective amount of radiation in a complete inspection(12.56mSv in average)for split-bolus method is lower than conventional one(18.27mSv in average) and it is significantly different(P〈0.05). In addition, split-bolus method can also provide more information for diagnosis by using other features in dual energy software. ConN,'on Split-bolus injection combined with dual energy scan can effectively reduce the radiation dose of CTU, improve the quality of images.
出处 《中国CT和MRI杂志》 2016年第9期110-113,共4页 Chinese Journal of CT and MRI
关键词 X线体层摄影术 泌尿系成像 排泄期 双能量成像 X-ray Tomography Urography Excretion Dual Energy Imaging
  • 相关文献

参考文献13

  • 1Dillman JR, Caoili EM, Cohan RH. Nulti-detector CT urography: a one-stop renal and urinary tract imaging modality[J].Abdom Imaging, 2007,32(4):519-529.
  • 2周硕,林美福,陈彩龙,周庆伟,陈国宝,王尔祯.MSCTU诊断泌尿系统先天性疾病的价值:与IVU、MRU对比研究[J].中国CT和MRI杂志,2008,6(5):38-41. 被引量:13
  • 3Yanaga Y, Awai K, Funama Y, et al. Low-dose MDCT urography: feasibility study of low-tube- voltage technique and adaptive noise reduction filter[J]. Am J Roentgenol, 2009,193(3):220- 229.
  • 4Portnoy O, 6uranda L, apter S, et al. Optimization of 64-MDCT urography: effect of dual-phase imaging with furosemide on collecting system opacification and radiation dose[J]. Am J Roent-genol, 2011,197(5):882- 886.
  • 5刘畅,刘斌,王万勤.分次团注对比剂法CTU在泌尿系统检查中的应用价值[J].中国CT和MRI杂志,2011,9(4):44-46. 被引量:9
  • 6胡娟,王鹤,王霄英,杨学东,邱健星.双能CT成像在分次团注双期CTU中的临床应用[J].放射学实践,2012,27(3):259-263. 被引量:12
  • 7帅桃,李真林,阳琴,赵利娜,袁元,陈国勇,张凯.能谱纯化技术在泌尿系虚拟平扫成像中的应用价值[J].放射学实践,2013,28(12):1196-1199. 被引量:7
  • 8Mangold S, Thomas C, Fenchel M, et al. Virtual nonenhanced dual-energy CT urography with tin-filter technology:determinants of detection of urinary calculi in the renal collecting system[J]. Radiology, 2012,264(1):119-125.
  • 9Mafuji A, Hara M. Sasaki S, et al. Usefulness of dual- energy CT scanning at 80 kVp for identifying hilar and mediastinal structures: evaluation of contrast enhancement of the pulmonary vessels and lymph nodes[J].Jpn J Radiol, 2012,30(1):69-77.
  • 10Chandarana H, Godoy MC, Vlahos I, et al. Abdominal aorta: evaluation with dual-source dual-energy multidetector CT after endovascular repair of aneurysms- initial observations[J]. Radiology, 2008,249(2):692-700.

二级参考文献36

  • 1刘勇,张华.ALARA理论及CT低剂量应用研究进展[J].中国医学影像技术,2008,24(S1):17-20. 被引量:28
  • 2McTavish JD, Jinzaki M, Zou KH, et al. Multi-detector row CT urography : comparison of strategies for depicting the normal urina ry collecting system, Radiology, 2002,225(3) : 783-790.
  • 3Caoili EM, Cohan RH, Korobkin M, et al. Urinary tract abnormalities: initial experience with multi-detector row CT urography. Radiology, 2002,222(2) :353-360.
  • 4Nolte-Ernsting CC, Wildberger JE, Borders H, et al. Multi-slice CT urography after diuretic injection: initial results. Rofo, 2001, 173(3) :176- 180.
  • 5Tsili AC, Hremedis SC, Kalef-Ezra J, et al. Muhidetectovrow CT urography on a 16 row CT scanner in the evaluation of urothelial tumors. Eur Radiol, 2007,17(4) : 1046-1054.
  • 6Morcos SK. Computed tomography urography: technique. indications and limitations. Curt ()pin Urology, 2007,17(1):56-64.
  • 7CEC, Quality criteria for computed tomograghy, European Guoidelines. Commission of the European Communities, 1999.
  • 8UNSCEAR. United Nations Scientific Committee on the effects of atomic radiation UNSCEAR 2000 report to the general assembly, with Scientific annexes. United Nations, 2000.
  • 9White KS. Helical/spiral CT scanning: a pediatric radiology perspeetive. Pediatr Radiol, 1996,26 ( 1 ) : 5-14.
  • 10Chow LC, Kwan SW, Olcott EW, et al. Split bolus MDCT urog raphy with synchronous nephrographic and excretory phase enhancement. AJR AmJ Roentgenol, 2007,189(2) :314-322.

共引文献50

同被引文献34

引证文献6

二级引证文献19

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部