PET/CT成像伪影分析 被引量:4

ANALYSIS OF ARTIFACT IN PET/CT IMAGINg
原文传递
导出
摘要 目的分析PET/CT图像伪影产生的原因,探讨其可能预防的措施。方法回顾性分析了南京总医院核医学科1 100例全身PET/CT图像,分析伪影对图像质量的影响程度、伪影的类型、常见及不常见的生理摄取。结果在1 100例患者中,有117例存在伪影,其中33例(28.2%)伪影对临床诊断有影响。依次为呼吸运动伪影50例(42.7%);注射点及放射性污染伪影27例(23.1%);高密度物质伪影19例(16.2%);高血糖肌肉摄取伪影13例(11.1%);体位移动伪影6例(5.1%);口服造影剂伪影2例(1.7%)。结论正确认识伪影并加以消除,有利于提高图像质量,提高影像诊断准确性。 Objective To analyze the causes of PET/CT image artifacts and to discuss the possible preven- tive measures. Methods A retrospective analysis was performed on 1 100 cases who received body PET/ CT image in auhtor's hospital. The artifact types, common, uncommon physiological uptake and the degree of influence on image quality were analyzed. Results There were 117 of 1 100 patients presented artifacts, 33 (28. 2%) of them affected clinical diagnostic. Followed by respiratory motion artifacts in 50 cases (42.7%) ; injection site and radioactive contamination artifact were 27 cases (23.1%) ; high density mate- rial artifacts in 19 cases (16.2%); hyperglycemia muscle uptake artifacts 13 cases (11.1%); postural mo- tion artifacts 6 cases (5.1% ) ; oral contrast agent artifacts 2 cases ( 1.7% ). Conclusion Correct under- standing of artifacts and to eliminate them, will help improve the image quality, improve the accuracy of imaging diagnosis.
出处 《中国煤炭工业医学杂志》 2016年第8期1132-1135,共4页 Chinese Journal of Coal Industry Medicine
基金 南京军区南京总医院课题(编号:2014054)
关键词 PET/CT图像 伪影 PET/CT image Artifact
  • 相关文献

参考文献13

  • 1Hany TF, Steinert HC, GoelTes GW, et al. PET di- agnostic accuracy: improvement with in- line PET- CT system- initial results[J]. Radiology, 2002, 225 (2) : 575- 581.
  • 2许全盛,袁克虹,于丽娟,王文志,叶大田.PET/CT图像呼吸运动伪影校正研究进展[J].中国生物医学工程学报,2009,28(4):573-580. 被引量:22
  • 3Goerres GW, Kamel E, Seifert B, et al. Accuracy of image coregistration of pulmonary lesions in patients with non-small cell lung cancer using an integrated PET/CT system[J]. Nucl Med, 2002,43 ( 11 ) .. 1469 - 1475.
  • 4Osman MM, Cohade C, Nakamoto Y, et al. Clinically significant inaccurate localization of lesions with PET/CT:frequency in 300 patients[J]. J Nuel Med, 2003,44 (2) : 240 - 243.
  • 5Kamel EM,Burger C,Buck A,et al. Impact of metal- lic dental implants on CT- based attenuation correc- tion in a combined PET/CT acanner [J]. Eur Radio], 2003,13(4) 724- 728.
  • 6Vande Berg B, Malghem J, Maldague B, et al. Multi- detector CT imaging in thePostoperative orthopedic patient with metal hardware[J]. Eur J Radiol,2006. 60(3) 470- 479.
  • 7陈泽龙,谢康宁,漆家学,钱根年,赵春雷,陈杰.PET/CT图像基于CT衰减校正伪影的实验研究[J].医疗卫生装备,2014,35(3):81-84. 被引量:5
  • 8程娟.整体护理在PET-CT检查病人中的应用[J].全科护理,2012,10(15):1352-1353. 被引量:4
  • 9付占立,王荣福.能量底物环境对^(18)F-FDG显像的影响[J].国外医学(放射医学核医学分册),2000,24(2):55-58. 被引量:34
  • 10王进东,孙静涛,戴秀华,李为,张再兴.PET/CT和MRI在鼻咽癌早期诊断中的价值[J].中国煤炭工业医学杂志,2012,15(12):1845-1846. 被引量:4

二级参考文献92

  • 1张红梅,周纯武.PET/CT的特点及其在肿瘤学中的临床应用[J].中国医学影像技术,2005,21(8):1296-1298. 被引量:4
  • 2陈香,赵军,管一晖.PET-CT常见伪影[J].国外医学(放射医学核医学分册),2005,29(5):201-205. 被引量:8
  • 3李天然,陈自谦.功能影像设备PET-CT[J].医疗卫生装备,2005,26(12):27-30. 被引量:9
  • 4Pan Tinsu, Mawlawi O, Nehmeh S, et al. Attenuation correction of PET images with respiration-averaged CT image in PET/CT [ J ]. Nuclear Medicine, 2005, 36(6) : 1481 - 1487.
  • 5王学明,沈克涵.医学成像系统[M].北京:清华大学出版社.2006.
  • 6Daou D. Respiratory motion handling is mandatory to accomplish the high-resolution PET destiny [J]. European Journal of Nuclear Med Mol Imaging, 2008, 35:1961- 1970.
  • 7Visvikis D, Lamare F, Bruyant P, et al. Respiratory motion in positron emission tomography for oncology applications: Problems and solutions [ J]. Nuclear Instruments and Methods in Physics Research, 2006, A569: 453- 457.
  • 8Chuang Keh-Shih, Chen Tzong-Jer, Chang Chia-Chi, et al. Reduction of motion artifacts for PET imaging by respiratory correlated dynamic scanning [J]. Nuclear Instruments and Methods in Physics Research, 2006, A569: 458- 461.
  • 9Zhu Zhiyu, Tsui BMW, Segars WP. A Simulation Study of the Effect of Gating Scheme on respiratory Motion Blurting in FDG Lung PET [J]. IEEE Transactions on Nuclear Science, 2003, 13(3): 1554 - 1558.
  • 10Segars WP. Development of a new dynamic NURBS-based cardiactorso (NCAT) phantom [ D]. North Carolina: University of North Carolina, 2001.

共引文献69

同被引文献34

引证文献4

二级引证文献15

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部