不同材料烤瓷冠与金属冠磁共振成像伪影的比较 被引量:5

Comparison of artifacts on magnetic resonance imaging caused by different porcelain-fused-to-metal crowns and metal crowns
下载PDF
导出
摘要 目的分析不同材料烤瓷冠和铸造冠在磁共振成像(MRI)图像上产生伪影的大小。方法在临床随机选取的右侧上颌第二磨牙模型制作钴铬烤瓷冠、钴铬金属冠、镍铬烤瓷冠、镍铬金属冠各一个,并以丙烯酸树脂冠作为对照,采用1.5 T MRI扫描仪对5个牙冠进行扫描,观察不同材料牙冠成像伪影的大小。结果同一序列中不同材料试件所产生伪影面积的差异均有统计学意义,不同材料试件两两比较,其差异也有统计学意义。钴铬金属冠产生的伪影面积最大,镍铬烤瓷冠产生的伪影面积最小。结论 4种牙冠对MRI的影响不同,口腔金属修复体产生的伪影与材料类型密切相关。 Objective This study aimed to investigate the artifacts of different porcelain-fused-to-metal crowns and metal crowns on magnetic resonance imaging(MRI). Methods One Co-Cr porcelain crown, one Co-Cr metal crown, one Ni-Cr porcelain crown, and one Ni-Cr metal crown were placed in a cylindrical plastic vessel filled with distilled water. An arcrylic resin crown was used as control and then scanned using roution head sequence, including FRFSET2WI, GRET2*WI, and T1FLAIR, on a 1.5 T scanner. The MRI artifact areasof different porcelain-fused-to-metal crowns and metal crowns were calculated and analyzed with variance test. Results ANOVA of factorial design indicated significant differences among the crowns. Co-Cr metal crown produced the largest artifact, whereas Ni-Cr porcelain crown produced the least. Conclusion Various crowns differently affected MRI images. The artifacts produced by the crownswere closely related to the type of material.
作者 秦琨 张东坡
出处 《国际口腔医学杂志》 CAS 北大核心 2016年第5期507-510,共4页 International Journal of Stomatology
关键词 磁共振成像 伪影 镍铬合金 钴铬合金 magnetic resonance imaging artifact Ni-Co alloy Co-cr alloy
  • 相关文献

参考文献11

  • 1Shafiei F, Honda E, Takahashi H, et al. Artifacts from dental casting alloys in magnetic resonance imaging[J]. J Dent Res, 2003, 82(8):602-606.
  • 2Eggers G, Rieker M, Kress B, et al. Artefacts in ma- gnetic resonance imaging caused by dental material [J]. MAGMA, 2005, 18(2): 103-111.
  • 3胡艳艳,魏斌.固定修复体引起磁共振伪影的研究进展[J].上海口腔医学,2008,17(4):441-443. 被引量:2
  • 4Viano AM, Gronemeyer SA, Haliloglu M, et al. Improved MR imaging for patients with metallic implants[J]. Magn Reson Imaging, 2000, 18(3):287- 295.
  • 5Thomsen M, Schneider U, Breusch S J, et al. Arte- facts and ferromagnetism dependent on different metal alloys in magnetic resonance imaging. An experimental study[J]. Orthopade, 2001, 30(8):540- 544.
  • 6Fache JS, Price C, Hawbolt EB, et al. MR imaging artifacts produced by dental materials[J]. AJNR Am J Neuroradiol, 1987, 8(5):837-840.
  • 7Behr M, Fellner C, Bayreuther G, et al. MR-imaging of the TMJ: artefacts caused by dental alloys[J]. Eur J Prosthodont Restor Dent, 1996, 4(3):111-115.
  • 8Suh JS, Jeong EK, Shin KH, et al. Minimizing arti- facts caused by metallic implants at MR imaging: experimental and clinical studies[J]. A JR Am J Roentgenol, 1998, 171(5):1207-1213.
  • 9黄海蓉,李蓉,郦妙尔,石卓瑾.钴铬合金磁共振成像伪影范围的研究[J].口腔医学研究,2011,27(9):778-780. 被引量:11
  • 10关小菊,刘娟,李长福.口腔不同材料烤瓷冠磁共振成像伪影的实验研究[J].中华老年口腔医学杂志,2010,8(2):107-109. 被引量:11

二级参考文献40

  • 1赖寿伟,杨华岳.金属固定假牙对头部MR图像的影响研究[J].实用医技杂志,2005,12(02A):350-352. 被引量:11
  • 2刘玉华,孙樱琳.固定义齿修复材料对MRI图像的影响[J].现代口腔医学杂志,2005,19(6):588-589. 被引量:19
  • 3戴永雨,张晶,高培毅,何志华.牙科金属对头部磁共振成像的影响[J].中华放射学杂志,1996,30(6):409-409. 被引量:4
  • 4孙樱林,刘玉华.镍铬合金固定修复体对MRI影像的影响[J].现代口腔医学杂志,2007,21(1):72-74. 被引量:17
  • 5Harris T,Farid rad M,Dickon J.The benefits of aesthetic orthodontic brackets in patients requiting multiple MRI scanning[J].Orthdontics,2006,33(2):90-93.
  • 6Shafiei F,Honda E,Takahashi H,et al.Artifacts from dental casting alloys in magnetic ressonance imagingLJ].J Dent Res,2003,82(8):602-604.
  • 7T.M.J.Harris,M.R.Faridrad,J.A.S.Dickson.The Benefits of Aesthetic Orthodontic Brackets in Patients Requiring Multiple MRI Scanning[J].Journal of Orthodontics,2006,33:90-94.
  • 8Mi-Fung Lee,Sungjun Kim,Sung-Ah Lee,et al.Overcoming Artifacts from Metallic Orthopedic Implants at High-Field-Strength MR Imaging and Multidetedtor CT[J].RG,2007,27(3):791-802.
  • 9HidekiMATSUURA,TakashiINOUE,KuniakiOGASAWARA,et al.Quantitative Analysis of Magnetic Resonance Imaging Susceptibility Artifacts Caused by Neurosurgical Biomaterials:Comparison of 0.5,1.5,and 3.0 Tesla Magnetic Fields[J].Neurol Med Chir,2005,45:395-399.
  • 10Eggers G,Rieker M,Kress B.et al.Artefacts in magnetic resonance imaging caused by dental material[J].MAGMA,2005,18(2):103-111.

共引文献29

同被引文献39

引证文献5

二级引证文献8

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部