期刊文献+

拉伸试验应注意的几个问题 被引量:3

Some experiences in tensile testing
原文传递
导出
摘要 通过长期拉伸试验研究,发现软件系统自动给出的R_(eL)、R_(eH)、R_(p0.2)、A_(gt)值不一定准确,应根据拉伸曲线对试验结果进行验证,如果发现问题应进行修正;还应对试验机设置的试验速度是否满足GB/T 228.1—2010中应变速率控制要求进行验证;试样的加工精度不满足标准要求时,应采取相应的解决措施。 According to the experiences in tensile testing,it was found that the values of ReL,ReH,Rp0.2 and Agtwhich were automatically provided by software system may be inaccurate. Therefore,the testing results should be verified based on the tensile curve. If there was any problem,the results should be corrected. Moreover,the testing speed setting of tester should meet the control requirements of strain rate in GB/T 228.1—2010. When the machining precision of sample cannot satisfy the standard requirements,the corresponding measures should be taken.
出处 《物理测试》 CAS 2016年第4期22-24,共3页 Physics Examination and Testing
关键词 试验结果 试验速度 加工精度 testing results testing speed machining precision
  • 相关文献

参考文献4

  • 1卢国清,齐彪,陈德,吴伟,白银兰,张磊.拉伸试验速率对测试结果的影响[J].物理测试,2015,33(5):1-5. 被引量:9
  • 2于兆斌,吴倩颖,杨雷岗,杨新生,祝铁柱.材料试验机针对拉伸试验新标准的改进[J].物理测试,2014,32(2):54-56. 被引量:3
  • 3中华人民共和国国家质量监督检验检疫总局,中国国家标准化管理委员会.GB/T228.1-2010《金属材料拉伸试验室温拉伸方法》[S].北京:中国标准出版社,2010.
  • 4中华人民共和国国家质量监督检验检疫总局,中国国家标准化管理委员会.GB/T228-2002《金属材料室温拉伸试验方法》[S].北京:中国标准出版社,2002.