溶血和乳糜微粒对ALT检测的影响 被引量:2

Effect of hemolysis and chylomicron on ALT detection
下载PDF
导出
摘要 目的探讨溶血和乳糜微粒对ALT检测的影响。方法根据中华人民共和国卫生行业标准(WS/T)和美国临床和实验室标准研究院(CLSI)相关文件,进行干扰实验验证。结果干扰实验结果显示,样本在反复冻融1次以上和震荡2 min以上时,以及外源性干扰物血红蛋白(HB)>2.5 g/L时ALT检测均产生正干扰,乳糜微粒>5 g/L对ALT检测有明显负干扰。结论血红蛋白产生的是正干扰,乳糜微粒产生的是负干扰,通过评估溶血和脂血标本的干扰方向和干扰偏差的大小,对如何处理生物参考区间上限附近的标本具有指导作用。 Objective To investigate the effect of hemolysis and chylomicron on ALT detection. Methods Interference tests were conducted on the relevant documents from the health industry standards( WS / T) in the People’s Republic of China and the Clinical and Laboratory Standards Institute( CLSI). Results Interference tests showed that when samples were frozen and thawed for at least 1 time and shocked for more than 2 minutes,and when exogenous HB 〉 2. 5 g / L,positive interference was imparted on ALT detection. However,when chylomicron 〉 5 g / L,a significant negative interference was produced. Conclusion Hemoglobin makes positive interference and chylomicrons make negative interference. By evaluating the direction of interference and the size of the interference bias of the hemolysis and lipid blood sample,guidance is provided on dealing with the sample near the upper limit of the biological reference interval.
机构地区 辽宁省血液中心
出处 《中国输血杂志》 CAS 北大核心 2016年第7期758-760,共3页 Chinese Journal of Blood Transfusion
关键词 血红蛋白 乳糜微粒 干扰实验 hemoglobin chylomicron interference testing
  • 相关文献

参考文献10

二级参考文献56

  • 1林莉,石文,彭玉莲,黄宪章.罗氏COBAS 400全自动生化分析仪检测系统性能验证试验[J].现代医学仪器与应用,2007,0(3):50-53. 被引量:7
  • 2郝冬连.溶血对生化检验准确性的影响及纠正探析[J].求医问药(下半月),2013(2):631-631. 被引量:6
  • 3程正江.血红蛋白和胆红素干扰临床化学分析的机理初探[J].国外医学(临床生物化学与检验学分册),2004,25(6):488-491. 被引量:27
  • 4邱玲,程歆琦,鄢盛恺,沈瑛,刘茜,刘荔,刘福成,葛增梅.临床化学方法验证与表格化管理[J].现代检验医学杂志,2006,21(4):5-9. 被引量:10
  • 5杨有业,张秀明.临床检验方法学评价[M].北京:人民卫生出版社,2008:118-127.
  • 6Simonsen O,Grubb A,Thysell H.The blood serum concentration of cystatin C (γ-trace) as a measure of glomerular filtration rate.Scand J Clin Lab Invest,1985,45:97-101.
  • 7White C,Akbari A,Hussain N,et al.Estimating glomerular filtration rate in kidney transplantation:a comparison between serum creatinine and cystatin C-based methods.J Am Soc Nephrol,2005,16:3763-3770.
  • 8Perkins BA,Nelson RG,Ostrander BE,et al.Detection of renal function decline in patients with diabetes and normal or elevated GFR by serial measurements of serum cystatin C concentration:results of a 4-year follow-up study.J Am Soc Nephrol,2005,16:1404-1412.
  • 9Hoek FJ,Kemperman FAW,Krediet RT.A comparison between cystatin C,plasma creatinine and the Cockcroft and Gault formula for the estimation of glomerular filtration rate.Nephrol Dial Transplant,2003,18:2024-2031.
  • 10Tan GD,Lewis AV,James TJ,et al.Clinical usefulness of cystatin C for the estimation of glomerular filtration rate in type 1 diabetes.Diabetes Care,2002,25:2004-2009.

共引文献66

同被引文献15

引证文献2

二级引证文献9

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部