摘要
以微制程技术及电化学法制备PANI/Au/Al2O3电极,并组装成电阻式丙酮气体传感器。以恒电流法在不同电流密度下进行三阶段聚合,并利用混合物实验设计法对纳米结构PANI/Au/Al2O3电极的最佳制备工艺条件进行了优化。结果表明:分别固定聚合电流密度为40,60,80p.A/cm2进行三阶段聚合,当聚合电流密度按由小到大顺序,所得的PANI/Au/Al2O3电极稳定性最好。固定聚合电量为35mC,当三阶段电量比为0.35:0.33:0.32时,对于9958×10-6丙酮得到最大的感测电阻变化率为7.55%。丙酮气体浓度在(475—1189)×10-6与(23~118)×10-6范围内,对应灵敏度分别为1.2×10-3和6.4×10-3%/10-6,平均响应时间为3min。
In order to obtain a better performance of resistive acetone gas sensor, based on micro process technology and electrochemical methods, PANI/Au/Al2O3 electrode is prepared and assembled. Three-phase polymerization is carried out with different current density. Use mixture experimental design method, preparation process of nanostructure of PANI/Au/Al2O3 electrode is optimized. Results show that the stability of PANI/Au/ Al2O3 electrode is the best when polymerization current density is fixed at 40,60 and 80μA/cm2, respectively, and the aggregate current density change from small to large. Maintaining polymerization charge at 35 mC, as three- phase charge ratio is 0.35 : 0. 33 : 0.32, the maximum sensor resistance rate of the 9 958× 10-6 acetone is 7.55 %. As the acetone gas concentration in the range of (475 - 1 189 ) ×10-6 and (23-118 ) × 10-6, corresponding sensitivities are 1.2 × 10 -3 and 6.4 × 10 -3 %/10 -6, and the average response time is 3 min.
出处
《传感器与微系统》
CSCD
2016年第9期66-68,共3页
Transducer and Microsystem Technologies
基金
台湾科学委员会项目(NSC101-2221-E-167-027-MY3)
四川省教育厅科研项目(15ZB0378)
成都大学校青年基金资助项目(2015XJZ02)
关键词
混合物实验设计法
气体传感器
丙酮
电极
选择性
mixture experimental design method
gas sensor
acetone
electrode
selectivity