基于SiPM的乳腺专用PET探测模块研制与评价 被引量:3

Development and Evaluation of SiPM Based Detector for Positron Emission Mammography
下载PDF
导出
摘要 基于单点型硅光电倍增管(Si PM)与LYSO闪烁晶体,研制了一种新型乳腺专用PET探测模块,并对其性能进行研究。该探测模块采用28×28根1.9 mm×1.9 mm×15 mm的LYSO晶体组成阵列,与8×8个单点型Si PM组成的阵列通过光导耦合,所产生的脉冲信号分别经过模拟、定时电路处理后送入高速采样电路完成信号处理。该探测模块能够实现1.9 mm位置分辨,12.73%平均能量分辨以及2.66 ns双探头符合时间分辨,具有较好的分辨性能,能够满足乳腺专用PET成像的设计需求。 We have developed a detector module based on Silicon Photomultiplier (SiPM) and LYSO for Posi- tron Emission Mammography (PEM) imaging. The module consists of a 28 x 28 array of 1.9 x 1.9 x 15 mm3 ( pitch 2 mm) LYSO elements. The scintillation array is optically coupled to an 8 x 8 single channel SiPM array through a light guide. The moduleg signals were read out using analog circuit, timing circuit and then processed by the high speed data acquisition system. The results of performance test are presented. The detector module has a position resolution of 1.9 mm; has an average energy resolution of 12.7% ; has a coincidence timing reso- lution of 2.66 ns. The results show that the detector module has good resolutions performance, which could meet the needs of PEM imaging.
出处 《核电子学与探测技术》 CAS 北大核心 2016年第3期316-320,共5页 Nuclear Electronics & Detection Technology
基金 国家自然科学基金资助项目(11475206) 中科院高能所科技创新项目(542013IHPZZBS80162)资助
关键词 硅光电倍增管(SiPM) 乳腺专用PET 探测器 能量分辨 符合时间分辨 SiPM position emission mammography (PEM) detector energy resolution coincidence time res-olution
  • 相关文献

参考文献10

  • 1H Zaidi, C Thompson. Evolution and Developments in Instrumentation for Positron Emission Mammography [J]. Pet Clinics, 2009, 4(4) :317 -327.
  • 2C. J. Thompson. Feasibility study for positron emis- sion mammography [J]. MEDICAL PHYSICS, 1994,21 (4) :529 -538.
  • 3MC Abreu, JD Aguiar, F'G Almeida, et al. Design and Evaluation of the Clear - PEM Scanner for Posi- tron Emission Mammography [K]. IEEE Transactions on Nuclear Science, 2006, 53 (1) :71 - 77.
  • 4LI Lin, GU Xiao - yue, LI Dao-wu, et al. Perform- ance Evaluation and Initial Clinical Test of the Posi- tron Emission Mammography System (PEMi) [J]. IEEE Transactions on Nuclear Science, 2015, 62 (5):1 -9.
  • 5Guerra A D, Belcari N, Bisogni M G, et al. Silicon Photomultipliers (SiPM) as novel photodetectors for PET [J]. Nuclear Instruments & Methods in Physics Research, 2011, 648(17) :$232 $235.
  • 6V Popov, S Majewski, AG Weisenberger. Readout e- lectronics for multi - anode photomultiplier tubes with pad matrix anode layout [J]. IEEE Nuclear ScienceSymposium Conference Record, 2003, 567( 1 ):2156 -2159 Vol. 3.
  • 7帅磊,丰宝桐,胡婷婷,李道武,魏书军,黄欢,李可,柴培,李琳,单保慈,魏龙,徐文贵.乳腺专用PET连续采样电子学系统设计与性能测试[J].原子能科学技术,2011,45(5):628-632. 被引量:2
  • 8Neves J A. The Clear PEM breast imaging scanner [J]. Nuclear Instruments & Methods in Physics Re- search A, 2011, 628(1):444-447.
  • 9Moadel R M. Breast cancer imaging devices [J]. Seminars in Nuclear Medicine, 2011, 41 ( 3 ) : 229 241.
  • 10Li L, Gu X, Li D, et al. Ability of the Positron E- mission Mammography system, PEMi, in detection of millimeter- sized lesions [C]// Nuclear Science Symposium and Medical Imaging Conference (NSS/ MIC), 2013 IEEE. IEEE, 2013:1 -7.

二级参考文献1

共引文献1

同被引文献15

引证文献3

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部